Intellijoint HIP Generation 2B System

K171525 · Intellijoint Surgical, Inc. · OLO · Oct 30, 2017 · Neurology

Device Facts

Record IDK171525
Device NameIntellijoint HIP Generation 2B System
ApplicantIntellijoint Surgical, Inc.
Product CodeOLO · Neurology
Decision DateOct 30, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

The Intellijoint HIP Generation 2B System is a computer-controlled, optical localizer intended to provide intra-operative measurements to a surgeon to aid in selection and positioning of orthopedic implant system components, where a reference to a rigid anatomical structure can be identified. The Intellijoint HIP Generation 2B System is indicated for patients undergoing orthopedic surgery, and where the use of stereotactic surgery is considered safe and effective. The system aids the surgeon in performing intra-operative measurements including measurements of limb position, joint center-of-rotation, and implant component positioning. Example orthopedic surgical procedures include, but are not limited to: - Total Hip Arthroplasty - Minimally Invasive Hip Arthroplasty - Revision Hip Arthroplasty

Device Story

Imageless optical navigation system for orthopedic surgery; provides intra-operative assessment of leg length, offset, anterior-posterior change, hip center-of-rotation change, and acetabular cup angle. System components: infrared camera, tracker, computer workstation, software, and bone fixation hardware. Used in OR by surgeons during hip arthroplasty. Inputs: optical tracking data from trackers fixed to patient anatomy. Processing: system calculates spatial relationships between anatomical structures and implant components. Output: real-time measurements displayed on workstation for surgeon. Benefits: aids in accurate selection and positioning of orthopedic implants.

Clinical Evidence

No clinical data. Evidence consists of benchtop accuracy testing using calibrated fixtures, software functional/unit testing, and simulated use validation in anatomical phantoms and cadaver labs by orthopedic surgeons to confirm clinical accuracy and user needs.

Technological Characteristics

Imageless optical navigation system; infrared camera-based tracking; computer workstation; bone fixation instruments. Software-based measurement of spatial parameters. Connectivity: workstation-based. Sterilization: not specified.

Indications for Use

Indicated for patients undergoing orthopedic surgery (e.g., Total Hip Arthroplasty, Minimally Invasive Hip Arthroplasty, Revision Hip Arthroplasty) where stereotactic surgery is safe and effective.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized eagle. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with "ADMINISTRATION" underneath. October 30, 2017 Intellijoint Surgical Inc. Brandon Gingrich Quality and Regulatory Affairs Manager 60 Bathurst Drive. Unit 6 Waterloo, Ontario N2V 2A9 Canada Re: K171525 Trade/Device Name: Intellijoint HIP Generation 2B System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: October 5, 2017 Received: October 6, 2017 Dear Brandon Gingrich: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K171525 Device Name Intellijoint HIP Generation 2B System #### Indications for Use (Describe) The Intellijoint HIP Generation 2B System is a computer-controlled, optical localizer intended to provide intra-operative measurements to a surgeon to aid in selection and positioning of orthopedic implant system components, where a reference to a rigid anatomical structure can be identified. The Intellijoint HIP Generation 2B System is indicated for patients undergoing orthopedic surgery, and where the use of stereotactic surgery is considered safe and effective. The surgeon in performing intra-operative measurements including measurements of limb position, joint center-of-rotation, and implant component positioning. Example orthopedic surgical procedures include, but are not limited to: - Total Hip Arthroplasty - Minimally Invasive Hip Arthroplasty - Revision Hip Arthroplasty | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary # 1. Submitter Information | Submitter: | Intellijoint Surgical Inc. | |----------------|-----------------------------------------------------------| | Address: | 60 Bathurst Dr., Unit 6<br>Waterloo, ON<br>Canada N2V 2A9 | | Telephone: | (519) 342-3178 | | Fax: | (226) 317-0471 | | Contact: | Brandon Gingrich | | Date Prepared: | May 23, 2017 | ### 2. Device Information | Trade Name: | Intellijoint HIP® Generation 2B System | |----------------------|----------------------------------------| | Common Name: | Orthopedic Stereotaxic Instrument | | Classification: | Class II per 21 CFR 882.4560 | | Classification Name: | Orthopedic Stereotaxic Instrument | | Product Code: | OLO | ### 3. Purpose of Submission The purpose of this submission is to gain clearance for updates to the indications for use for a previously cleared Computer-Assisted Orthopedic Surgery System. # 4. Predicate Device Information The Intellijoint HIP® Generation 2B System described in this submission is substantially equivalent to the following predicates: | Predicate Device | Manufacturer | 510(k) No. | |----------------------------------------|----------------------------|------------| | Intellijoint HIP® Generation 2A System | Intellijoint Surgical Inc. | K162364 | | Stryker OrthoMap Versatile Hip System | Stryker Leibinger GmbH | K162937 | # 5. Device Description The Intellijoint HIP® Generation 2B System is an imageless optical navigation system intended for use in orthopedic surgery. The device provides intra-operative assessment of patient leg length, offset, anterior-posterior change, hip center of rotation change, and acetabular cup angle during Total Hip Arthroplasty procedures. The system is composed of an infrared Camera, Tracker, computer workstation, software, and bone fixation instruments/hardware. The Intellijoint HIP® Generation 2B System is an update to the Intellijoint HIP® System previously cleared in 510(k) K162364. This submission provides the addition of revision hip arthroplasty (RHA) procedures to the indications for use, and the use of either linear {4}------------------------------------------------ Image /page/4/Picture/2 description: The image shows the logo for Intellijoint Surgical. The logo consists of a blue geometric shape on the left and the word "intellijoint" in gray on the right. Below "intellijoint" is the word "surgical" in a smaller font size, also in gray. The geometric shape appears to be made up of three triangles. or non-linear acetabular cup impactors. Other minor modifications made to the device are also provided. #### 6. Intended Use The Intellijoint HIP Generation 2B System is a computer-controlled, optical localizer intended to provide intra-operative measurements to a surgeon to aid in selection and positioning of orthopedic implant system components, where a reference to a rigid anatomical structure can be identified. The Intellijoint HIP Generation 2B System is indicated for patients undergoing orthopedic surgery, and where the use of stereotactic surgery is considered safe and effective. The system aids the surgeon in performing intra-operative measurements including measurements of limb position, joint center-of-rotation, and implant component positioning. Example orthopedic surgical procedures include, but are not limited to: - -Total Hip Arthroplasty - -Minimally Invasive Hip Arthroplasty - Revision Hip Arthroplasty - #### 7. Comparison of Technological Characteristics The substantial equivalence of the Intellijoint HIP® Generation 2B System to the predicates is shown by similarity in intended use, indications for use, and performance. #### 8. Performance Data This submission is for updates to the Intellijoint HIP® Generation 2A System cleared in 510(k) K162364. The following tests were performed to demonstrate the substantial equivalence of the updated Intellijoint HIP® Generation 2B System to its predicate devices: | Test | Summary | Result | |------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------| | Verification | | | | Benchtop Accuracy | Verified clinical accuracy requirements using<br>calibrated benchtop test fixtures. | All accuracy<br>requirements<br>were met. | | Software Functional<br>and Unit Tests | Verified that the software application<br>satisfies functional requirements and<br>performs as intended. Algorithms and<br>measurement calculations were also verified<br>in these tests. | Software<br>satisfied all<br>requirements and<br>specifications. | | Bone Fixation<br>Functional and<br>Performance Tests | Provides confirmation that new hardware<br>(Femoral Disc and Femoral Rod) satisfy<br>functional and performance requirements. | All functional<br>and performance<br>requirements<br>met. | {5}------------------------------------------------ Image /page/5/Picture/2 description: The image shows the logo for Intellijoint Surgical. The logo consists of a geometric shape on the left and the company name on the right. The geometric shape is made up of three triangles in different shades of blue and gray. The company name is written in a sans-serif font, with the word "intellijoint" in bold and the word "surgical" in a smaller font below it. | Validation | | | |--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------| | Anatomical Phantom<br>Simulated Use and<br>Clinical Accuracy | Simulated use testing was performed on bone<br>models by orthopedic surgeons in a<br>simulated THA procedure following a typical<br>workflow. This test validated that the<br>Intellijoint HIP® Generation 2B System<br>satisfies user needs, intended use and clinical<br>accuracy requirements. Accuracy was<br>assessed by comparing simulated use<br>measurements with ground truth values. | All user needs<br>and clinical<br>accuracy<br>requirements<br>were met. | | Cadaver Simulated<br>Use | Simulated use testing was also performed in<br>a cadaver lab. This test validated that the<br>Intellijoint HIP® Generation 2B System<br>satisfies clinical use requirements and<br>performs as intended when:<br>- Operated by a surgeon<br>- Used on human specimens<br>- Used in a realistic operating room<br>environment | All clinical use<br>requirements<br>were met. | The testing demonstrated that the Intellijoint HIP® Generation 2B System is substantially equivalent to the legally marketed predicate devices for its intended use. # 9. Conclusion Based on the indications for use, technological characteristics, performance testing, and comparison to the predicate devices, the Intellijoint HIP® Generation 2B System has been shown to be substantially equivalent to the legally marketed predicate devices identified in this submission, and does not present any changes to safety or effectiveness.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...