← Product Code [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO) · K163379

# KneeAlign 2 System (K163379)

_Orthalign, Inc. · OLO · Mar 2, 2017 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K163379

## Device Facts

- **Applicant:** Orthalign, Inc.
- **Product Code:** [OLO](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/OLO.md)
- **Decision Date:** Mar 2, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The KneeAlign® system is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical structures during stereotactic orthopedic surgical procedures. The KneeAlign system facilitates the accurate positioning of implants and instrumentation, relative to these alignment axes. Orthopedic surgical procedures include but are not limited to: Total Knee Arthroplasty Unicompartmental Knee Arthroplasty - Tibial Transverse Resection

## Device Story

KneeAlign 2 System is a non-invasive, computer-assisted surgical navigation system for knee arthroplasty. It uses a palm-sized, single-use computer module and reusable reference sensors to track anatomical landmarks. Surgeon registers landmarks (malleoli, tibial plateau) to establish reference frames; device calculates mechanical axes (femoral/tibial) and determines varus/valgus and flexion/extension/slope angles. Output is displayed on an integrated graphical interface on the module, guiding the surgeon in positioning cutting blocks for bone resection. Used in the OR by surgeons to improve implant alignment accuracy. Benefits include precise bone cuts and optimized implant positioning, potentially improving surgical outcomes.

## Clinical Evidence

No clinical data. Evidence consists of bench testing and simulated-use validation. Bench testing verified navigated resection plane angular and depth accuracy using mechanical fixtures and foam models. A prospective cadaver validation was conducted in a simulated operating room environment by a surgeon to confirm the system meets design requirements for unicompartmental knee arthroplasty procedures.

## Technological Characteristics

System comprises a single-use computer module and reusable instrumentation (jigs, cutting blocks). Sensing principle: inertial sensors, microcontroller, and digital signal processor. Connectivity: standalone. Biocompatibility: ISO 10993-1 compliant. Sterilization: validated for reusable components. Software: algorithm-based spatial coordinate calculation.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- OrthAlign Plus® System ([K153237](/device/K153237.md))
- KneeAlign® 2 System ([K103829](/device/K103829.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, with flowing lines extending from the bottom of the profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 2, 2017

OrthAlign, Inc. % David Vancelette Director QA/RA Orthalign, Inc. 120 Columbia, Suite 500 Aliso Viejo, California 92656

Re: K163379

Trade/Device Name: KneeAlign 2 System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: November 30, 2016 Received: December 1, 2016

Dear David Vancelette:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K163379

Device Name KneeAlign 2 System

Indications for Use (Describe)

The Knee Align® system is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical structures during stereotactic surgical procedures. The KneeAlien system facilitates the accurate positioning of implants and instrumentation, relative to these alignment axes.

Orthopedic surgical procedures include but are not limited to:

- · Total Knee Arthroplasty
- · Unicompartmental Knee Arthroplasty Tibial Transverse Resection

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

{3}------------------------------------------------

#### 5. 510(K) SUMMARY

This 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92(c).

| DATE                      | November 30, 2016                                                                                                                                                     |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT                 | OrthAlign, Inc.<br>120 Columbia<br>Suite 500<br>Aliso Viejo, CA 92656<br>Tel: (949) 715-2424<br>Fax: (949) 831-9500                                                   |
| OFFICIAL<br>CORRESPONDENT | David Vancelette<br>OrthAlign, Inc.<br>120 Columbia,<br>Suite 500<br>Aliso Viejo, CA 92656<br>dvancelette@orthalign.com<br>Tel: (949) 525-9034<br>Fax: (949) 831-9500 |
| TRADE NAME                | KneeAlign® 2 System                                                                                                                                                   |
| COMMON NAME               | Stereotaxic Instrument                                                                                                                                                |
| DEVICE<br>CLASSIFICATION  | Class II, 21 CFR §882.4560                                                                                                                                            |
| PRODUCT CODES             | OLO: Orthopedic Stereotaxic Instrument                                                                                                                                |
| PREDICATE<br>DEVICES      | OrthAlign Plus® System (K153237)<br>KneeAlign® 2 System (K103829)                                                                                                     |
| SUBMISSION TYPE           | Traditional 510(k). The subject device is a modification<br>to the previously cleared KneeAlign® 2 System<br>(K103829).                                               |

#### SUBSTANTIALLY EQUIVALENT TO:

The KneeAlign®2 System is substantially equivalent to the previously cleared KneeAlign® 2 System (K103829) and OrthAlign Plus® System (K153237).

#### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The KneeAlign® 2 System is a non-invasive computer assisted surgical navigation system for use in knee arthroplasty procedures. The KneeAlign® 2 System is configured to detect, measure, and display angular and positional measurement changes in a triaxial

{4}------------------------------------------------

format.

The KneeAlign® 2 System utilizes a palm-sized computer module and reference sensor to generate positional information in orthopedic procedures providing a sequence of steps for registration of anatomical landmarks, calculation of mechanical axes, and positioning of instruments relative to the mechanical axes.

In total knee arthroplasty procedures, the device assists the surgeon in:

- . Establishing the mechanical axis of the femur, determining the varus/valgus angle and the flexion/extension angle of the cutting block relative to the femur.
- Establishing the mechanical axis of the tibia, determining the varus/valgus . angle and the posterior slope angle of the cutting block relative to the tibia.

In unicompartmental knee arthroplasty procedures, the device assists the surgeon in:

- Establishing the mechanical axis of the tibia, determining the varus/valgus . angle and the posterior slope angle of the cutting block relative to the tibia, for the transverse resection.
The KneeAlign® 2 System comprises a single use computer module and reusable instrumentation.

### INDICATIONS FOR USE:

The KneeAlign® 2 System has the same indications for use as the previously cleared KneeAlign® 2 System (K103829). Additional functionality has been added to the predicate device to enable the navigation of tibial transverse resections in unicompartmental knee arthroplasty. Also, Indications for Use are common to the OrthAlign Plus® System (K153237). Thus, the Indications for Use are as follows:

### KneeAlign® 2 System:

The KneeAlign® system is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical structures during stereotactic orthopedic surgical procedures. The KneeAlign system facilitates the accurate positioning of implants and instrumentation, relative to these alignment axes.

Orthopedic surgical procedures include but are not limited to:

- . Total Knee Arthroplasty
- . Unicompartmental Knee Arthroplasty - Tibial Transverse Resection

### TECHNICAL CHARACTERISTICS (COMPARED TO PREDICATE):

The KneeAlign® 2 System was cleared under K103829. The KneeAlign® 2 System comprises a single use computer module, a reusable reference sensor, a reusable femoral iig and a reusable tibial jig. The device utilizes algorithms to convert sensor outputs into spatial coordinates, providing graphical and numerical representation of instruments and 

{5}------------------------------------------------

anatomy on the user display screen. The KneeAlign®2 System is being updated to include navigation functionality for unicompartmental knee arthroplasty, as in the predicate device OrthAlign Plus® System (K153237). All other features and principles of operation remain unchanged.

#### PERFORMANCE DATA:

Device performance testing confirms that the KneeAlign® 2 System can be used according to its intended use. The KneeAlign® 2 System has been verified and validated according to OrthAlign's procedures for product design and development. Performance testing addressed only the added Unicompartments and surgical procedure steps. Performance testing included:

- System hardware verification/validation testing to ensure the instruments meet . their mechanical requirements.
- . Instrumentation cleaning, sterilization and shipping validations for the specified processes.
- . Navigation device sterilization, packaging, shelf life, environmental conditions and shipping validations for the specified ranges of conditions involved in each process. (Summary data for the identical predicate device is referenced.)
- System components biocompatibility assessment per ISO 10993-1 (2009). ●
- Customer requirements / usability validation in cadaver with an advising ● surgeon to validate the system meets design input requirements for its functions in a simulated use environment.
- . System accuracy testing: bench testing with mechanical fixtures and foam models to verify navigated resection plane angular and depth accuracy.

For simulated use testing, a prospective cadaver validation was done in a simulated operating room environment with a surgeon conducting the procedures.

This testing regime demonstrates that the subject device is as safe and effective as the predicate devices. This testing regime demonstrates that the subject device is substantially equivalent to the legally marketed predicate devices, for its intended use in the accurate navigation of tibial resection planes in unicompartmental knee arthroplasty.

The information provided by OrthAlign in this 510(k) application confirms that the KneeAlign® 2 System is substantially equivalent to predicate devices such as the KneeAlign® 2 System (K103829) and OrthAlign Plus® System (K153237).

#### BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE:

A technological comparison and bench and simulated use testing demonstrate the substantial equivalence of the KneeAlign® 2 System to the predicate devices.

{6}------------------------------------------------

The subject device is identical to the predicate KneeAlign® 2 System (K103829), with the following exceptions:

- The subject device allows for the navigation of the tibial transverse resection . plane in unicompartmental knee arthroplasty.
The subject device is substantially equivalent in terms of unicompartmental knee arthroplasty function to the predicate device OrthAlign Plus® System (K153237).

The table below summarizes the features of the subject device as compared to the predicate devices.

{7}------------------------------------------------

# 510(K) SUMMARY

## Table 1. KneeAlign® 2 System Comparison to Predicates

| Property                                           | KNEEALIGN® 2 SYSTEM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | KNEEALIGN® 2 SYSTEM (K103829)<br>PREDICATE 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | ORTHALIGN PLUS® SYSTEM (K153237)<br>PREDICATE 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use                                | Total Knee Arthroplasty /<br>Unicompartmental Knee Arthroplasty:<br>The KneeAlign® system is a computer-<br>controlled system intended to assist the<br>surgeon in determining reference<br>alignment axes in relation to anatomical<br>structures during stereotactic orthopedic<br>surgical procedures. The KneeAlign<br>system facilitates the accurate positioning<br>of implants and instrumentation, relative<br>to these alignment axes.<br>Orthopedic surgical procedures include<br>but are not limited to:<br>Total Knee Arthroplasty Unicompartmental Knee<br>Arthroplasty - Tibial<br>Transverse Resection | Total Knee Arthroplasty<br>The KneeAlign® system is a computer-<br>controlled system intended to assist the<br>surgeon in determining reference<br>alignment axes in relation to anatomical<br>structures during stereotactic orthopedic<br>surgical procedures. The KneeAlign<br>system facilitates the accurate<br>positioning of implants and<br>instrumentation, relative to these<br>alignment axes.<br>Orthopedic surgical procedures include<br>but are not limited to:<br>Total Knee Arthroplasty <span>▶</span> Unicompartmental Knee<br>Arthroplasty function is being<br>added. | Total Knee Arthroplasty / Total Hip<br>Arthroplasty / Unicompartmental Knee<br>Arthroplasty:<br>The OrthAlign Plus®System is a<br>computer-controlled system intended to<br>assist the surgeon in determining<br>reference alignment axes in relation to<br>anatomical and instrumentation<br>structures during stereotactic orthopedic<br>surgical procedures. The OrthAlign<br>Plus® System facilitates the accurate<br>positioning of implants, relative to these<br>alignment axes. The system aids the<br>surgeon in controlling leg length and<br>offset discrepancies in Total Hip<br>Arthroplasty.<br>Example orthopedic surgical procedures<br>include but are not limited to:<br>Total Knee Arthroplasty Total Hip Arthroplasty:<br>Anterior/Posterior Unicompartmental Knee<br>Arthroplasty: Tibial transverse<br>resection <span>▶</span> Same Indications for Use. |
| Property                                           | KNEEALIGN® 2 SYSTEM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | KNEEALIGN® 2 SYSTEM (K103829)<br>PREDICATE 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | ORTHALIGN PLUS® SYSTEM (K153237)<br>PREDICATE 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Technological Principles                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Computer generation of<br>positional information   | Uses inertial sensors, microcontroller and<br>digital signal processor to generate<br>positional information.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Registration of anatomy                            | Lateral and medial malleoli and the center<br>of the tibial plateau are registered to<br>establish a tibial reference frame.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Navigation of tibial<br>transverse resection plane | Posterior slope is in the sagittal plan of<br>the tibia. Varus/Valgus is in the coronal<br>plane of the tibia.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Controls resection depth                           | Via mechanical instruments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Surgical work flow                                 | Patient preparation Instrument setup Instrument attachment to patient Anatomic registrations Cutting block navigation Resection                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Design Elements                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Main System Components                             | Single-use computer unit Navigation software Reusable instrument set Registration instruments Cutting blocks                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| User Interface                                     | Integrated graphical user interface, on<br>single-use unit that attached to<br>instrumentation.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Tibial Cutting Blocks                              | Mechanical instrument that interfaces<br>with tibial jig for variable positioning                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |

{8}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K163379](https://fda.innolitics.com/submissions/OR/subpart-e%E2%80%94neurological-surgical-devices/OLO/K163379)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
