Navigated XIA 4.5 Polyaxial Screwdriver

K143597 · Stryker Corporate · OLO · Sep 4, 2015 · Neurology

Device Facts

Record IDK143597
Device NameNavigated XIA 4.5 Polyaxial Screwdriver
ApplicantStryker Corporate
Product CodeOLO · Neurology
Decision DateSep 4, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Stryker Xia 4.5 Polyaxial Screwdriver is a Stryker Navigated Manual Surgical Instrument. Stryker Navigated Manual Surgical Instruments are intended to be used as accessories to the Stryker Spine Navigation System, when used with the SpineMap 3D Navigation software. They are manual surgical instruments intended to be used in spine surgery to facilitate placement of Stryker Spine implants. Stryker Navigated Spine Instruments may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified. Stryker Navigated Spine Instruments are intended for exclusive use with the Rotational Adaptor and associated Trackers to facilitate the placement of the Stryker Spine XIA® 3, XIA® 4.5, MANTIS®, MANTIS® Redux, and ES2® Spinal Fixation Systems using the STRYKER SpineMap® 3D Navigation System. The surgeon must also refer to the Stryker Spine XIA® 3, XIA® 4.5, MANTIS® Redux, or ES2® package insert/instructions for use, product label, and surgical technique guide to obtain detailed product information and recommended surgical procedure.

Device Story

Manual surgical instrument for bone screw placement, insertion, and removal in spine surgery; interfaces with handle and Rotational Navigation Adapter; connects to Stryker Spine Navigation System via instrument tracker. Used by surgeons in clinical settings for intraoperative guidance. Device provides physical interface for screw manipulation; navigation functionality provided by external system accessories (Rotational Navigation Adapter and Instrument Tracker). Output is physical screw placement guided by navigation system visualization. Benefits include improved accuracy in pedicle screw placement during spinal fixation procedures.

Clinical Evidence

No clinical studies were performed. Evidence consists of bench testing, including simulated use in a cadaveric setting and navigation accuracy verification. Accuracy was verified to be within ± 2 mm and ± 2° (80th percentile < ± 1.5 mm) per ASTM F2554:2010.

Technological Characteristics

Manual surgical instrument; stainless steel/metal construction; interfaces with Rotational Navigation Adapter and Instrument Tracker. Navigation accuracy verified per ASTM F2554:2010. No internal electronics or software; passive tracking via external system components.

Indications for Use

Indicated for use in spine surgery for patients requiring computer-assisted planning and surgery for the placement of Stryker Spine XIA 3, XIA 4.5, MANTIS, MANTIS Redux, and ES2 spinal fixation systems, where a reference to a rigid anatomical structure can be identified.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing right, with flowing lines representing connection and support. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Stryker Leibinger GmbH & Company KG Ms. Becky Ditty Senior Staff Regulatory Affairs Specialist 4100 E. Milham Avenue Kalamazoo, Michigan 49001 September 4, 2015 Re: K143597 Trade/Device Name: Navigated XIA® 4.5 Polyaxial Screwdriver Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: OLO Dated: August 5, 2015 Received: August 7, 2015 Dear Ms. Ditty: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {2}------------------------------------------------ Page 2 - Ms. Becky Ditty forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K143597 Device Name Navigated XIA® 4.5 Polyaxial Screwdriver #### Indications for Use (Describe) The Stryker Xia 4.5 Polyaxial Screwdriver is a Stryker Navigated Manual Surgical Instrument. Stryker Navigated Manual Surgical Instruments are intended to be used as accessories to the Stryker Spine Navigation System, when used with the SpineMap 3D Navigation software. They are manual surgical instruments intended to be used in spine surgery to facilitate placement of Stryker Spine implants. Stryker Navigated Spine Instruments may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative guidance where a reference to a rigid anatomical structure can be identified. Stryker Navigated Spine Instruments are intended for exclusive use with the Rotational Adaptor and associated Trackers to facilitate the placement of the Stryker Spine XIA® 3, XIA® 4.5, MANTIS®, MANTIS® Redux, and ES2® Spinal Fixation Systems using the STRYKER SpineMap® 3D Navigation System. The surgeon must also refer to the Stryker Spine XIA® 3, XIA® 4.5, MANTIS® Redux, or ES2® package insert/instructions for use, product label, and surgical technique guide to obtain detailed product information and recommended surgical procedure. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED#### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary This chapter provides a summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92. # Section 5.1 Submitter | Applicant Name: | Stryker Leibinger GmbH & Co. KG<br>Boetzinger Strasse 41<br>D-79111 Freiburg, Germany<br>Phone number: +49-761-4512117<br>Fax number: +49-761-451249117 | |-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------| | Registration No.: | 8010177 | | Name of Contact Person: | Becky Ditty<br>Sr. Staff Regulatory Affairs Specialist<br>4100 E. Milham Ave<br>Kalamazoo, MI 49001<br>becky.ditty@stryker.com<br>(269) 389-3434 | | Date Prepared: | September 4, 2015 | # Section 5.2 Device | Trade Name: | Navigated XIA® 4.5 Polyaxial Screwdriver | | | | | |-------------------------|------------------------------------------|------------------------------------------------|----------------------|-------|-----------------| | Common Name: | Navigated Screwdriver | | | | | | Classification<br>Name: | Product<br>Code | Device | Regulation<br>Number | Class | Review<br>Panel | | | Primary<br>Code<br>OLO | Orthopedic<br>Stereotaxic<br>Instrument | 21 CFR<br>§882.4560 | II | Orthopedic | | | Secondary<br>Code: LXH | Orthopedic<br>Manual<br>Surgical<br>Instrument | 21 CFR<br>§882.4540 | I | Orthopedic | {5}------------------------------------------------ ### Section 5.3 Predicate Devices | Trade Name | Stryker Navigation System –<br>Spine & Fluoroscopy Module:<br>Stryker Adapted Smart Manual<br>Instruments (Primary Predicate) | XIA ® 4.5 Polyaxial Screwdriver | |---------------|-------------------------------------------------------------------------------------------------------------------------------|----------------------------------------| | 510(k) Number | K012380 | N/A — Class I/ 510(k) exempt<br>device | | Product Code | HAW | LXH | | Manufacturer | Stryker Leibinger GmbH & Co. KG | Stryker Spine SAS | These predicate devices have not been subject to a design-related recall. # Section 5.4 Device Description The Navigated XIA® 4.5 Polyaxial Screwdriver, a Stryker Navigated Spine Instrument, is a manual tool for bone screw placement, insertion, and removal. The tip of the navigated screwdriver inserts into the screw head to turn it. The screwdriver also has an interface to allow a connection to the Rotational Navigation Adapter, a Stryker Spine Navigation System accessory, and to a handle. Image /page/4/Figure/6 description: The image shows two different views of a navigated XIA 4.5 polyaxial screwdriver. The top view shows the screwdriver with the text "XIA 4.5 SCREWDRIVER" printed on the side. The bottom view shows the screwdriver with labels pointing to the handle, rotational navigation adapter, instrument tracker, shaft, and bone screw. The figure is labeled as "FIGURE 5-1: NAVIGATED XIA 4.5 POLYAXIAL SCREWDRIVER". FIGURE 5-2: NAVIGATED XIA 4.5 POLYAXIAL SCREWDRIVER ATTACHED TO ROTATION NAVIGATION ADAPTER, INSTRUMENT TRACKER AND HANDLE {6}------------------------------------------------ #### Section 5.5 Indications for Use The Stryker Xia 4.5 Polyaxial Screwdriver is a Stryker Navigated Manual Surgical Instrument. Stryker Navigated Manual Surgical Instruments are intended to be used as accessories to the Stryker Spine Navigation System, when used with the SpineMap 3D Navigation software. They are manual surgical instruments intended to be used in spine surgery to facilitate placement of Stryker Spine implants. Stryker Navigated Spine Instruments may be used as part of the Stryker Spine Navigation System, which is indicated for any medical condition in which the use of computer assisted planning and surgery may be appropriate. The system can be used for intraoperative quidance where a reference to a rigid anatomical structure can be identified. Stryker Navigated Spine Instruments are intended for exclusive use with the Rotational Navigational Adaptor and associated Trackers to facilitate the placement of the pedicle screws of the Stryker Spine XIA® 3, XIA® 4.5, MANTIS® Redux, and ES2® Spinal Fixation Systems using the STRYKER SpineMap® 3D Navigation System. The surgeon must also refer to the Stryker Spine XIA® 3, XIA® 4.5, MANTIS®, MANTIS® Redux, or ES2® package insert/instructions for use, product label, and surgical technique guide to obtain detailed product information and recommended surgical procedure. # Section 5.6 Comparison of Technological Characteristics with the Predicate Device The Navigated XIA 4.5 Polyaxial Screwdriver (subject device) and the predicate devices are intended to be used during the preparation and placement of Stryker Spine screws during spinal surgery, including insertion and removal of bone screws. The subject and predicate devices have similar technologies, such that the devices interface with a handle and have a tip that the user inserts into the screw head to turn it. Additionally, the subject device and the Stryker Adapted Smart Manual Instruments predicate device are both designed to interface with Stryker Navigation Systems via the previously cleared Stryker Spine Navigation System accessories: the Rotational Navigation Adapter (K012380) and an Instrument Tracker (K141941). The {7}------------------------------------------------ attachment of the Rotational Navigation Adapter and Instrument Tracker to the instrument allows for the localization and navigation of the manual surgical instruments and thus the technology is not included in the subject device itself. The subject device includes minor design changes as compared to the predicate devices. The main change is a longer shaft to enable the attachment of the Rotational Navigation Adapter; however the subject device also has a modified locking mechanism, and can interface with a wider range of handles. The instrument modifications detailed in this submission have no impact on the technological characteristics of either the existing instruments or the Stryker Spine Navigation Systems. None of the changes identified affect how the spinal screw is placed or removed. The navigation of the manual instrument is not performed by the instrument itself, but by previously cleared Stryker Spine Navigation System accessories. The accuracy testing provided demonstrates that the Stryker Spine Navigation System accessories (Rotational Navigation Adapter and Instrument Tracker) can be sufficiently connected to the Navigated XIA 4.5 Polyaxial Screwdriver such that the accuracy requirements of the Stryker Spine Navigation System can be achieved. #### Section 5.7 Performance Data The following performance data were provided in support of the substantial equivalence decision: | Test | Description | |-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Simulated Use | Tested the Navigated XIA 4.5 Polyaxial Screwdriver together with the<br>Stryker Spine Navigation System and its components and accessories in a<br>cadaveric setting according to its requirements. | | Navigation Accuracy | Verified the navigation accuracy of $\pm$ 2 mm and $\pm$ 2° of the Navigated XIA<br>4.5 Polyaxial Screwdriver by verifying its tracking accuracy of (80th<br>percentile < $\pm$ 1.5 mm) according to ASTM F2554:2010. | | TABLE 5.1: V&V OVERVIEW | | #### Performance Testing - Bench #### Performance Testing - Animal No animal studies were performed to support substantial equivalence. #### Performance Testing - Clinical No clinical studies were performed to support substantial equivalence. {8}------------------------------------------------ # Section 5.8 Conclusion The non-clinical data support the safety of the device and verification and validation demonstrate that the Navigated XIA 4.5 Polyaxial Screwdriver device performs as intended in the specified use conditions. The data also demonstrates that the Navigated XIA 4.5 Polyaxial Screwdriver performs comparably to the predicate devices that are currently marketed for the same intended use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...