Uni-FuZe-P Bone Putty
K201781 · Zavation Medical Products, LLC · MQV · Dec 28, 2020 · Orthopedic
Device Facts
| Record ID | K201781 |
| Device Name | Uni-FuZe-P Bone Putty |
| Applicant | Zavation Medical Products, LLC |
| Product Code | MQV · Orthopedic |
| Decision Date | Dec 28, 2020 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3045 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
Uni-FuZe-P Bone Putty is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. Uni-FuZe-P Bone Putty is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e.,extremities and pelvis). These defects may be surgically created osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.
Device Story
Uni-FuZe-P Bone Putty is a resorbable, bioactive, osteoconductive bone graft substitute. It consists of a matrix of bovine collagen, beta-tricalcium phosphate (Beta-TCP), Bioglass 45S5, and polyethylene glycol (PEG). The device is provided sterile for single-use. It is applied by a clinician (surgeon) by gently packing the moldable/malleable putty into bony voids or gaps in the extremities or pelvis. The putty acts as a scaffold, filling the defect while resorbing over time to be replaced by the patient's own bone during the natural healing process. It is intended for non-load-bearing applications where the bone void does not compromise structural stability.
Clinical Evidence
No human clinical data. Evidence includes bench testing and an animal study. Biocompatibility confirmed per ISO 10993-1 and ISO 10993-6. In vivo performance evaluated in a rabbit functional femoral critical defect model (n=not specified) at 1 day, 6 weeks, and 12 weeks, comparing subject device to predicate. Endpoints included radiographic, micro-CT, and histological analyses. Results showed normal osteoconductive healing, absence of adverse inflammatory reactions, and performance equivalent to the predicate. Bioactivity confirmed per ISO/FDIS 23317; material composition verified via SEM, XRD, and SDS-Page.
Technological Characteristics
Materials: Type I bovine collagen (ASTM F2212), Beta-TCP (ASTM F1088), Bioglass 45S5 (ASTM F1538), and PEG (1450/600 blend). Form: Malleable putty. Principle: Bioactive osteoconductive scaffold. Sterilization: Gamma irradiation (VDmax, ANSI/AAMI/ISO 11137-2, 25 kGy). Packaging validated per ASTM F88/F88M and ASTM F2096. Shipping validated per ASTM D7386.
Indications for Use
Indicated for bony voids or gaps in the skeletal system (extremities and pelvis) not intrinsic to structural stability. Applicable to surgically created osseous defects or defects from traumatic injury. No specific age or gender contraindications listed.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Predicate Devices
- MasterGraft Putty (K051386)
Reference Devices
- Sorrento Bioglass Bone Graft Substitute (K173933)
Related Devices
- K051386 — MASTERGRAFT PUTTY BONE VOID FILLER · Medtronic Vascular · Dec 16, 2005
- K180080 — FIBERGRAFT BG Matrix Bone Graft Substitute · Prosidyan, Inc. · Apr 6, 2018
- K143533 — FIBERGRAFT BG PUTTY BONE GRAFT SUBSTITUTE · Prosidyan, Inc. · Mar 25, 2015
- K173933 — Sorrento Bioglass Bone Graft Substitute · Xenco Medical, LLC · Sep 13, 2018
- K110368 — NOVABONE PUTTY - BIOACTIVE SYNTHETIC BONE GRAFT · Novabone Products, LLC · Apr 14, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
December 28, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Zavation Medical Products, LLC % Robert Poggie, Ph.D. President BioVera Inc. 65 Promenade Saint Louis Notre-Dame-del-L'Ile-Perrot, QC J7V 7P2 Canada
Re: K201781
Trade/Device Name: Uni-FuZe-P Bone Putty Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: December 2, 2020 Received: December 4, 2020
Dear Dr. Poggie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K201781
Device Name
Uni-FuZe-P Bone Putty
Indications for Use (Describe)
Uni-FuZe-P Bone Putty is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. Uni-FuZe-P Bone Putty is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e.,extremities and pelvis). These defects may be surgically created osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K201781 SUMMARY for Zavation Uni-FuZe-P Bone Putty
In accordance with 21 CFR 807.92 of the Federal Code of Regulations, the following information is a summary of safety and effectiveness of Zavation Uni-FuZe-P Bone Putty.
## A. SUBMITTERS INFORMATION
| Submitter Name: | BioVera, Inc. |
|------------------------|------------------------------------------------------------|
| Submitter Address: | 65 Promenade Saint-Louis, NDIP, Québec, J7V 7P2,<br>CANADA |
| Contact Person: | Robert A. Poggie, PhD |
| Phone and Fax Numbers: | 514-901-0796 |
| Date of Submission: | June 29, 2020 |
## B. DEVICE IDENTIFICATION & MANUFACTURER
| Manufacturer Name: | Zavation Medical Products, LLC |
|-----------------------|-------------------------------------|
| Manufacturer Address: | 220 Lakeland Parkway |
| | Flowood, MS 39232 USA |
| Registration Number: | 3008583793 |
| Contact Name: | Jeffrey Johnson |
| Title: | CEO |
| Device Trade Name: | Uni-FuZe-P Bone Putty |
| Device Common Name: | Bone Void Filler |
| Classification Name: | Filler, bone void, calcium compound |
| Classification Code: | MQV |
| Classification Panel: | Orthopedic |
| Regulation Number: | 21 CFR sections 888.3045 |
| C1. PRIMARY PREDICATE DEVICE | |
|------------------------------|-------------------------------------------------------------|
| K051386 | Medtronic Sofamor Danek USA, MasterGraft Putty |
| C2. REFERENCE DEVICE | |
| K173933 | Xenco Medical, LLC, Sorrento Bioglass Bone Graft Substitute |
{4}------------------------------------------------
### D. DEVICE DESCRIPTION
Uni-FuZe-P Bone Putty is a resorbable bone void filler made from a matrix of bovine collagen (ASTM F2212), beta tricalcium phosphate (Beta-TCP per ASTM F1088), Bioglass 45S5 (ASTM F1538), and polyethylene glycol (PEG, 1450/600 blend). The implant is provided as sterile and is for single use.
### E. INDICATIONS FOR USE
Uni-FuZe-P Bone Putty is indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure. Uni-FuZe-P Bone Putty is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e. extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process.
#### F. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Uni-FuZe-P Bone Putty is a bioactive osteoconductive, resorbable, biocompatible bone graft substitute in putty form. The product is composed of matrix of bovine collagen per ASTM F2212, beta tricalcium phosphate (Beta-TCP per ASTM F1088), Bioglass 45S5 per ASTM F1538, and polyethylene glycol (PEG, 1450/600 blend). Uni-FuZe-P Bone Putty can be applied directly to the defect site or used as a moldable, and malleable material to be placed at the defect site.
Uni-FuZe-P Bone Putty is similar in composition to the reference device, Xenco Medical's Sorrento Bioglass Bone Graft Substitute, and has similar clinications for use as the predicate MasterGraft Putty device and reference Sorrento Bioglass Bone Graft Substitute. The subject, predicate, and reference devices are comprised of similar base materials (collagen, HA / TCP, bioglass), resorbable in relatively short period of time, available in similar volumes, and can be manipulated to fit the bone void. The table below summarizes the technological characteristics of the subject, predicate, and reference devices and supports equivalency of the products.
| | Subject Device Zavation Uni-<br>FuZe-P Bone Putty | Predicate Device, K051386,<br>Medtronic MasterGraft Putty | Reference Device, K173933,<br>Sorrento Bioglass Bone Graft |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Regulatory<br>Classifications | MQV | MQV | MQV |
| Materials | Beta-TCP ASTM F1088<br>Type I bovine collagen per<br>ASTM F2212<br>Bioglass 45S5 per ASTM F1538<br>Polyethylene glycol | Beta-TCP and HA<br>Type I bovine collagen | Beta-TCP ASTM F1088<br>Type I bovine collagen per<br>ASTM F2212<br>Bioglass 45S5 per ASTM F1538 |
| Physical form | Putty | Putty | Strip |
| Dosage | 2.5, 5, 10 cc | | 2, 5, 10, 12, 20, 24 cc |
| Resorbable | Yes | Yes | Yes |
| Porosity | Granular, high surface area | Granular, high surface area | Granular, high surface area |
| Sterile and<br>single use? | Yes | Yes | Yes |
{5}------------------------------------------------
# G. PERFORMANCE DATA
Assessment of biocompatibility of Uni-FuZe-P Bone Putty was performed by Nelson Labs according to ISO 10993-1:2018, and for biological effects according to FDA Guidance, "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." Material-Mediated Pyrogenicity, Irritation, Sensitization, and Implantation tests demonstrated acceptable biological safety profiles.
Biocompatibility per ISO 10993-6 was confirmed through short- and long-term implantation at time points of 4 and 13 weeks. Bone healing and biological response testing were evaluated using an established rabbit functional femoral critical defect model at time points of 1 day and 6- and 12-weeks following implantation for the subject device Uni-FuZe-P Bone Putty and the predicate device MasterGraft Putty. Biological performance was measured and documented with radiographic images, micro-CT, and histological analyses.
The macro observations of the implant sites demonstrated healthy tissue absent of adverse inflammatory reactions for Uni-FuZe-P Bone Putty. Radiographic and microCT analyses indicated no adverse reactions and a normal progression in healing over time. Histopathology assessment showed normal osteoconductive healing response. The study confirmed the biocompatibility and normal osteoconductive healing response associated with the Uni-FuZe-P and demonstrated substantially equivalent in vivo performance to the predicate device across all endpoints.
The bioactivity of the bioglass component of the Uni-FuZe-P Bone Putty was confirmed per ISO/FDIS 23317; the presence of HA was verified by SEM and XRD.
The stability of the collagen within the subject device was verified using SDS-Page.
Packaging seal strength and integrity was validated via peel strength (per ASTM F88/F88M) and bubble emission (per ASTM F2096); acceptance criteria were met.
Shipping and handling validations was performed per ASTM D7386; acceptance criteria were met.
Sterilization was validated to an SAL of 10° using method VDmax of ANSI/AAMI/JSO 11137-2 with 25 kGy selected as the minimum sterilization dose. Limulus Amebocyte Lysate (LAL) testing demonstrated the maximum dose for one surgery has less than 20 endotoxin units (< 20 EU).
#### H. CONCLUSIONS
Uni-FuZe-P Bone Putty has similar indications for use, technological characteristics, and principles of operation as the predicate and reference devices. The minor technological differences between Uni-FuZe-P Bone Putty and the predicate and reference devices do not raise new issues of safety or effectiveness. The performance data demonstrate that Uni-FuZe-P Bone Putty is substantially equivalent to the predicate device.