TRICOS A RESORBABLE BONE SUBSTITUTE
K081717 · Baxter Healthcare Corp · MQV · Oct 10, 2008 · Orthopedic
Device Facts
| Record ID | K081717 |
| Device Name | TRICOS A RESORBABLE BONE SUBSTITUTE |
| Applicant | Baxter Healthcare Corp |
| Product Code | MQV · Orthopedic |
| Decision Date | Oct 10, 2008 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3045 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
TricOs A Resorbable Bone Substitute is indicated for use as a bone void filler for voids and gaps that are not intrinsic to the stability of the bony structure. It is indicated for surgically created osseous defects or osseous defects resulting from traumatic injury. TricOs A Resorbable Bone Substitute is intended to be packed into bony voids or gaps of the skeletal system (i.e., the extremities, spine, and pelvis). Following placement in bony voids or gaps, TricOs A Resorbable Bone Substitute is resorbed while providing support for the in-growth of bone during the healing process. TricOs A Resorbable Bone Substitute is a bone void filler without initial mechanical properties, therefore rigid fixation techniques are recommended.
Device Story
TricOs A is a resorbable bone void filler composed of an inorganic calcium phosphate scaffold mixed with a heterologous human fibrin matrix. The fibrin matrix acts as a binder to improve handling and provides a 3D scaffold for bone in-growth. The device is applied by a surgeon to fill osseous gaps or voids in the skeletal system (extremities, spine, pelvis). It lacks initial mechanical strength, necessitating the use of rigid fixation techniques. The device is resorbed over time as natural bone heals and replaces the scaffold. It benefits patients by supporting bone regeneration in non-load-bearing defects.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing, biocompatibility testing, and animal studies using a rabbit femoral condyles model to verify performance as a bone void filler.
Technological Characteristics
Inorganic calcium phosphate scaffold with heterologous human fibrin matrix binder. Resorbable. Form factor: bone void filler. Sterilization: standard processing. Complies with ISO 14971:2007 for risk management.
Indications for Use
Indicated for patients with surgically created osseous defects or traumatic osseous defects in the skeletal system (extremities, spine, pelvis) requiring bone void filling. Contraindicated for voids intrinsic to bony structure stability; requires rigid fixation.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Predicate Devices
- TricOs T Resorbable Bone Substitute (K051722)
- TricOs T Resorbable Bone Substitute (K073571)
Related Devices
- K051722 — TRICOS T · Baxter Healthcare Corp · Nov 18, 2005
- K081721 — TRICOS A RESORBABLE SUBSTITUTE · Baxter Healthcare Corp · Aug 6, 2008
- K073571 — TRICOS T RESORBABLE BONE SUBSTITUTE · Baxter Healthcare Corporation · Apr 8, 2008
- K213111 — MagnetOs granules · Kuros Biosciences B.V · Jan 10, 2022
- K143692 — GranOSTM - beta tricalcium phosphate synthetic bone granules · Tcm Associates, Ltd. · Aug 21, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
and the comments of the comments of the comments of the contraction of the contribution of the contribution of the contribution of the contribution of the contribution of the
______________________________________________________________________________________________________________________________________________________________________________
OCT 1 . 2008
Page 1 of 3 Section 5, 510(k) Šummary
# K 081 717
| 5. 510(K) SUMMARY |
|-------------------|
|-------------------|
| DATE: | June 16, 2008 | | | | | | | | | | |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--------------------------------------|--------------|------------------|-----------------|------------------------------------------|--------|----------|---------------|-----|
| OWNER: | Baxter Healthcare Corporation | | | | | | | | | | |
| | One Deerfield Parkway | | | | | | | | | | |
| | Deerfield, IL 60015 | | | | | | | | | | |
| CONTACT PERSON: | Barbara Barbeau | | | | | | | | | | |
| | Senior Director, Global Regulatory Affairs | | | | | | | | | | |
| | 1620 Waukegan Road, MPGR-AL | | | | | | | | | | |
| | McGaw Park, IL 60085 | | | | | | | | | | |
| | Telephone: 847-270-4174 | | | | | | | | | | |
| | Fax: 847-785-5116 | | | | | | | | | | |
| | Email: <a href="mailto:Barbara_Barbeau@baxter.com">Barbara_Barbeau@baxter.com</a> | | | | | | | | | | |
| DEVICE NAME: | | | | | | | | | | | |
| | <table><tr><th>Trade Name:</th><td>Tricos A¹ Resorbable Bone Substitute</td></tr><tr><th>Common Name:</th><td>Bone Void Filler</td></tr><tr><th>Classification:</th><td>Resorbable Calcium Salt Bone Void Filler</td></tr><tr><th>Class:</th><td>Class II</td></tr><tr><th>Product Code:</th><td>MQV</td></tr></table> | Trade Name: | Tricos A¹ Resorbable Bone Substitute | Common Name: | Bone Void Filler | Classification: | Resorbable Calcium Salt Bone Void Filler | Class: | Class II | Product Code: | MQV |
| Trade Name: | Tricos A¹ Resorbable Bone Substitute | | | | | | | | | | |
| Common Name: | Bone Void Filler | | | | | | | | | | |
| Classification: | Resorbable Calcium Salt Bone Void Filler | | | | | | | | | | |
| Class: | Class II | | | | | | | | | | |
| Product Code: | MQV | | | | | | | | | | |
| PREDICATE DEVICES: | Previously cleared 510(k)s for | | | | | | | | | | |
| | Baxter Healthcare Corporation, Tricos T Resorbable Bone Substitute product. | | | | | | | | | | |
| Previous 510(k) | Indication | Clearance Date |
|-----------------|--------------------------------------------------------------|-------------------|
| K051722 | Bone void filling of<br>the skeletal system | November 18, 2005 |
| K073571 | Bone void filling of<br>the oral and<br>maxillofacial region | April 8, 2008 |
<sup>1</sup> TricOs A is a trademark of Baxter Healthcare Corporation
{1}------------------------------------------------
## DEVICE DESCRIPTION:
TricOs A Resorbable Bone Substitute consists of an inorganic calcium phosphate scaffold that is mixed with a heterologous human fibrin matrix prior to application. The fibrin matrix acts as a binder for the calcium phosphate scaffold, imparting favorable handling characteristics to the product to facilitate the surgical procedure, and acting as a three-dimensional matrix that supports the in-growth of bone.
STATEMENT OF TricOs A Resorbable Bone Substitute is indicated for INTENDED USE: use as a bone void filler for voids and gaps that are not intrinsic to the stability of the bony structure. It is indicated for surgically created osseous defects or osseous defects resulting from traumatic injury. TricOs A Resorbable Bone Substitute is intended to be packed into bony voids or gaps of the skeletal system (i.e., the extremities, spine, and pelvis). Following placement in bony voids or gaps, TricOs A Resorbable Bone Substitute is resorbed while providing support for the in-growth of bone during the healing process. TricOs A Resorbable Bone Substitute is a bone void filler without initial mechanical properties, therefore rigid fixation techniques are recommended.
## TECHNOLOGICAL CHARACTERISTICS:
TricOs A Resorbable Bone Substitute for the skeletal system is substantially equivalent to Baxter's current legally marketed TricOs T Resorbable Bone Substitute cleared by 510(k) K051722 with regard to intended use. TricOs A Resorbable Bone Substitute for the skeletal system is substantially equivalent to TricOs T Resorbable Bone Substitute cleared by 510(k)s K051722 and K073571 with regard to technological characteristics and performance.
{2}------------------------------------------------
## ASSESSMENT OF NONCLINICAL DATA:
Baxter Heathcare conducted a risk assessment according to the requirements of ISO 14971:2007 Medical Devices - Application of Risk Management to Medical Devices. The TricOs T and TricOs A device material and component specifications are similar. The device components continue to meet the same material testing standards and sterilization processing standards. Device performance as a bone void filler of the skeleton system has been verified in animal studies using a rabbit femoral condyles model and through functional and biocompatibility testing.
## CONCLUSIONS:
The TricOs A Resorbable Bone Substitute is substantially equivalent to the predicate devices. Testing has been verified against established standards and guidelines for its intended use and demonstrates that the proposed device is as safe and effective for the intended use as the predicate devices.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the eagle. The logo is black and white.
Public Health Service
OCT 1 0 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Baxter Healthcare Corporation % Ms. Valerie Followell Manager, Global Regulatory Affairs 1620 Waulkegan Road, MPGR-AL McGaw Park, Illinois 60085
Re: K081717
Trade/Device Name: TricOs A Resorbable Bone Substitute Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: II Product Code: MQV Dated: October 06, 2008 Received: October 07, 2008
Dear Ms. Followell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Or ug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 – Ms. Valerie Followell
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N Mulhearn
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### 4. INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
New special 510(k)
Device Name:
TricOs A | Resorbable Bone Substitute
Indication(s) for Use:
TricOs A Resorbable Bone Substitute is indicated for use as a bone void filler for voids and gaps that are not intrinsic to the stability of the bony structure. It is indicated for surgically created osseous defects or osseous defects resulting from traumatic injury. TricOs A Resorbable Bone Substitute is intended to be packed into bony voids or gaps of the skeletal system (i.e., the extremities, spine, pelvis). Following placement in bony voids or gaps, TricOs A Resorbable Bone Substitute is resorbed while providing support for the in-growth of bone during the healing process. TricOs A Resorbable Bone Substitute is a bone void filler without initial mechanical properties, therefore rigid fixation techniques are recommended.
| Prescription Use: | <div> <span style="font-size: 16px;"> </span> </div> <div> </div> |
|-----------------------|-------------------------------------------------------------------|
| 21 CFR 801 Subpart D | |
| Over-the-Counter Use: | <div> <span style="font-size: 16px;"> </span> </div> <div> </div> |
| 21 CFR Subpart C | |
Neil RP Onder for nxm
41-51-28
Concurrence of CDRH, Office of Device Evalualizes of
Division of General, Restorative, and Neurological Devices
1 TricOs A is a trademark of Baxter Healthcare Corporation
**510(k) Number** K081717