DYNA-LOK SPINAL SYSTEM
K983672 · Sofamor Danek USA,Inc. · MNI · Dec 16, 1998 · Orthopedic
Device Facts
| Record ID | K983672 |
| Device Name | DYNA-LOK SPINAL SYSTEM |
| Applicant | Sofamor Danek USA,Inc. |
| Product Code | MNI · Orthopedic |
| Decision Date | Dec 16, 1998 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3070 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
When used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients, the DYNA-LOK® Spinal System is indicated for one or more of the following: (1) degenerative spondylolisthesis with objective evidence of neurologic impairment. (2) fracture, (3) dislocation, (4) scoliosis, (5) kyphosis, (6) spinal tumor, and/or (7) failed previous fusion (pseudarthrosis). In addition, when used as a pedicle screw fixation system, the DYNA-LOK® Spinal System is indicated for skeletally mature patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (4) who are having the device removed after the development of a solid fusion mass. When used as a posterior, non-cervical, non-pedicle screw fixation system, the DYNA-LOK® Spinal System is intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (5) fracture, (6) pseudarthrosis, (7) tumor resection, and/or (8) failed previous fusion. When used as an anterolateral thoracic/lumbar system, the DYNA-LOK® Spinal System is intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (5) fracture, (6) pseudarthrosis, (7) tumor resection, and/or (8) failed previous fusion.
Device Story
DYNA-LOK® Spinal System is a spinal stabilization construct consisting of plates, bolts, screws, and hooks. Used in posterior non-cervical or anterolateral thoracic/lumbar procedures to facilitate solid spinal fusion. Implanted by surgeons in clinical settings. Provides mechanical stabilization to the spine during the fusion process. Components are fabricated from medical-grade stainless steel or titanium alloy. System is provided sterile or non-sterile.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Construct components: plates, bolts, screws, hooks. Materials: medical-grade stainless steel (ASTM F138, ISO 5832-1/9) or titanium alloy (ASTM F136, ISO 5832-3). Mechanical stabilization system.
Indications for Use
Indicated for skeletally mature patients requiring spinal stabilization for degenerative spondylolisthesis, fracture, dislocation, scoliosis, kyphosis, spinal tumor, pseudarthrosis, degenerative disc disease, spinal stenosis, or tumor resection. Specific pedicle screw use for severe L5-S1 spondylolisthesis (Grades 3-4) with autogenous bone graft.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Related Devices
- K020155 — DYNA-LOK CLASSIC SPINAL SYSTEM · Medtronic Sofamor Danek · Feb 5, 2002
- K964254 — TITANIUM LIBERTY POSTERIOR SPINAL SYSTEM · Sofamor Danek USA,Inc. · Apr 21, 1997
- K253335 — CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · Oct 24, 2025
- K033271 — DYNALOK CLASSIC SPINAL SYSTEM · Medtronic Sofamor Danek · Nov 6, 2003
- K200322 — Altus Spine Pedicle Screw System · Altus Partners, LLC · Jun 2, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
K983672
# DYNA-LOK® Spinal System 510(k) Summary
## October, 15 1998
- l. Sofamor Danek USA Company: 1800 Pyramid Place Memphis, TN 38132 (901) 396-3133
#### Proposed Proprietary Trade Name: DYNA-LOK® Spinal System .
#### Description lll.
The DYNA-LOK® Spinal System consists of plates, bolts, screws, hooks, CROSSLINK® Plates and other components used to build a spinal construct. Instrumentation is also available to facilitate implantation of the device components. The purpose of the DYNA-LOK® Spinal System is to provide stabilization during the development of a solid spinal fusion.
The DYNA-LOK® Spinal System implant components are fabricated from medical grade stainless steel described by such standards as ASTM F138 or ISO 5832-1 or ISO 5832-9. Alternatively, the entire system may be made out of medical grade titanium alloy described by such standards as ASTM F136 or ISO 5832-3. The DYNA-LOK® Spinal System implant Components may be sold either sterile or non-sterile
#### Indications for Use IV.
When used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients, the DYNA-LOK® Spinal System is indicated for one or more of the following: (1) degenerative spondylolisthesis with objective evidence of neurologic impairment. (2) fracture, (3) dislocation, (4) scoliosis, (5) kyphosis, (6) spinal tumor, and/or (7) failed previous fusion (pseudarthrosis).
In addition, when used as a pedicle screw fixation system, the DYNA-LOK® Spinal System is indicated for skeletally mature patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (4) who are having the device removed after the development of a solid fusion mass.
{1}------------------------------------------------
When used as a posterior, non-cervical, non-pedicle screw fixation system, the DYNA-LOK® Spinal System is intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (5) fracture, (6) pseudarthrosis, (7) tumor resection, and/or (8) failed previous fusion.
When used as an anterolateral thoracic/lumbar system, the DYNA-LOK® Spinal System is intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (5) fracture, (6) pseudarthrosis, (7) tumor resection, and/or (8) failed previous fusion.
#### V. Substantial Equivalence
The DYNA-LOK® Spinal System is substantially equivalent to itself. The new pedicle screw classification indications described in the classification Final Rule published in the July 27.1998 Federal Register were used to expand the previously cleared indications for the DYNA-LOK® Spinal Svstem.
©1998 Sofamor Danek
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference of the circle. Inside the circle is an image of three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# DEC 1 6 1998
Richard W. Treharne, Ph.D. Vice President Research and Regulatory Affairs Sofamor Danek 1800 Pvramid Place 38132 Memphis, Tennessee
Re: K983672 DYNA-LOK® Spinal System Trade Name: Regulatory Class: II Product Codes: MNI, MNH, KWP, and KWQ Dated: October 15, 1998 Received: October 19, 1998
Dear Dr. Treharne:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Richard W. Treharne, Ph.D.
This letter will allow you to begin marketing your device as The FDA described in your 510(k) premarket notification. finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K983672 510(k) Number (if known):
Device Name: DYNA-LOK® Spinal System
### Indications for Use:
When used as a pedicle screw fixation system of the non-cervical posterior spine in skeletally mature patients, the DYNA-LOK® Spinal System is indicated for one or more of the following: (1) degenerative spondylolisthesis with objective evidence of neurologic impairment, (2) fracture, (3) dislocation, (4) scoliosis, (5) kyphosis, (6) spinal turnor, and/or (7) failed previous fusion (pseudarthrosis).
In addition, when used as a pedicle screw fixation system, the DYNA-LOK® Spinal System is indicated for skeletally mature patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and below); and (4) who are having the device removed after the development of a solid fusion mass.
When used as a posterior, non-cervical, non-pedicle screw fixation system, the DYNA-LOK® Spinal System is intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) spinal deformities (i.e., scoliosis, kyphosis, and/or lordosis), (5) fracture, (6) pseudarthrosis, (7) tumor resection, and/or (8) failed previous fusion.
When used as an anterolateral thoracic/lumbar system, the DYNA-LOK® Spinal System is intended for the following indications: (1) degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), (2) spinal stenosis, (3) spondylolisthesis, (4) spinal deformities (i.e., scoliosis, and/or lordosis), (5) fracture, (6) pseudarthrosis, (7) tumor resection, and/or (8) failed previous fusion.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Evaluation (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | X |
|------------------------------------------|---|
|------------------------------------------|---|
OR
| Over-The-Counter Use | |
|----------------------|-------------------|
| | (Optional 1-2-96) |
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number. K983672