CORE PEDICLE SCREW SYSTEM
K131522 · Core-Nexus · MNI · Aug 9, 2013 · Orthopedic
Device Facts
| Record ID | K131522 |
| Device Name | CORE PEDICLE SCREW SYSTEM |
| Applicant | Core-Nexus |
| Product Code | MNI · Orthopedic |
| Decision Date | Aug 9, 2013 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3070 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Core Pedicle Screw System is a posterior pedicle screw system indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. In addition, the Core Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudarthrosis). The Core Hook System is intended for use as a posterior, noncervical, nonpedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma(i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion. The Core Hook System can be used in conjunction with Core Pedicle Screw System.
Device Story
Top-loading posterior spinal fixation system; components include pedicle screws, rods, hooks, set screws, connectors, and transverse linking mechanism. Used by surgeons in clinical settings to build spinal implant constructs for stabilization and fusion. Provides immobilization of spinal segments for acute/chronic instabilities or deformities. Implants are single-use; supplied non-sterile. System facilitates surgical correction of spinal pathologies; benefits patient through structural stabilization of spine during fusion process.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Posterior spinal fixation system; components: pedicle screws, rods, hooks, set screws, connectors, transverse links. Material: Titanium alloy (Ti-6Al-4V ELI) per ASTM F136. Non-sterile, single-use. Mechanical fixation principle.
Indications for Use
Indicated for skeletally mature patients requiring spinal stabilization/fusion. Conditions: severe spondylolisthesis (Grade 3-4, L5-S1), degenerative spondylolisthesis with neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, spinal stenosis, degenerative disc disease, and pseudarthrosis. Used in thoracic, lumbar, and sacral spine.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Predicate Devices
- Dio Medical Rexious Spinal Fixation System (K113324, K111362, K100765)
Related Devices
- K081145 — MODIFICATION TO 4CIS VANE SPINE SYSTEM · Solco Biomedical Co., Ltd. · Nov 5, 2008
- K222229 — Advanced LumFix Spinal Fixation System · CG Bio Co., Ltd. · May 25, 2023
- K111362 — REXIOUS SPINAL FIXATION SYSTEM · Dio Medical Co., Ltd. · Jul 21, 2011
- K222415 — Rexious Spinal Fixation System · Dio Medical Corporation · Oct 25, 2022
- K060702 — 4CIS VANE SPINE SYSTEM · Solco Biomedical Co., Ltd. · Jun 22, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
### 510(k) Summary Core Pedicle Screw System Premarket Notification
AUG 9 2013
SUBMITTED BY
Core-Nexus 15760 Ventura Blvd 7th Floor Encino, CA 91436
ESTABLISHMENT REGISTRATION NUMBER
# Pending
Pending
Thomas Palmer President Core-Nexus
May 15, 2013
Class II
OWNER/OPERATOR NUMBER
CONTACT PERSON
SUBMISSION PREPARED BY
Lisa Peterson Kaedon Consulting, LLC Email: Ipeterson@kaedonconsulting.com Phone: 512-507-0746
Email: tpalmer@core-nexus.com
Phone: 818-424-4270
DATE PREPARED
CLASSIFICATION NAME
DEVICE CLASS
REGULATION NUMBER
COMMON NAME
PROPRIETARY NAME
Core Pedicle Screw System
888.3070 (Product Code MNI, MNH) 888.3050 (Product Code KWP)
Pedicle Screw Spinal Fixation System
IDENTIFICATION OF PREDICATE DEVICE(S)
Dio Medical Rexious Spinal Fixation System (K113324, K111362, K100765)
Spondylolisthesis Spinal Fixation, Spinal Pedicle Fixation
{1}------------------------------------------------
#### DEVICE DESCRIPTION
The Core Pedicle Screw System is a top-loading multiple component, posterior spinal fixation system that consists of pedicle screws, rods, hooks, set screws, connectors, and a transverse (cross) linking mechanism. The Core Pedicle Screw System allows surgeons to build a spinal implant construct to stabilize and promote spinal fusion.
#### INDICATIONS
The Core Pedicle Screw System is a posterior pedicle screw system indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
In addition, the Core Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
The Core Hook System is intended for use as a posterior, noncervical, nonpedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma(i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis); tumor; pseudoarthrosis; and failed previous fusion. The Core Hook System can be used in conjunction with Core Pedicle Screw System.
#### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The purpose of this 510(k) is to obtain clearance to market the subject device as the Core Pedicle Screw System. The subject device was previously cleared as the Dio Medical Rexious Spinal Fixation System (K113324, K111362, K100765). Documentation the legal right to distribute the Core Pedicle Screw System is maintained at Core-Nexus.
The subject device is identical to the previously cleared Rexious System in terms of indications for use, device dimensions, instrumentation, manufacturing process, cleaning/sterilization process and labeling. The System implant components are supplied non-sterile, are single use, and are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F136.
## DISCUSSION OF NON-CLINICAL TESTING
Non-clinical testing was not performed as part of this submission.
#### CONCLUSIONS
The subject device is identical to the predicate device in terms of indications for use, device dimensions. instrumentation, manufacturing process, cleaning/sterilization process and labeling. Documentation was provided as part of this 510(k) to demonstrate that the Core Pedicle Screw System is substantially equivalent to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a stylized human figure with outstretched arms.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 9, 2013
Core-Nexus % Kaedon Consulting, LLC Ms. Lisa Peterson Principal Consultant 14001 Hunters Pass Austin, Texas 78734
Re: K131522
Trade/Device Name: Core Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, MNH, KWP Dated: June 24, 2013 Received: June 27, 2013
Dear Ms. Peterson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Ms. Lisa Peterson
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Erin I. Keith
For
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known): K131522
Core Pedicle Screw System Device Name:
Indications for Use:
The Core Pedicle Screw System is a posterior pedicle screw system indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
In addition, the Core Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudarthrosis).
The Core Hook System is intended for use as a posterior, noncervical, nonpedicle system indicated for degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma(i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion. The Core Hook System can be used in conjunction with Core Pedicle Screw System.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Core-Nexus
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Colin O'Neill for RPJ
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K131522