ZAVATION SPINAL SYSTEM
K112484 · Zavation, LLC · MNI · Nov 22, 2011 · Orthopedic
Device Facts
| Record ID | K112484 |
| Device Name | ZAVATION SPINAL SYSTEM |
| Applicant | Zavation, LLC |
| Product Code | MNI · Orthopedic |
| Decision Date | Nov 22, 2011 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3070 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Zavation Spinal System is a pedicle screw system intended to provide Immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The Zavation Spinal Systems is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. The Zavation Spinal Systems when used as anterior thoracic/lumbar screw fixation systems, is indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis).
Device Story
Zavation Spinal System; rod-based spinal fixation system; components include polyaxial pedicle screws, rods, and crosslinks. Used for single or multiple level spinal fixation; provides immobilization and stabilization as an adjunct to fusion. Components manufactured from titanium alloy. System implanted by surgeons in clinical/OR settings to treat spinal instabilities or deformities. Mechanical performance validated via static/dynamic compression, bending, and torsion testing per ASTM F1717. Output is structural stabilization of spinal segments; facilitates bone fusion; intended to reduce pain and restore spinal alignment.
Clinical Evidence
Bench testing only. Static compression bending, torsion, and dynamic compression bending performed per ASTM F1717 on worst-case constructs. Results demonstrate performance equivalent or superior to predicate devices.
Technological Characteristics
Rod-based spinal fixation system; polyaxial pedicle screws; variable length cross connectors; straight and pre-lordosed rods. Material: Titanium alloy (Ti-6Al-4V) per ASTM F136. Mechanical testing per ASTM F1717.
Indications for Use
Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion. Treats thoracic, lumbar, and sacral spine instabilities/deformities including degenerative spondylolisthesis, fractures, dislocations, scoliosis, kyphosis, spinal tumors, pseudarthrosis, severe spondylolisthesis (Grades 3-4 at L5-S1), degenerative disc disease, and spinal stenosis.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Predicate Devices
- Depuy Spine Moss Miami Spine System (K955348)
- Depuy Spine Expedium Spine System (K033901)
Related Devices
- K153404 — Zavation Spinal System · Zavation, LLC · Apr 21, 2016
- K190317 — Zavation Spinal System · Zavation Medical Products, LLC · Apr 11, 2019
- K222614 — Zavation Spinal System · Zavation Medical Products, LLC · Dec 5, 2022
- K190361 — Zavation Spinal System · Zavation Medical Products, LLC · Mar 18, 2019
- K023498 — MODIFICATION TO BLACKSTONE SPINAL FIXATION SYSTEM · Blackstone Medical, Inc. · Nov 13, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
K11 248<-1
NOV 222 2011
# 510K Summary
| Date: | August 27, 2011 |
|----------------------------|-------------------------------------------------------------------------------------------------|
| Submitter: | Zavation LLC<br>501 Avalon Way<br>Brandon, MS 39047<br>Phone: 601-919-1119<br>Fax: 800-447-1302 |
| Contact person: | John Walker |
| Type of 510(k) submission: | Traditional |
| Trade name: | Zavation Spinal System |
| Common name: | Spinal Fixation System |
| Classification regulation: | 888.3070 (MNH, MNI)<br>888.3060 (KWQ) |
| Device classification: | Class II |
| Classification Panel: | Orthopedic |
| Product code: | MNH, MNI, KWQ |
| Basis for submission: | New device |
Device Description: The Zavation Spinal System is comprised of polyaxial pedicle screws, rods and crosslinks. The Zavation Spinal System can be used for single or multiple level fixations. The pedicle screws are available in various lengths and diameters. The rods are available in straight and pre-lordosed (curved) configurations. The system has variable length cross connectors.
Intended Use: The Zavation Spinal System is a pedicle screw system intended to provide lmmobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).
{1}------------------------------------------------
K1124841
The Zavation Spinal Systems is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
The Zavation Spinal Systems when used as anterior thoracic/lumbar screw fixation systems, is indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis).
#### Materials:
The Zavation Spinal System components are manufactured from titanium alloy (Ti-6Al-4V) as described by ASTM F136.
### Predicate Device:
Depuy Spine Moss Miami Spine System (K955348) Depuy Spine Expedium Spine System (K033901)
#### Technological Characteristics:
The Zavation Spinal System possesses the same technological characteristics as the predicate devices. These include: basic design (rod based fixation system having polyaxial pedicle screws with various screw and rod diameters and lengths), material (titanium alloy), mechanical safety and performances, and intended use (as described above).
### Performance Data:
Static compression bending and torsion, and dynamic compression bending were performed according to ASTM F1717 on a worst-case construct. The mechanical test results demonstrated that the Zavation Spinal System performs as well as or better than the predicate devices.
#### Conclusion:
The Zavation Spinal System is substantially equivalent to the device referenced above and is therefore safe and effective for its intended use.
Pg 2 of 2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
NOV 2 2 2011
Zavation LLC % Mr. John Walker Engineering Manager 501 Avalon Way Brandon, Mississippi 39047
Re: K112484
Trade/Device Name: Zavation Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, MNH, KWP Dated: October 06, 2011 Received: October 14, 2011
Dear Mr. Walker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
Page 2 - Mr. John Walker
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erin Keith
orMark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
### 510(k) Number: K112484
Device Name: Zavation Spinal System
Indications For Use:
The Zavation Spinal System is a pedicle screw system intended to provide Immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
The Zavation Spinal Systems is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
The Zavation Spinal Systems when used as anterior thoracic/lumbar screw fixation systems, is indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis).
| Prescription Use | X |
|------------------|---|
|------------------|---|
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
| 510(k) Number | K112484 |
|---------------|---------|
| Page | 1 of 1 |
Pg 1 of 1