ROMEO POSTERIOR OSTEOSYNTHESIS SYSTEM

K112108 · Spineart · MNI · Aug 19, 2011 · Orthopedic

Device Facts

Record IDK112108
Device NameROMEO POSTERIOR OSTEOSYNTHESIS SYSTEM
ApplicantSpineart
Product CodeMNI · Orthopedic
Decision DateAug 19, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Intended Use

ROMEO posterior osteosynthesis system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudarthrosis). When used as a posterior, non-cervical, non-pedicle screw fixation system, ROMEO posterior osteosynthesis system is intended for the following indications for the following indications; degenerative disc disease(DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

Device Story

ROMEO posterior osteosynthesis system consists of spinal implants including pedicle screws and hooks; used for immobilization and stabilization of thoracic, lumbar, and sacral spine segments; adjunct to fusion. Components are implanted by surgeons in clinical/OR settings. System provides mechanical support to spinal column to treat instabilities, deformities, and degenerative conditions. Device components are supplied sterile or non-sterile. Clinical benefit derived from stabilization of spinal segments to facilitate fusion and address patient pain or deformity.

Clinical Evidence

Bench testing only. Mechanical testing including static compression bending and static torsion performed according to ASTM F1717-10. Results demonstrate that additional components perform as safely and effectively as predicate devices.

Technological Characteristics

Pedicle screw spinal system and spinal interlaminal fixation orthosis. Components include pedicle screws and hooks. Mechanical testing performed per ASTM F1717-10. No software or electronic components.

Indications for Use

Indicated for skeletally mature patients requiring spinal immobilization and stabilization as an adjunct to fusion for thoracic, lumbar, and sacral spine conditions including severe spondylolisthesis (L5-S1), degenerative spondylolisthesis with neurologic impairment, fracture, dislocation, scoliosis, kyphosis, lordosis, spinal tumor, spinal stenosis, pseudoarthrosis, and failed previous fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ AUG 19 2011 ## 510(k) SUMMARY As required by section 807.92 | Submitter | SPINEART<br>International Center Cointrin 20 route de pré-bois CP1813<br>1215 GENEVA 15<br>SWITZERLAND | |---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contacts | Franck PENNESI Director of Industry & Quality<br>Phone : +41 22 799 40 25<br>Fax : +41 22 799 40 26<br>Mail : fpennesi@spineart.ch<br>Regulatory contact : Dr Isabelle DRUBAIX (Idée Consulting)<br>idrubaix@nordnet.fr | | Trade Name | ROMEO posterior osteosynthesis system | | SPECIAL 510k | Modification to ROMEO posterior osteosynthesis system<br>(Extension of range of products) | | CFR section | 888.3070 and 888.3050 | | Classification Name | Pedicle screw spinal system / Spinal interlaminal fixation orthosis | | Class | II | | Product Codes | KWP Spinal interlaminal fixation orthosis<br>MNH orthosis, spondylolisthesis spinal fixation<br>MNI orthosis, spinal pedicle fixation | | Device panel | ORTHOPEDIC | | Legally marketed<br>predicate devices | ELLIPSE posterior osteosynthesis system (K081165) and<br>ROMEO posterior osteosynthesis system (K093170 and K<br>101678) manufactured by SPINEART | Description: The modifications to ROMEO posterior osteosynthesis system (K081165, K093170, K101678) manufactured by SPINEART consist of addition of addition of R093170, 11010707 manufacturers - Public Public and pedicle hooks and pedicle hooks (angled, Ma These components are supplied either sterile or not sterile. Intended Use ROMEO posterior osteosynthesis system is intended to provide Intended - 35 - 10 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 1 infinobilization and beautinent of the following acute and chronic instabilities or aujunct to tasion in the increases and sacral spine: severe spondylolisthesis (grades aelormities of the therefora; degenerative spondylolisthesis with objective evidence 9 und 17 of the 20 02 76; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and or nearonogie infpaintinosis). When used as a posterior, non-cervical, nonpedicle screw fixation system, ROMEO posterior osteosynthesis system is intended for pealcie screw indications for the following indications: degenerative disc disease(DDD) the following marcations discogenic origin with degeneration of the disc confirmed by (i.e., e. b.a. pan. ban pan. studies); spondylolisthesis; trauma (i.e., fracture or mistory and rainosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. {1}------------------------------------------------ Performance data ROMEO posterior osteosynthesis additional components conforms to special control established for Pedicle screw spinal system and to « Spinal System Special - Suidance for Industry and FDA Staff Document » issued on: May 3, 2004. Mechanical testing including static compression bending, static torsion, dynamic Mcchanical testing moraing performed according to ASTM F1717-10. Results compression that additional components perform as safely and effectively as their predicate devices. Substantial equivalence ROMEO posterior osteosynthesis system additional Substantial - agently and inter predicate device in terms of interms of intended use, material, design, mechanical properties and function. ass, matther, performance testing according to special control demonstrate that Non- ellined. Ponents are as safe, as effective, and performs as safely and effectively as their predicate devices. Preparation date, July 18, 2011 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle symbol, with three overlapping wing shapes, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Spineart % Mr. Franck Pennesi International Center Cointrin 20 Route de pre bois CP1813 1215 Geneva 15 Switzerland AUG 1 9 2011 Re: K112108 Trade/Device Name: ROMEO Posterior Osteosynthesis System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, MNH, KWP Dated: June 21, 2011 Received: June 22, 2011 Dear Mr. Pennesi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ ## Page 2 - Mr. Franck Pennesi or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Eil Keith -fr Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health · Enclosure {4}------------------------------------------------ ## INDICATIONS FOR USE 510(k) Number (if known): K112108 Device Name: ROMEO posterior osteosynthesis system Indications for Use: ROMEO posterior osteosynthesis system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; spinal tumor; and failed previous fusion (pseudarthrosis). When used as a posterior, non-cervical, non-pedicle screw fixation system, ROMEO posterior osteosynthesis system is intended for the following indications for the following indications; degenerative disc disease(DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. Over-The-Counter Use _ Prescription Use > AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (OD (Divisio Division of Surgical, Orthopedic, and Restorative Devices KII2108 510(k) Number __ Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...