THEKEN ATOLL OCT SPINAL SYSTEM

K083863 · Theken Spine, LLC · MNI · Mar 27, 2009 · Orthopedic

Device Facts

Record IDK083863
Device NameTHEKEN ATOLL OCT SPINAL SYSTEM
ApplicantTheken Spine, LLC
Product CodeMNI · Orthopedic
Decision DateMar 27, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Atoll OCT Spinal System is indicated to promote fusion of the occipito-cervicothoracic regions of the spine (Occiput - T3). The intended indications are as follows: - Degenerative Disc Disease (as identified by neck or back pain of discogenic origin a with degeneration of the disc confirmed by history and radiographic studies) - Spondylolisthesis . - Spinal Stenosis . - Fracture/Dislocation - . Tumors - Pseudoarthrosis . - Atlanto/axial fracture with instability - Occipitocervical dislocation . - Revision of previous cervical and upper thoracic spine surgery . The occipital bone screws are limited to occipital fixation only. The use of the polyaxial and standard screws is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. Screws are not intended for use in the cervical spine. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine. The Atoll OCT Spinal System can also be linked to the Theken Coral Spinal System with the use of transitional rods and rod connectors

Device Story

Posterior cervical/thoracic spinal fixation system; components include screws, hooks, rods, occipital plates, and connectors. Used by surgeons to stabilize spine segments to promote fusion. Components implanted during surgery; provide mechanical support for spinal conditions including fractures, tumors, and degenerative disease. System allows for configuration variety to match patient physiology; can link to Theken Coral Spinal System via transitional rods. Benefits include stabilization of occipito-cervico-thoracic regions.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Constructed from medical implant grade titanium alloy (Ti-6Al-4V) per ASTM F136 and ISO 5832-3. System consists of screws, hooks, rods, occipital plates, and connectors. Mechanical fixation device; no software or energy source.

Indications for Use

Indicated for patients requiring fusion of occipito-cervico-thoracic spine (Occiput-T3) due to degenerative disc disease, spondylolisthesis, spinal stenosis, fracture/dislocation, tumors, pseudoarthrosis, atlanto/axial fracture with instability, occipitocervical dislocation, or revision surgery. Contraindications: Occipital bone screws limited to occipital fixation; polyaxial/standard screws limited to upper thoracic spine (T1-T3) and not for cervical spine use.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows a logo with the word "Theken" in a stylized font. The "T" is large and angular, with a curved top extending over the rest of the word. The remaining letters are in a bold, sans-serif font. Below the word is a textured, rectangular shape that appears to be a stylized line or border. A. Submitter Information # 510(k) Summary (21 CFR Part 807.92) MAR 2 7 2009 #### Submitter's Name: Theken Spine, LLC Establishment Registration #: 1530901 Address: 1800 Triplett Blvd. Akron, Ohio 44306 Telephone Number: 330-475-8600 Fax Number: 330-773-7697 Contact Person: Dale Davison Date Prepared: 12/23/08 B. Device Information AtollTM OCT Spinal System Trade Name: Common Name: Posterior Cervical Instrumentation Classification: Class II System with the corresponding product codes: KWP 888.3050 - Spinal Interlaminal Fixation Orthosis MNI 888.3070(b) (1) – Pedicle Screw Spinal System MNH 888.3070(b) (1) - Pedicle Screw Spinal System Predicate Devices: Theken - Atoll Cervico-Thoracic System (K070638) Comparable Device: DePuy Spine (USA) - Mountaineer OCT Spinal System (K080828) Synthes Spine (USA) OC Fusion System (K053418) Material Composition: Implant Grade Titanium Alloy (Ti-6Al-4V) per ASTM F136 and ISO 5832-3 Subject Device Description: The Atoll OCT Spinal System is intended for use as an aid in spine fusion. It consists of screws, hooks, rods, occipital plates and connectors. These components are available in a variety of sizes to allow for a variety of configurations to better fit each individual patient physiology. The Atoll OCT Spinal System components are manufactured from medical implant grade titanium alloy Ti-6Al-4V (ELI) per ASTM F136 and ISO 5832-3. To achieve the best results. unless otherwise specifically described in another Theken Spine document, do not use the Atoll OCT Spinal System components in conjunction with components from any other manufacturer. The purpose of this submission is the addition of an occipital plate. Page 1 of 2 {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the word "Theken" in a stylized font. The "T" is large and angular, with a curved top. The rest of the letters are smaller and connected. Below the word is a line of small, outlined rectangles. Intended Use: The Atoll OCT Spinal System is indicated to promote fusion of the occipito-cervicothoracic regions of the spine (Occiput - T3). The intended indications are as follows: - Degenerative Disc Disease (as identified by neck or back pain of discogenic origin a with degeneration of the disc confirmed by history and radiographic studies) - Spondylolisthesis . - Spinal Stenosis . - Fracture/Dislocation � - . Tumors - Pseudoarthrosis . - Atlanto/axial fracture with instability � - Occipitocervical dislocation . - Revision of previous cervical and upper thoracic spine surgery . The occipital bone screws are limited to occipital fixation only. The use of the polyaxial and standard screws is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. Screws are not intended for use in the cervical spine. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine. The Atoll OCT Spinal System can also be linked to the Theken Coral Spinal System with the use of transitional rods and rod connectors ### C. Substantial Equivalence The characteristics of the Atoll OCT Spinal System are similar to the following predicate devices: - Atoll Cervico-Thoracic System (K070638) manufactured by Theken Spine. 1. - Mountaineer OCT Spinal System (K080828) manufactured by DePuy Spine. 2. - OC Fusion System (K053418) manufactured by Synthes Spine. প Equivalence for the Atoll OCT Spinal System is based on similarities of intended use, design, and physical characteristics when compared to the predicate devices. Therefore, Theken Spine believes that there is sufficient evidence to conclude that the Atoll OCT Spinal System is substantially equivalent to existing legally marketed devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services, USA. The seal features an eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are written in a circle around the eagle. The eagle is facing to the right and has three lines above it that represent feathers. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Theken Spine LLC % Mr. Dale Davison 1800 Triplett Blvd Akron, Ohio 44306 MAR 2 7 2009 Re: K083863 Trade/Device Name: Theken Atoll OCT Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: II Product Code: MNH, MNI, KWP Dated: December 23, 2008 Received: December 29, 2008 Dear Mr. Davison: We have reviewed your Section 510(k) premarket notification of intent to market the device indication We nave reviewed your Section 510(x) premainted is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreative to regally manated povice Americal Device Ameral Food Drug commerce prof to May 20, 1976, the enaonable with the provisions of the Federal Food, Drug, devices that have been recuire approval of a premarket approval application (PMA). and Cosmetic Act (Act) that do not require approval of a premarket approval application (PM and Cosment Act (Act) that do not require approval controls provisions of the Act. The Act. The Act. The Act. The You may, merelore, market the device, baoject to the ments for annual registration, listing of the and general controls provisions of the Fect menate required in the manage inst misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your device is classified (see above) into e. Existins major regulations affecting your device. FDA It may be subject to such additional versions, Title 21, Parts 800 to 898. In addition, FDA can be found in the Code of I ceeras concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised fill I DA s issualled of a substance of the other requirements of the Act that FDA has made a determination that your device complies with other water that FDA has made a decornmanon that your correst by other Federal agencies. You must or any Federal statutes and regulations and intel to: registration and listing (21 Comply with an the Act 3 requirements, news 801); good manufacturing practice requirements as set a CFR Part 807), labeling (21 CFRT art 820); good if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ # Page 2 – Mr. Dale Davison This letter will allow you to begin marketing your device as described in your Section 510(k) I ins letter will anow you to begin maineg of substantial equivalence of your device to a legally promatics notification: "The PDs in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please IT you ucate spocitie advices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to (240) 270-0120. This, prease now as cogning regarding regarding postmarket surveillance, premarket nontact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket prease contact CDACT 3 Office of Day westions regarding the reporting of device adverse events Surveillance at (240) 270 317 17 17 DVD)), please contact the Division of Surveillance Systems (Medical Device Reporting (11271)), prec.org. information on your responsibilities under the at (240) 270 370 1. 1 0 a may acturers, International and Consumer Assistance At tolli the Dribion of (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrl/industry/support/index.html. Sincerely yours, erely yours, For Pete D Rumin mo Dep Dir r. N. Melkerson Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): ## Device Name: Atoll™ OCT Spinal System Indications For Use: The Atoll OCT Spinal System is indicated to promote fusion of the occipito-cervico-thoracic regions of the spine (Occiput - T3). The intended indications are as follows: - Degenerative Disc Disease (as identified by neck or back pain of discogenic origin with . degeneration of the disc confirmed by history and radiographic studies) - . Spondylolisthesis - Spinal Stenosis . - Fracture/Dislocation . - Tumors . - Pseudoarthrosis . - Atlanto/axial fracture with instability . - Occipitocervical dislocation . - Revision of previous cervical and upper thoracic spine surgery The occipital bone screws are limited to occipital fixation only. The use of the polyaxial and standard screws is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. Screws are not intended for use in the cervical spine. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine. The Atoll OCT Spinal System can also be linked to the Theken Coral Spinal System with the use of transitional rods and rod connectors. Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation (ODE) Radtl Division Sign-Off Division of Ceneral, Restorative, and Nearclogical Devices Page 1 of 1 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...