XIA SPINAL SYSTEM
K022160 · Howmedica Osteonics · MNI · Aug 1, 2002 · Orthopedic
Device Facts
| Record ID | K022160 |
| Device Name | XIA SPINAL SYSTEM |
| Applicant | Howmedica Osteonics |
| Product Code | MNI · Orthopedic |
| Decision Date | Aug 1, 2002 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 888.3070 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Xia Spinal System, which includes the subject Xia Titanium Hooks, and the Xia Stainless Steel System are intended for use in the noncervical spine. When used as a pedicle screw fixation system, the Xia Spine System and Xia Stainless Steel Systems are intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the fifth lumbar - first sacral (L5-S1) vertebral joint: (b) who are receiving fusions using autogenous bone graft only: (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass. When used as a pedicle screw fixation system, the Xia Spinal System and Xia Stainless Steel Systems are also intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). When used as an anterior screw fixation system or a posterior hook and sacral/iliac screw fixation system, the Xia Spinal System and Xia Stainless Steel Systems are indicated for patients with degenerative disc disease which is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, spinal deformities such as scoliosis, kyphosis, tumor, pseudoarthrosis or revision of failed fusion attempts.
Device Story
Xia Titanium Hooks are line extension components for Xia Spinal System; intended for noncervical spinal fixation. Device consists of titanium alloy hooks; modification from predicate stainless steel hooks involves material change and dimensional adjustments. System provides immobilization/stabilization of spinal segments as adjunct to fusion. Used by surgeons in clinical settings; components attached to thoracic, lumbar, or sacral spine. Output is mechanical stabilization of vertebral segments; facilitates bone fusion. Benefits include correction of spinal deformities and stabilization of unstable segments.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Components fabricated from titanium alloy. System includes hooks, screws, rods, and connectors. Designed for noncervical spinal fixation. Mechanical properties verified via bench testing to be comparable to stainless steel predicate.
Indications for Use
Indicated for skeletally mature patients requiring noncervical spinal stabilization/immobilization. Conditions include severe spondylolisthesis (L5-S1), degenerative disc disease, fracture, dislocation, scoliosis, kyphosis, spinal tumor, spinal stenosis, and failed previous fusion (pseudarthrosis).
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Predicate Devices
- Xia Stainless Steel System (K022160)
Reference Devices
- Osteonics Spinal System
- Opus Spinal System
Related Devices
- K052181 — MODIFICATION TO XIA SPINAL SYSTEM · Stryker Spine · Sep 8, 2005
- K063428 — XIA STAINLESS STEEL SPINAL SYSTEM · Stryker Spine · Dec 11, 2006
- K061854 — XIA TITANIUM SPINAL SYSTEM AND XIA STAINLESS STEEL SPINAL SYSTEM · Stryker Spine · Jul 27, 2006
- K132471 — CD HORIZON SPINAL SYSTEM · Medtronic Sofamor Danek USA, Inc. · Oct 8, 2013
- K053115 — XIA SPINAL SYSTEM · Stryker Spine · Dec 6, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
Special 510(k) Premarket Notification
# AUG 0 1 2002
# Special 510(k) Summary: Line Extension to the Xia Spinal System
#### Submission Information
| Name and Address of the Sponsor<br>of the 510(k) Submission: | Howmedica Osteonics Corp<br>59 Route 17<br>Allendale, NJ 07401-1677 |
|--------------------------------------------------------------|---------------------------------------------------------------------|
| Contact Person: | Karen Ariemma<br>Regulatory Affairs Specialist |
| Date of Summary Preparation: | June 28, 2002 |
| Device Identification | |
| Proprietary Name: | Xia Spinal System |
| Common Name: | Spinal Fixation Appliances |
| Proprietary Name: | Xia Spinal System |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Spinal Fixation Appliances |
| Classification Name and Reference: | Spinal Interlaminal Fixation Orthosis,<br>21 CFR 888.3050<br>Spinal Intervertebral Body Fixation Orthosis<br>21 CFR 888.3060<br>Pedicle Screw Spinal System<br>21 CFR 888.3070 |
### Predicate Device Identification
The Xia Stainless Steel System consists of Monoaxial and Polyaxial Screws, Washer, Hooks, Blocker, Rods, Connectors and Multi-Axial Cross Connectors (MACs). The Xia Spinal System consists of Monoaxial and Polyaxial Screws, Hooks, Blockers, Rods, Washer and Connectors.
## Description of Device Modification
This submission is intended to address a material modification to the Xia Stainless Steel Hooks. The subject components, named the Xia Titanium Hooks, are a line extension of the Xia Spinal System. The predicate Xia Stainless Steel Hooks are fabricated from stainless steel. The subject Xia Titanium Hooks are fabricated from titanium alloy. In addition, dimensional changes were made to the Xia Stainless Steel Hooks to create the subject Xia Titanium Hooks.
## Intended Use:
The Xia Spinal System, which includes the subject Xia Titanium Hooks, and the Xia Stainless Steel System are intended for use in the noncervical spine. When used as a pedicle screw fixation system, the Xia Spine System and Xia Stainless Steel Systems are intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the fifth lumbar - first sacral (L5-S1) vertebral joint: (b) who are receiving fusions using autogenous bone graft only: (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass.
When used as a pedicle screw fixation system, the Xia Spinal System and Xia Stainless Steel Systems are also intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
{1}------------------------------------------------
When used as an anterior screw fixation system or a posterior hook and sacral/iliac screw fixation system, the Xia Spinal System and Xia Stainless Steel Systems are indicated for patients with degenerative disc disease which is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, spinal deformities such as scoliosis, kyphosis, tumor, pseudoarthrosis or revision of failed fusion attempts.
### Statement of Technological Comparison:
The Xia Titanium Hooks share the same intended use and basic design concepts as that of the predicate Xia Stainless Steel Hooks. Mechanical testing demonstrated comparable mechanical properties to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, with three human profiles incorporated into the design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Re:
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Karen Ariemma Regulatory Affairs Specialist Howmedica Osteonics Corp 59 Route 17 Allendale, New Jersey 07401-1677
K022160 Xia Titanium Hooks Trade Name: 21 CFR 888.3070, 21 CFR 888.3060, 21 CFR 888.3050 Regulation Number: Pedicle Screw Spinal System, Spinal Intervertebral Body Fixation Regulation Name: Orthosis, Spinal Interlaminal Fixation Orthosis Regulatory Class: Class II MNI, MNH, KWP, KWQ Product Code: June 28, 2002 Dated: July 3, 2002 Received:
AUG 01 2002
Dear Ms. Ariemma:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 – Ms. Karen Ariemma
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html.
Sincerely yours,
Sincerely yours,
Mark M Millman
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): ≤ 022160
Device Name: Xia Spinal System
The 6 mm diameter rods from the Osteonics® Spinal System and Opus Spinal System are intended to be used with the other components of the Xia Spinal System.
Indications For Use:
The XIA Spine System is intended for use in the noncervical spine. When used as a pedicle screw fixation system, the XIA Spine System is intended for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass.
When used as a pedicle screw fixation system, the XIA Spine System is also intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
When used an anterior screw fixation system or a posterior hook and sacral/iliac screw fixation system, the XIA Spine System is indicated for patients with degenerative disc disease which is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal deformities such as scoliosis, kyphosis, lordosis, turnor, pseudoarthrosis or revision of failed fusion attempts.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use OR
Over-The-Counter Use (Per 21 CFR 801.109)
(Optional Format 1-2,96)
Mark Mel
(Division Sign-Off) Division of General, Restorative and Neurological Device
510(k) Number