MICOMED POSTERIOR DOUBLEROD SYSTEM

K021275 · Micomed Ortho GmbH · MNI · Jul 18, 2002 · Orthopedic

Device Facts

Record IDK021275
Device NameMICOMED POSTERIOR DOUBLEROD SYSTEM
ApplicantMicomed Ortho GmbH
Product CodeMNI · Orthopedic
Decision DateJul 18, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Micomed Posterior Doublerod System is a pedicle screw system indicated for the treatment of severe spondyloisthesis (Grades 3 and 4) at the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. The Micomed Posterior Doublerod System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities of deformities of the thoracic, lumbar, and sacral spine: degenerative spondyloisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudoarthrosis). When used as a non-pedicle screw fixation system, the Micomed Posterior Doublerod System is also intended for scoliotic, lordotic deformities such as scoliosis, Scheuermann's disease, degernerative disk disease defined as back pain of discogenic origin with degeneration of the disk confirmed by patient history and radiographic studies, and fractures of the posterior thoracolumbar spine from levels T4 to S1.

Device Story

Low-profile, top-loading spinal fixation system; components include pedicle screws, polyaxial screws, open/closed hooks, and fluted/threaded rods. Used by surgeons in clinical settings to provide immobilization and stabilization of spinal segments as an adjunct to fusion. Titanium and stainless steel construction. Polyaxial screws available only in titanium. System provides mechanical support to the spine; aids in fusion process; intended for removal after solid fusion attainment in specific spondylolisthesis cases.

Clinical Evidence

Bench testing only. Biomechanical performance of the polyaxial screw was evaluated per ASTM F1717-96 and shown to be equivalent to predicate devices.

Technological Characteristics

Materials: Titanium (ASTM F136) and Stainless Steel (ASTM F1314). Components: Pedicle screws, polyaxial screws (titanium only), hooks, fluted/threaded rods. Mechanical fixation system. No software or energy source.

Indications for Use

Indicated for skeletally mature patients requiring spinal fusion for severe spondylolisthesis (Grades 3-4, L5-S1), acute/chronic instabilities or deformities (thoracic, lumbar, sacral spine) including degenerative spondylolisthesis with neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, failed fusion, Scheuermann's disease, and degenerative disc disease.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) Notification #K021275 01/2 Micomed Posterior Doublerod System (MPDS) July 9, 2002 JUL 1 8 2002 ## Attachment 2 ## Micomed -- Posterior Doublerod System 510(K) Summary ## Prepared July 1, 2002 Micomed Posterior Doublerod System Tradename: Spinal Fixation Device System Generic Name Classification: Class II MNI and MNH, 21CFR888.3070 KWP, 21CFR888.3050 Micomed Ortho GmbH & Co. KG Company: Schorndorfer Strasse 96 Schorndorf 73614 Germany Phone: 011 49 71 81 48 99 85 87 Fax: 011 49 71 81 48 99 84 info@micomed.com Corrine M. Taflinger, RAC Contact: RA Consultant (858) 481-1638 Predicate Cross Medical SynergyTM Spinal System, 510(k) # K974749 Depuy Inc. Moss® Spinal System, 510(k) #984378 Devices The Micomed Posterior Doublerod System is a low profile, top loading Description: spinal fixation system available in titanium and stainless steel. The system consists of pedicle screws, polyaxial screws*, open and closed hooks, and fluted and threaded rods. A set of instruments is available for use with the Micomed Posterior Doublerod System. * polyaxial screws are only available in titanium The components of the Micomed Posterior Doublerod System are available Material: in titanium in conformance with ASTM F136 as well as Stainless Steel in conformance with ASTM F1314. Polyaxial Screws are only available in titanium. Page /D {1}------------------------------------------------ 510(k) Notification #K021275 842 Micomed Posterior Doublerod System (MPDS) r July 9, 2002 - - ۰۰ | | 510(K) Summary - continued | |-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications: | The Micomed Posterior Doublerod System is a pedicle screw system<br>indicated for the treatment of severe spondyloisthesis (Grades 3 and 4) at<br>the L5-S1 vertebra in skeletally mature patients receiving fusion by<br>autologous bone graft having implants after the attainment of a solid fusion.<br>The Micomed Posterior Doublerod System is a pedicle screw system<br>intended to provide immobilization and stabilization of spinal segments in<br>skeletally mature patients as an adjunct to fusion in the treatment of the<br>following acute and chronic instabilities of deformities of the thoracic.<br>lumbar, and sacral spine: degenerative spondyloisthesis with objective<br>evidence of neurological impairment, fracture, dislocation, scoliosis,<br>kyphosis, spinal tumor and failed previous fusion (pseudoarthrosis).<br>When used as a non-pedicle screw fixation system, the Micomed Posterior<br>Doublerod System is also intended for scoliotic, lordotic, or kyphotic | | | deformities such as scoliosis, Scheuermann's disease, degernerative disk<br>disease defined as back pain of discogenic origin with degeneration of the<br>disk confirmed by patient history and radiographic studies, and fractures of<br>the posterior thoracolumbar spine from levels T4 to S1. | | Testing and<br>Performance: | The polyaxial screw, to be used with the Micomed Posterior Doublerod<br>System. has been shown to have acceptable biomechanical performance per<br>ASTM F 1717-96 and to function in an equivalent manner to the polyaxial<br>screws currently used in the predicate device systems. | | Substantial<br>Equivalence: | The Micomed polyaxial screw for use with the Posterior Doublerod System<br>is substantially equivalent with respect to intended use, design, and<br>performance to the Cross Medical polyaxial screw (as part of the Synergy<br>Spinal System) cleared for market entry under 510(k) # K974749 on March<br>13, 1998, and the Depuy Moss® Spinal System cleared for market entry<br>under 510(k) #984378 on December 30, 1998. | : 上一篇: : : ・ 000 2017 {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES #### Public Health Service Food and Drug Administration · 9200 Corporate Boulevard Rockville MD 20850 # JUL 1 8 2002 Micomed Ortho GmbH & Co. KG c/o Corrine M. Taflinger, RAC Regulatory Affairs Consultant 13195 Seagrove Street Lake Forest, California 92130 Re: K021275 Trade/Device Name: Micomed Posterior Doublerod System Regulatory Number: 21 CFR 888.3070. 21 CFR 888.3050 Regulation Name: Pedicle screw spinal system, Spinal interlaminal fixation orthosis Regulatory Class: II Product Code: MNH, MNI, KWP Dated: April 17, 2002 Received: April 22, 2002 Dear Ms. Taflinger: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Corrine M. Taflinger, RAC This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Mark M Millerson Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Notification #K021275 Micomed Posterior Doublerod System (MPDS) July 9, 2002 ### Attachment 3 #### Statement of Indication for Use ### 510(k) Number: K021275 Device Name: Micomed Posterior Doublerod System Indications for Use: The Micomed Posterior Doublerod System is a pedicle screw system indicated for the treatment of severe spondyloisthesis (Grades 3 and 4) at the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. The Micomed Posterior Doublerod System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities of deformities of the thoracic, lumbar, and sacral spine: degenerative spondyloisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudoarthrosis). When used as a non-pedicle screw fixation system, the Micomed Posterior Doublerod System is also intended for scoliotic, lordotic deformities such as scoliosis, Scheuermann's disease, degernerative disk disease defined as back pain of discogenic origin with degeneration of the disk confirmed by patient history and radiographic studies, and fractures of the posterior thoracolumbar spine from levels T4 to S1. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The Counter Use Mark n Mill (Optional Format 1-2-96) (Division Sign-Off) Division of General, Restorative and Neurological Devices Micomed Ortho GmbH 510(k) Number K0212 Page / 2
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...