MITEK MICRO ANCHOR
Device Facts
| Record ID | K962511 |
|---|---|
| Device Name | MITEK MICRO ANCHOR |
| Applicant | Mitek Products |
| Product Code | MBI · Orthopedic |
| Decision Date | Nov 6, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3040 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The device is used to anchor suture into bone. The suture is subsequently used by the surgeon to reattach the repositioned/injured soft tissue to bone. The purpose of this 510(k) submission is to obtain clearance for the Mitek Micro Anchor for the collateral ligament(s) around the (PIP) joint. The intended use is the same as the predicate device, i.e., suture, when used to approximate soft tissue to bone during a period of rehabilitation.
Device Story
Bone fixation device; anchors USP #4/0 suture below bone surface. Components: 3.7mm x 1.3mm titanium anchor with Nitinol arc; delivery system includes drill bit and inserter. Used by surgeons in clinical settings to reattach injured soft tissue to bone. Provides mechanical fixation for collateral ligaments of PIP joint; facilitates tissue approximation during rehabilitation. Offers greater strength than suture alone.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Dimensions: 3.7mm length, 1.3mm diameter. Materials: Titanium 6Al 4V, Nitinol arc. Delivery system: drill bit and inserter. Mechanical fixation principle.
Indications for Use
Indicated for anchoring suture into bone to reattach soft tissue to bone, specifically for collateral ligament(s) around the proximal interphalangeal (PIP) joint.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- SUTURE
- Mitek GII Mini Anchor (K915089, K930892)
Related Devices
- K962793 — MITEK MICRO ANCHOR · Mitek Products · Oct 1, 1996
- K961094 — MITEK THREADED ANCHOR (MODIFICATION) · Mitek Products · Oct 3, 1996
- K022207 — AXYA MODEL 3000 AXYALOOP TITANIUM BONE ANCHOR SYSTEM · Axya Medical, Inc. · Oct 4, 2002
- K982420 — MITEK MICRO ANCHOR · Mitek Products · Aug 24, 1998
- K041115 — GII QUICKANCHOR PLUS · Depuy Mitek, A Johnson & Johnson Company · May 28, 2004