Suture Wing

K202663 · Osteonic Co., Ltd. · MBI · Jun 7, 2021 · Orthopedic

Device Facts

Record IDK202663
Device NameSuture Wing
ApplicantOsteonic Co., Ltd.
Product CodeMBI · Orthopedic
Decision DateJun 7, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

Suture wing is intended for fixation of soft tissue to bone, using suture, in orthopedic surgery.

Device Story

Suture Wing is a sterile, single-use bone anchor system used in orthopedic surgery to reattach soft tissues like ligaments, tendons, and capsules to bone. The device consists of a non-absorbable all-suture anchor and a Nitinol driver shaft. It is operated by a surgeon in a clinical setting. The surgeon uses the driver shaft to insert the anchor into the bone, providing a mechanical fixation point for the sutures to secure the soft tissue. This fixation facilitates healing by holding the tissue in the desired anatomical position. The device is supplied sterile via ethylene oxide (EtO) gas.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included mechanical performance (ASTM F543, ASTM F1839), material characterization (ASTM F2848, ASTM F2063), sterilization validation (ISO 11135, ISO 11138, ISO 11140, ISO 11737), packaging integrity (ISO 11607, ASTM F1980, ASTM F88, ASTM F1929), and bacterial endotoxin/pyrogenicity testing (USP <85>, <161>, <151>, ISO 10993-11).

Technological Characteristics

Materials: UHMWPE (ASTM F2848) and Nitinol (ASTM F2063). Components: All-suture anchor, Nitinol driver shaft, handle. Sizes: 1.7mm and 2.8mm. Sterilization: Ethylene oxide (EtO). Single-use. Mechanical fixation principle.

Indications for Use

Indicated for fixation of soft tissue (ligaments, tendons, capsules) to bone in orthopedic surgery.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ June 7, 2021 Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white. To the right of the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue. Osteonic Co., Ltd. % Sanglok Lee Manager Wise Company Inc. #303, 142, Gasan digital 1-ro Geumcheon-gu, Seoul 08507 Korea, Republic of Re: K202663 Trade/Device Name: Suture Wing Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: May 4, 2021 Received: May 6, 2021 Dear Sanglok Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Pooja Panigrahi -S for Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K202663 Device Name Suture Wing Indications for Use (Describe) Suture wing is intended for fixation of soft tissue to bone, using suture, in orthopedic surgery. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Osteonic. The logo consists of a blue geometric shape on the left and the word "OSTEONIC" in bold, black letters on the right. Below the word "OSTEONIC" is the tagline "The Medical Explorer" in a smaller, lighter font. # The assigned 510(k) Number: K202663 ## 01. Date of Submission: 2020.09.11 # 02. Applicant Dakyung Ham OSTEONIC Co., Ltd. 303Ho, 405Ho, 505-2Ho, 505-3Ho, 902Ho, 1004Ho, 1201Ho, 1202Ho, 1206Ho, 38 Digital-ro 29-gil, Guro-gu, Seoul, Korea, 102Ho, 103Ho, 24, Digital-ro 27-gil, Guro-gu, Seoul, Korea TEL: +82-2-6902-8456 FAX: +82-2-6902-8401 Email: dakham@osteonic.com #### 03. Submission Correspondent Sanglok, Lee Wise COMPANY Inc. #303, 142, Gasan digital 1-ro, Geumcheon-gu, Seoul, Korea TEL: +82 70 8812 3619 / +82 2 831 3615 FAX: +82 50 4031 3619 Email: info@wisecompany.org ## 04. Subject Device Identification Trade Name: Suture Wing Common Name: Sterile bone anchor-all suture Classification Name: fastener, fixation, non-biodegradable, soft tissue Product Code: MBI Panel: Orthopedic Regulation Number: 21 CFR 888.3040 Device Class: II #### 05. Indication for use Suture wing is intended for fixation of soft tissue to bone, using suture, in orthopedic surgery. #### 06. Predicate devices Predicate device 510(k) Number: K133224 Device Name: Y-Knot RC All-suture anchor w/Two and Three #2 Hi-Fi Sutures Manufacturer: ConMed Corporation # 07. Device Description This product is used for orthopedic surgery, which soft tissue fix such as ligaments, tendons, and capsules to bone. The anchor and sutures is made of non-absorbable and consists of Driver Shaft(Nitinol) for anchor insertion. This product is sterilized product and single use only. Sterile bone anchor-all suture is supplied EO gas sterile state and it was packed in tyvek and PE film. #### 08. Non-Clinical Test Conclusion Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards: {4}------------------------------------------------ Image /page/4/Picture/1 description: The image contains a logo for Osteonic. The logo consists of a blue geometric shape on the left and the word "OSTEONIC" in bold, dark gray letters on the right. Below "OSTEONIC" is the tagline "The Medical Explorer" in a smaller, lighter gray font. # ▪ Material - ASTM F2848: 2017 Standard Specification for Medical-Grade Ultra-High Molecular Weight Polyethylene Yarns - ASTM F2063: 2012 Standard Specification for Wrought Nickel-Titanium Shape Memory Alloys for Medical Devices and Surgical Implants # ▪ Mechanical performance - ASTM F543: 2013 Standard specification and test methods for metallic medical bone screws - ASTM F1839: 2008 (Reapproved 2016), Standard specification for rigid polywethane foam for use as a standard material for testing orthopaedic devices and instruments # ▪ Sterilization, shelf-life and packaging for sterile product - ISO 11135:2014, Sterilization of health-care products Ethylene oxide Requirements for the development validation and routine control of a sterilization and routine control of a sterilization process for medical devices - ISO 11138-1:2006, Sterilization of health care products Biological indicators Part 1: General requirements - ISO 11138-2:2009, Sterilization of health care products - Biological indicators - Part 2: Biological indicators for ethylene oxide sterilization processes - ISO 11140-1:2014, Sterilization of health care products Chemical indicators Part 1: General requirements - ISO 11737-1:2018 Sterilization of medical devices Microbiological methods- Part 1: Estimation of microorganisms on products - ISO 11737-2:2009 Sterilization of medical devices Microbiological methods- Part 2. Tests of sterility performed in the validation of a sterilization process - ISO 11607-1:2006/AMD1:2014 Packaging for terminally sterilized medical devices part 1: requirements for materials, sterile barrier systems and packaging system - ISO 11607-2:2006/AMD1:2014 Packaging for terminally sterilized medical devices Part 2: Validation requirements for forming, sealing and assembly processes - ASTM F1980:2016 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices - ASTM F88/F88M:2015 Standard test method for seal strength of flexible barrier materials. - ASTM F1929:2015 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration # ▪ Bacterial Endotoxin - USP <85> Bacterial Endotoxin Test - USP <161> Medical Devices-Bacterial Endotoxin and Pyrogen Tests - Material mediated pyrogenicity - USP <151> Pyrogenicity Test - ISO 10993-11: Tests for systemic toxicity # ▪ Insertion test - FDA guidance : Bone Anchors Premarket notification (510(K)) Submissions - ASTM F1839: 2008 (Reapproved 2016), Standard specification for rigid polyurethane foam for use as a standard material for testing orthopaedic devices and instruments # 09. Substantially Equivalent Conclusion # Table 1: Substantial Equivalence Comparison | Product<br>Name | SUBJECT Device | PREDICATE Device<br>Y-Knot RC All-suture anchor w/Two and<br>Three #2 Hi-Fi Sutures (K133224) | Equivalence<br>Discussion | |----------------------|-----------------|-----------------------------------------------------------------------------------------------|---------------------------| | Product<br>code | MBI | MBI | Same | | Regulatory<br>class | Class II | Class II | Same | | Regulation<br>Number | 21 CFR 888.3040 | 21 CFR 888.3040 | Same | {5}------------------------------------------------ Image /page/5/Picture/1 description: The image contains a logo for a company called OSTEONIC. The logo consists of a blue geometric shape resembling a stylized diamond or cube, positioned to the left of the company name. Below the company name, there is a tagline that reads "The Medical Explorer" in a smaller font size. | Intended use | Suture wing is intended for fixation of soft<br>tissue to bone, using suture, in orthopedic<br>surgery. | The non-absorbable suture anchors are<br>intended to reattach soft tissue to bone in<br>orthopedic surgical procedures. | Same | |-------------------------|-----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------| | Operating<br>Principles | Bone fixation anchor that ties soft tissues<br>such as ligament, tendon, and the articular<br>capsules to bone. | Bone fixation anchor that ties soft tissues<br>such as ligament, tendon, and the articular<br>capsules to bone. | Same | | Material | Anchor and suture:<br>UHMWPE | Anchor and suture:<br>UHMWPE | Similar | | Structure | This product consists of a non-absorbable all<br>suture and driver shaft and handle. | This product consists of a non-absorbable all<br>suture and driver shaft and handle. | Similar | | Product<br>Size | 1.7mm/ 2.8mm | 1.8mm/ 2.8mm | Similar | | Sterilization | Sterile (EtO sterilization) | Sterile (EtO sterilization) | Same | | Single Use/<br>Reuse | Single use | Single use | Same | | Packaging | 1 EA / BOX | 1 EA / BOX | Same | | Shelf -life | 5 Years | 5 Years | Same | Based on above, the subject device, Suture Wing, is Substantially Equivalent (SE) to the predicate devices, Y-Knot RC All-suture anchor w/Two and Three #2 Hi-Fi Sutures (K133224) in respect of safety and effectiveness.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...