The K2M Growing Spine System implants are indicated in patients with potential for additional spinal growth under 10 years of age who require surgical treatment to obtain and maintain correction of severe, progressive, life threatening, early-onset spinal deformities associated with thoracic insufficiency, including early-onset scoliosis. The Growth Guidance implants may be used with any cleared K2M rod construct ranging in diameter 4.5mm to 6.35mm. The Growth Rod Conversion implants may be used with 4.5mm and 5.5mm rod constructs. The K2M Growing Spine System is not intended to be used in conjunction with staples.
Device Story
System consists of spinal screw and connector components designed for use in growing rod constructs. Implants accommodate spinal growth in pediatric patients under 10 years of age. Used by surgeons in clinical settings to obtain and maintain correction of severe, progressive, life-threatening, early-onset spinal deformities, including early-onset scoliosis. Growth Guidance implants compatible with K2M rod constructs (4.5mm–6.35mm); Growth Rod Conversion implants compatible with 4.5mm and 5.5mm rod constructs. Device provides mechanical stabilization of the spine while allowing for growth, potentially reducing deformity progression and improving thoracic function.
Clinical Evidence
Bench testing only. Performance evaluations conducted on worst-case constructs for static torsion, static compression, and dynamic compression bending per ASTM F1717. Bacterial endotoxin testing performed.
Technological Characteristics
System comprises screw and connector components for growing rod constructs. Materials, design features, and sizes are substantially equivalent to predicate systems. Testing performed in accordance with ASTM F1717 for mechanical performance.
Indications for Use
Indicated for patients under 10 years of age with potential for spinal growth requiring surgical correction of severe, progressive, life-threatening, early-onset spinal deformities associated with thoracic insufficiency, including early-onset scoliosis. Not for use with staples.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Predicate Devices
Medtronic CD HORIZON® Growth Rod Conversion Set (K133904)
K140750 — SHILLA GROWTH GUIDANCE SYSTEM · Medtronic Sofamor Danek USA, Inc. · Jul 17, 2014
K142114 — Xia® Growth Rod Conversion Set · Stryker Corporation · Oct 27, 2014
K180227 — Polaris Spinal Growth System · Zimmer Biomet Spine, Inc. · Mar 15, 2018
K241816 — VerteGlide Spinal Growth Guidance System · OrthoPediatrics Corp. · Mar 14, 2025
K172979 — NuVasive® Growth Rod Conversion Set · Nu Vasive, Incorporated · Nov 6, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency name in a lighter blue font. The logo is simple and professional, conveying the agency's authority and mission.
July 29, 2022
K2M, Incorporated Ms. Nancy Giezen Manager, Regulatory Affairs 600 Hope Parkway Southeast Leesburg, Virginia 20175
Re: K161028
Trade/Device Name: K2M Growing Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: PGM
Dear Ms. Giezen:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated July 22, 2016. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under regulation number, 21 CFR 888.3070.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Ronald Jean, OHT6: Office of Orthopedic Devices, (301)796-5650, Ronald.Jean@fda.hhs.gov
Sincerely,
Ronald P. Jean -S
Ronald P. Jean, Ph.D. Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 22, 2016
K2M, Incorporated Ms. Nancv Giezen Manager, Regulatory Affairs 600 Hope Parkway Southeast Leesburg, Virginia 20175
Re: K161028 Trade/Device Name: K2M Growing Spine System Regulatory Class: Unclassified Product Code: PGM Dated: June 30, 2016 Received: July 1, 2016
Dear Ms. Giezen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161028
Device Name K2M Growing Spine System
#### Indications for Use (Describe)
The K2M Growing Spine System implants are indicated in patients with potential for additional spinal growth under 10 years of age who require surgical treatment to obtain and maintain correction of severe, progressive, life threatening, early-onset spinal deformities associated with thoracic insufficiency, including early-onset scoliosis. The Growth Guidance implants may be used with any cleared K2M rod construct ranging in diameter 4.5mm to 6.35mm. The Growth Rod Conversion implants may be used with 4.5mm and 5.5mm rod constructs. The K2M Growing Spine System is not intended to be used in conjunction with staples.
Type of Use (Select one or both, as applicable):
| <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) SUMMARY K2M, Inc. K2M Growing Spine System
#### Submitter
K2M, Inc. 600 Hope Parkway SE Leesburg, VA 20175 Contact Person: Nancy Giezen Telephone: 571 919-2000 Date Prepared: 06/29/2016
## Classification
| Trade Name: | |
|-------------------|--|
| Common Name: | |
| Regulatory Class: | |
K2M Growing Spine System Growing Rod System Unclassified (PGM)
Classification Name(s): Growing Rod System (Unclassified, Product Code PGM)
## Predicate Device(s)
Primary Predicates: Medtronic CD HORIZON® Growth Rod Conversion Set (K133904)
Additional Predicates: Medtronic SHILLA Growth Guidance System (K140750) K2M Range/Denali/Mesa Spinal System (K070229, K120099, K121630, K141873) DePuy ISOLA® and EXPEDIUM® Growing Spine Systems (K141509) Stryker Xia® Growth Rod Conversion Set (K142114) Orthopedic Equipment Company (K781443, K781448, K781449)
## Device Description
The K2M Growing Spine System consists of screw and connector components. The purpose of this 510(k) is to enable these system components to be used as part of a growing rod construct.
Function: When used as part of a growing rod construct, the system implants are designed to accommodate growth in patients under 10 years of age.
## Indications For Use
The K2M Growing Spine System implants are indicated in patients with potential for additional spinal growth under 10 years of age who require surgical treatment to obtain and maintain correction of severe, progressive, life threatening, early-onset spinal deformities associated with thoracic insufficiency, including early-onset scoliosis. The Growth Guidance implants may be
{5}------------------------------------------------
used with any cleared K2M rod construct ranging in diameter 4.5mm to 6.35mm. The Growth Rod Conversion implants may be used with 4.5mm and 5.5mm rod constructs. The K2M Growing Spine System is not intended to be used in conjunction with staples.
#### Technological Comparison to Predicate(s)
The K2M Growing Spine System was compared to predicate systems and the design features, materials and sizes were found to be substantially the same as these systems.
## Non-clinical Performance Evaluation
Performance evaluations were previously conducted on constructs representing the worst case components (including static torsion, static compression and dynamic compression bending in accordance with ASTM F1717). In the current submission, engineering rationales determined that the proposed implants were substantially equivalent to the predicate devices. Bacterial endotoxin testing was also conducted in support of substantial equivalence.
#### Conclusion
There are no significant differences between the K2M Growing Spine System and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.