The Xia® Growth Rod Conversion Set is indicated in patients with potential for additional spinal growth under 10 years of age who require surgical treatment to obtain and maintain correction of severe, progressive, life-threatening, early-onset spinal deformities associated with thoracic insufficiency, including early-onset scoliosis. The Xia® Growth Rod Conversion Set may be used with any cleared Xia® 4.5 Spinal System rod construct. The Xia® Growth Rod Conversion Set is not intended for use in conjunction with staples.
Device Story
Xia® Growth Rod Conversion Set consists of titanium alloy connectors designed to convert traditional fusion spinal constructs into non-fusion growth-enabling constructs. Used in pediatric patients with early-onset spinal deformities; allows for periodic surgical lengthening as patient grows. Operates by integrating with existing Xia® 4.5 Spinal System components (hooks, screws, connectors, 4.5mm rods). Used in surgical settings by physicians; provides mechanical stabilization while permitting spinal growth. Benefits include maintenance of deformity correction in growing children without premature fusion.
Clinical Evidence
Bench testing only. Engineering analysis demonstrated that the conversion set does not adversely affect the performance of the Xia® 4.5 Spinal System.
Technological Characteristics
Titanium alloy components; designed for use with 4.5mm diameter rods, hooks, and screws. Mechanical connector system for spinal rod constructs. No software or energy source.
Indications for Use
Indicated for patients <10 years old with potential for spinal growth requiring surgical correction of severe, progressive, life-threatening, early-onset spinal deformities associated with thoracic insufficiency, including early-onset scoliosis. Contraindicated for use with staples.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Predicate Devices
Medtronic Sofamor Danek, CD HORIZON® Growth Rod Conversion Set (K133904)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 29, 2022
Stryker Spine Garry T. Hayeck, Ph.D. Senior Regulatory Affairs Specialist 2 Pearl Court Allendale, New Jersey 07401
Re: K142114
Trade/Device Name: Xia® Growth Rod Conversion Set Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: PGM
Dear Dr. Hayeck:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated October 27, 2014. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under regulation number, 21 CFR 888.3070.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Ronald Jean, OHT6: Office of Orthopedic Devices, (301)796-5650, Ronald.Jean@fda.hhs.gov
Sincerely,
Ronald P. Jean -S
Ronald P. Jean, Ph.D. Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol of medicine, with three intertwined snakes and a staff. The logo is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 27, 2014
Stryker Spine Garry T. Hayeck, Ph.D. Senior Regulatory Affairs Specialist 2 Pearl Court Allendale, New Jersey 07401
Re: K142114
Trade/Device Name: Xia® Growth Rod Conversion Set Regulatory Class: Unclassified Product Code: PGM Dated: July 31, 2014 Received: August 4, 2014
Dear Dr. Hayeck:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Ronald P. Jean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142114
Device Name Xia® Growth Rod Conversion Set
## Indications for Use (Describe)
The Xia® Growth Rod Conversion Set is indicated in patients with potential for additional spinal growth under 10 years of age who require surgical treatment to obtain and maintain correction of severe, progressive, life-threatening, earlyonset spinal deformities associated with thoracic insufficiency, including early-onset scoliosis. The Xia® Growth Rod Conversion Set may be used with any cleared Xia® 4.5 Spinal System rod construct. The Xia® Growth Rod Conversion Set is not intended for use in conjunction with staples.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) Summary: Stryker Spine Xia® Growth Rod Conversion Set | |
|--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Stryker Spine<br>2 Pearl Court<br>Allendale, NJ 07401 |
| Contact Person | Garry T. Hayeck, Ph.D.<br>Senior Regulatory Affairs Specialist<br>Phone: 201-760-8043<br>Fax: 201-760-8406<br>E-mail: garry.hayeck@stryker.com |
| Date Prepared | September 29, 2014 |
| Trade Name | Xia® Growth Rod Conversion Set |
| Common Name | Growing Rod System |
| Proposed Class | Unclassified |
| Product Code | PGM |
| Predicate and<br>Reference Devices | The Xia® Growth Rod Conversion Set was shown to be substantially<br>equivalent to the primary predicate device listed below:<br>Medtronic Sofamor Danek, CD HORIZON® Growth Rod Conversion<br>Set, K133904<br><br>The following reference device was also cited in support of this<br>submission:<br>Stryker Spine, Xia® 4.5 Spinal System, K121342 |
| Device Description | The Xia® Growth Rod Conversion Set consists of connectors designed<br>to convert a traditional fusion construct into a non-fusion growth<br>enabling construct that can be surgically lengthened on a periodic<br>basis as the patient grows.<br><br>The Xia® Growth Rod Conversion Set components are manufactured<br>from titanium alloy and are designed to interact with constructs<br>consisting of hooks, screws, connectors, and 4.5mm diameter rods.<br>The Xia® Growth Rod Conversion Set is intended for use only with Xia®<br>4.5 Spinal System fusion constructs cleared for pediatric use. |
| Indications for Use | The Xia® Growth Rod Conversion Set is indicated in patients with<br>potential for additional spinal growth under 10 years of age who<br>require surgical treatment to obtain and maintain correction of<br>severe, progressive, life-threatening, early-onset spinal deformities<br>associated with thoracic insufficiency, including early-onset scoliosis.<br>The Xia® Growth Rod Conversion Set may be used with any cleared<br>Xia® 4.5 Spinal System rod construct. The Xia® Growth Rod Conversion<br>Set is not intended for use in conjunction with staples. |
| Summary of<br>Technological<br>Characteristics | Characteristics of the Xia® Growth Rod Conversion Set are<br>substantially equivalent to those of the above mentioned CD<br>HORIZON® Growth Rod Conversion Set based on material, geometry,<br>intended use, and fundamental scientific technology, |
| Summary of the<br>Performance Data | Engineering analysis demonstrated that introduction of the Growth<br>Rod Conversion Set does not adversely affect performance of the<br>Xia® 4.5 Spinal System and does not represent a new, worst case<br>scenario. No additional performance data was provided. |
| Conclusion | This system is as safe and effective as the previously cleared CD<br>HORIZON® Growth Rod Conversion Set listed above due to their<br>shared technological and material characteristics, and principles of<br>operation. |