K232123 · Medacta International S.A. · OVD · Oct 18, 2023 · Orthopedic
Device Facts
Record ID
K232123
Device Name
MectaLIF Anterior Lag Extension
Applicant
Medacta International S.A.
Product Code
OVD · Orthopedic
Decision Date
Oct 18, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The MectaLIF Anterior is an interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The interior of the spacer component of the MectaLIF Anterior can be packed with autograft or autologous bone graft. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six months of non-operative treatment. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). The MectaLIF Anterior Stand-Alone system is intended to be used with bone screws provided and requires no additional supplementary fixation. The MectaLIF Anterior Simple requires additional supplementary fixation such as pedicle screws and rods or lumbar anterior plate system.
Device Story
MectaLIF Anterior Lag Extension is an intervertebral body fusion device system consisting of cages, plates, and screws. It is used in spinal surgery to facilitate fusion in patients with degenerative disc disease. The device is implanted by a surgeon; the spacer component is packed with autograft or autologous bone graft. The system includes a 'stand-alone' configuration using bone screws for fixation, eliminating the need for additional supplementary fixation, or a 'simple' configuration requiring supplemental fixation (e.g., pedicle screws/rods). This submission introduces a design change to the 'lag plate flush' and its associated 'anti-back-out' cover plate, which secures to the disc spacer via an interlocking mechanism to reduce the risk of screw migration. The device is used in a clinical/OR setting. It provides structural support to the intervertebral space to promote fusion, potentially alleviating back pain of discogenic origin.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by non-clinical performance testing, including simulated-use testing to confirm operational integrity and resistance testing of the flush plate-cover plate interface.
Technological Characteristics
Intervertebral body fusion device. Materials: Ti6Al4V ELI (ISO 5832-3/ASTM F136). Components: Cages, plates, and screws. Features an interlocking mechanism with an anti-back-out cover plate. Sterilization and packaging remain unchanged from the predicate.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1, including those with up to Grade I spondylolisthesis or retrolisthesis. Patients must have failed at least six months of non-operative treatment. Previous non-fusion spinal surgery at the involved level is permitted.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K212831 — MectaLIF Extension · Medacta International S.A. · Oct 29, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 18, 2023
Medacta International S.A. % Christopher Lussier Senior Director, Quality, Regulatory and Clinical Research Medacta USA 6386 Global Drive, Suite 101 Memphis. Tennessee 38141
Re: K232123
Trade/Device Name: MectaLIF Anterior Lag Extension Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVD Dated: July 17, 2023 Received: August 3, 2023
Dear Mr. Lussier:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Brent Showalter -S
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K232123
Device Name
MectaLIF Anterior Lag Extension
### Indications for Use (Describe)
The MectaLIF Anterior is an interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The interior of the spacer component of the MectaLIF Anterior can be packed with autograft or autologous bone graft. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies.
These patients should be skeletally mature and have had at least six months of non-operative treatment.
These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). The MectaLIF Anterior Stand-Alone system is intended to be used with bone screws provided and requires no additional supplementary fixation. The MectaLIF Anterior Simple requires additional supplementary fixation such as pedicle screws and rods or lumbar anterior plate system.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Senior Director, Quality, Regulatory, and Clinical Research, Medacta USA
Date Prepared: July 17, 2023
#### II. Device
| Device Proprietary Name: | MectaLIF Anterior Lag Extension |
|--------------------------|---------------------------------------------------------------|
| Common or Usual Name: | Intervertebral Fusion Device With Integrated Fixation, Lumbar |
| Classification Name: | Intervertebral body fusion device |
| Primary Product Code | OVD |
| Regulation Number: | 21 CFR 888.3080 |
| Device Classification | II |
#### Predicate Device III.
Substantial equivalence is claimed to the following primary predicate device:
- A MectaLIF Anterior Extension, K221545, Medacta International SA.
#### IV. Device Description
MectaLIF Anterior Stand Alone Fusion Device System, consisting of cages, plates and screws, does not require additional fixation. The MectaLIF Anterior cage can be coupled with four different profiles of stand-alone plates: flush, long, L5-S1 and hybrid.
The purpose of this submission is to gain the clearance for a design change performed on the MectaLIF Anterior lag plates flush and their related cover plate already cleared within K221545.
The lag plate flush is secured to the disc spacer via an interlocking mechanism and an additional antiback-out cover plate, used to reduce the risk of screw migration after the implantation.
No materials' changes have been performed, thus both the plates and the screws are made of Ti6Al4V ELI (ISO 5832-3/ASTM F136).
{4}------------------------------------------------
#### V. Indications for Use
The MectaLIF Anterior is an interbody fusion device indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The interior of the spacer component of the MectaLIF Anterior can be packed with autograft or autologous bone graft. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six months of non-operative treatment. These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). The MectaLIF Anterior Stand-Alone system is intended to be used with bone screws provided and requires no additional supplementary fixation. The MectaLIF Anterior Simple requires additional supplementary fixation such as pedicle screws and rods or lumbar anterior plate system.
#### VI. Comparison of Technological Characteristics
The subject MectaLIF Anterior Lag Extension implants are identical to the predicate devices with regards to the following characteristics:
- Indications for use;
- System's device components; ●
- General design: ●
- Material:
- Biocompatibility: ●
- . Device usage:
- . Sterility:
- Shelf-life; and
- Packaging. ●
The only differences between the subject and the predicate devices are related to the design changes implemented on the sub-components of lag plates flush and cover plate.
# Discussion
The technological differences between the subject and predicate devices do not raise new questions of safety and effectiveness as demonstrated by the risk analysis provided within the submission.
Medacta International SA has not made any change to the indications for use, manufacturing process, material, device usage, biocompatibility, sterility, shelf life and packaging of the subject devices with respect to the predicate devices.
The comparison of technological characteristics and performance data provided within the submission supports the substantial equivalence of the subject devices respect to the predicate devices.
#### VII. Performance Data
Based on the risk analysis, testing activities were conducted to written protocols. The following validations and tests support the substantial equivalence determination:
# Non-Clinical Studies
{5}------------------------------------------------
- PERFORMANCE TESTING ●
- Simulated-use testing to confirm that, during operation according to the surgical O technique, the device can achieve its intended use.
- Simulated-use testing to confirm that, in case of potential misuse generated by fastening о attempt with the Cover Plate off axis, the device can achieve its intended use.
- Resistance testing to confirm that the resistance of the Flush Plate-Cover Plate interface o is above the required strength.
- PYROGENICITY ●
- Bacterial endotoxin test (LAL test) according to European Pharmacopoeia §2.6.14 (which o is equivalent to USP chapter <85>)
- Pyrogen test according to USP chapter <151> for pyrogenicity determination o
- The subject devices are not labeled as non-pyrogenic or pyrogen free. O
- BIOCOMPATIBILITY evaluation ●
- SHELF-LIFE evaluation
# Clinical Studies:
- No clinical studies were conducted.
#### VIII. Conclusion
The information provided above supports that the subject devices are substantially equivalent to the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.