LDR SPINE ROI-A INTERBODY FUSION SYSTEM

K110327 · Ldr Spine USA · OVD · Sep 30, 2011 · Orthopedic

Device Facts

Record IDK110327
Device NameLDR SPINE ROI-A INTERBODY FUSION SYSTEM
ApplicantLdr Spine USA
Product CodeOVD · Orthopedic
Decision DateSep 30, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

When used as an intervertebral body fusion device, the ROI-A Implant System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.

Device Story

ROI-A Oblique implant is an interbody fusion device for lumbar spine; placed via anterolateral approach at ~25° offset from anterior axis. Device acts as a spacer to facilitate fusion when used with autograft and supplemental fixation. Surgeon implants device during spinal surgery. Tantalum markers provide radiographic visibility. Device performance evaluated via bench testing (static/dynamic compression, bending, expulsion, wear debris) and cadaver studies to ensure stability and prevent migration/loosening.

Clinical Evidence

No clinical data. Evidence consists of bench testing per ASTM F2077 (static shear compression, dynamic axial compression, dynamic bending), ASTM F-04.25.02.02 (expulsion), and ASTM F1877 (wear debris). Low and high cycle cadaver testing performed to evaluate clinical stability, device expulsion/pull-out, loosening, migration, and blade back-out.

Technological Characteristics

Material: PEEK OPTIMA LT1 (ASTM F2026); Tantalum markers (ASTM F560). Design: Intervertebral body fusion device for anterolateral placement. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for intervertebral body fusion of lumbar spine (L2-S1) in skeletally mature patients with degenerative disc disease (DDD) and up to Grade I spondylolisthesis, following six months of failed non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K110327 # 510(k) Summary of Safety and Effectiveness SEP 3 0 2011 LDR Spine ROI-A® Implant System | Owner's Name & Address:<br>Contact Person: | LDR Spine USA<br>4030 West Braker Lane, Suite 360<br>Austin, TX 78759<br>Phone: (512) 344-3333<br>Fax: (512) 344-3350<br><br>Maritza Elias<br>Regulatory Affairs Project Manager<br>LDR Spine USA<br>4030 West Braker Lane, Suite 360<br>Austin, TX 78759<br>Phone: (512) 344-3471<br>Fax: (512) 344-3350<br>Email: maritzaelias@ldrspine.com | |--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date 510(k) Summary Prepared: | September 29, 2011 | | Trade Name: | LDR Spine ROI-A® Implant System | | Common Name: | Intervertebral Body Fusion Device (MAX) | | Classification: | MAX 888.3080 - Intervertebral Fusion Device with Bone<br>Graft, Lumbar | | Legally Marketed Predicate Device: | LDR Spine ROI Implant System (K082262, K090507) | | Device Description | The ROI- A Oblique implant is intended for use as an interbody<br>fusion device in the lumbar spine. The device is manufactured<br>from medical grade PEEK OPTIMA® LT1 in accordance with<br>ASTM F2026 and has tantalum markers conforming to ASTM<br>F560 embedded in the implant extremities to facilitate visibility in<br>x-ray imaging. The subject device is designed for placement<br>using an anterolateral approach with an approximate offset angle<br>of 25° from the straight anterior axis. | | Indications for Use: | When used as an intervertebral body fusion device, the ROI-A<br>Implant System is indicated for intervertebral body fusion of the<br>lumbar spine, from L2 to S1, in skeletally mature patients who<br>have had six months of non-operative treatment. The device is<br>intended for use at either one or two contiguous levels for the<br>treatment of degenerative disc disease (DDD) with up to Grade I<br>spondylolisthesis. DDD is defined as back pain of discogenic<br>origin with degeneration of the disc confirmed by history and<br>radiographic studies. The device system is designed for use with<br>supplemental fixation and with autograft to facilitate fusion. | sheet 1 of 2 {1}------------------------------------------------ ### K110327 ### 510(k) Summary of Safety and Effectiveness LDR Spine ROI-A® Implant System Non-Clinical Performance Data: Testing was comprised of static shear compression, dynamic axial compression and dynamic bending (per ASTM F2077), expulsion (per ASTM F-04.25.02.02), wear debris analysis (per ASTM F1877), low cycle cadaver testing to evaluate clinical stability and device expulsion/pull-out, and high cycle cadaver testing to evaluate device loosening or migration, blade back-out and vertebral body damage. The results of this testing demonstrate that the performance of the LDR Spine ROI-A Oblique implant, when compared with its legally marketed predicate, is substantially equivalent. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 SEP 30 2011 LDR Spine USA % Ms. Maritza Elias Regulatorv Affairs Project Manager 4030 West Braker Lane. Suite 360 Austin, Texas 78759 Re: K110327 Trade/Device Name: LDR ROI-A Implant System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: II Product Code: OVD Dated: August 31, 2011 Received: September 1, 2011 Dear Ms. Elias: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or 10 devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Insting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ Page 2 – Ms. Maritza Elias CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.him for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Eunl Keith + Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # KI10327 ## INDICATIONS FOR USE 510(k) Number (if known): K110327 LDR Spine USA ROI-A® Implant System Device Name: ### Indications for Use: When used as an intervertebral body fusion device, the ROI-A Implant System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) sheet 1 of 1 Division of Surgical, Orthopedic, and Restorative Devices KII0327 510(k) Number_
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...