iFuse-TORQ® Implant System

K213667 · SI-BONE, Inc. · OUR · Jun 10, 2022 · Orthopedic

Device Facts

Record IDK213667
Device NameiFuse-TORQ® Implant System
ApplicantSI-BONE, Inc.
Product CodeOUR · Orthopedic
Decision DateJun 10, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The iFuse-TORQ Implant System is indicated for: Fusion of the sacroiliac joint for sacroiliac joint dysfunction including sacroiliac joint disruption and degenerative sacroiliitis Fracture fixation of the pelvis, including acute, non-acute and non-traumatic fractures

Device Story

iFuse-TORQ Implant System consists of cannulated, fully threaded or lag-design metallic implants and associated surgical instruments. Implants allow packing of autograft/allograft materials; optional washers provide support in poor bone quality. Compatible with 3.2 mm guidewires. Used by surgeons in clinical/OR settings for sacroiliac joint fusion and pelvic fracture fixation. Provides mechanical stabilization of bone segments to facilitate fusion or fracture healing. Benefits patients by restoring joint stability and addressing pelvic fractures.

Clinical Evidence

Bench testing only. Verification testing performed per ASTM F543, including axial pullout, torsional strength, and insertion/removal torque. Clinical literature review conducted to support expanded indications.

Technological Characteristics

Metallic bone fixation fasteners; cannulated design; fully threaded or lag configurations; optional washers. Compatible with 3.2 mm guidewires. Testing performed per ASTM F543. No software or electronic components.

Indications for Use

Indicated for patients requiring sacroiliac joint fusion due to dysfunction (disruption, degenerative sacroiliitis) or pelvic fracture fixation (acute, non-acute, non-traumatic).

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ June 10, 2022 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around a stylized image of a human figure. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked above the word "ADMINISTRATION". SI-BONE, Inc. Meirav Harsat Director, Regulatory Affairs 471 El Camino Real, Suite 101 Santa Clara, California 95050 Re: K213667 Trade/Device Name: iFuse-TORO® Implant System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: OUR, HWC Dated: May 17, 2022 Received: May 18, 2022 Dear Meirav Harsat: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K213667 Device Name iFuse-TORQ Implant System Indications for Use (Describe) The iFuse-TORQ Implant System is indicated for: · Fusion of the sacroiliac joint for sacroiliac joint dysfunction including sacroiliac joint disruption and degenerative sacroiliitis · Fracture fixation of the pelvis, including acute, non-acute and non-traumatic fractures **Type of Use (Select one or both, as applicable)** | ☑ Executive Director/CEO/COO/Subject Company | ☐ For-Profit Consultant/Advisor/Subject Company | |----------------------------------------------|-------------------------------------------------| |----------------------------------------------|-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY – iFuse-TORQ® Implant System #### I. SUBMITTER SI-BONE, Inc. 471 El Camino Real, Suite 101, Santa Clara, CA 95050 Phone: 408-207-0700; Fax: 408-557-8312 Contact Person: Meirav Harsat, Director of Regulatory Affairs, SI-BONE, Inc. Email: mharsat(@si-bone.com Phone: 650-862-4942 Date Prepared: June 7, 2022 #### II. DEVICE | Trade Name of Device: | iFuse-TORQ® Implant System | |-----------------------|---------------------------------------------------------------------| | Common or Usual Name: | Sacroiliac Joint Fixation | | Regulation Number: | 21 CFR 888.3040, Smooth or threaded metallic bone fixation fastener | | Product Code: | OUR and HWC | #### III. PREDICATE AND REFERENCE DEVICES | Primary Predicate: | iFuse-TORQ Implant System K203247 | |-----------------------|------------------------------------| | Additional Predicate: | Firebird SI Fusion System, K210667 | #### IV. DEVICE DESCRIPTION The iFuse-TORO Implant System consists of the iFuse-TORO Implants and associated Instruments. iFuse-TORQ implants are fully threaded or with a lag design and provided with optional washers. The washers are intended to add additional support under the head of the screw in situations where the bone quality is poor. The cannulated implants are compatible with off-the-shelf 3.2 mm guidewires. The iFuse-TORQ implants, that are available in various lengths and diameters, allow for packing of autograft and allograft materials. #### V. PURPOSE This 510(k) premarket notification is submitted to request clearance to expand the indications for use the subject device. #### VI. INDICATIONS FOR USE The iFuse-TORQ Implant System is indicated for: - . Fusion of the sacroiliac joint for sacroiliac joint dysfunction including sacroiliac joint disruption and degenerative sacroiliitis - Fracture fixation of the pelvis, including acute, non-acute and non-traumatic fractures {4}------------------------------------------------ ### VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The subject device has identical technological characteristics compared to the primary predicate device and similar technological characteristics compared to the additional predicate. The clinical literature review and the design verification test results confirm that the differences do not raise new questions of safety and effectiveness. #### VIII. PERFORMANCE DATA The verification testing conducted per ASTM F543, including axial pullout, torsional strength, and insertion and removal torque, demonstrated that the technological differences between the subject device and the additional predicate do not raise new questions of safety and effectiveness and that the subject device and the additional predicate are substantially equivalent. #### IX. CONCLUSIONS The proposed subject device has the same intended use as the predicates and an indication for use that is substantially equivalent to the predicate devices. Based on literature search results, it was demonstrated that the differences in the indication for use statement from previous cleared legally marketed predicates are not critical to the intended therapeutic, diagnostic, prosthetic, or surgical use of the device. The proposed device has identical technological characteristics to the primary predicate and similar technological characteristics to the additional predicate. Based on the performance data test results, it was demonstrated that the differences do not raise new questions of safety and effectiveness and do not affect the safety and effectiveness of the device when used as labeled. As such, the subject device has been shown to be substantially equivalent to the primary and additional predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...