nvc

K142328 · Nvision Biomedical Technologies, LLC · ODP · Dec 23, 2014 · Orthopedic

Device Facts

Record IDK142328
Device Namenvc
ApplicantNvision Biomedical Technologies, LLC
Product CodeODP · Orthopedic
Decision DateDec 23, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The nv is intended for spinal fusion procedures at one level, from C2-T1, in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine. DDD is defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies. One device is to be used per intervertebral space. Patients should receive six weeks of non-operative treatment prior to treatment with an intervertebral body fusion device. The ny devices must be used with supplemental fixation and are designed for use with autograft bone to facilitate fusion. The devices are to be implanted via an anterior approach.

Device Story

The nv is an intervertebral body fusion device for cervical spine procedures. It is implanted via an anterior approach following discectomy. The device features a hollow center to accommodate autograft bone and includes anti-migration surfaces to improve stability. It is used in conjunction with supplemental fixation. The device is intended for use by surgeons in a clinical/hospital setting. It provides structural support to the intervertebral space to facilitate fusion, potentially reducing neck pain associated with degenerative disc disease.

Clinical Evidence

Bench testing only. Performance testing included static and dynamic axial compression (ASTM F2077), static and dynamic torsion (ASTM F2077), subsidence (ASTM F2267), and expulsion testing.

Technological Characteristics

Materials: PEEK Optima® LT1 (ASTM F2026) with tantalum markers (ASTM F560). Design: Hollow center for bone graft, multiple footprints/heights, anti-migration surface features. Energy source: None (mechanical). Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) requiring spinal fusion at one level. DDD defined as neck pain of discogenic origin with confirmed disc degeneration. Requires six weeks of prior non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure in profile, with three faces overlapping to create a sense of depth and dimension. The figure is positioned to the right of the department's name, which is arranged in a circular pattern around the figure. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 23, 2014 Nvision Biomedical Technologies, LLC % Ms. Jennifer Palinchik JALEX Medical, LLC 27881 Clemens Road, Suite 2 Westlake, Ohio 44145 Re: K142328 Trade/Device Name: nv Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP Dated: November 25, 2014 Received: November 26, 2014 Dear Ms. Palinchik: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Lori A. Wiggins -S for Mark N. Melkerson Director, Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K142328 Device Name nv C #### Indications for Use (Describe) The nv is intended for spinal fusion procedures at one level, from C2-T1, in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine. DDD is defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies. One device is to be used per intervertebral space. Patients should receive six weeks of non-operative treatment with an intervertebral body fusion device. The ny' devices must be used with supplemental fixation and are designed for use with autograft bone to facilitate fusion. The devices are to be implanted via an anterior approach. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | Submitted By: | Nvision Biomedical Technologies, LLC<br>18618 Tuscany Stone, Suite 120 | |------------------------|------------------------------------------------------------------------| | | San Antonio, TX 78258 | | Date: | October 29, 2014 | | Contact Person: | Jennifer Palinchik | | | Regulatory Consultant | | Contact Telephone: | (440) 930-2015 | | Device Trade Name: | nvC | | Classification Name: | Intervertebral Body Fusion Device | | Device Classification: | Class II | | Reviewing Panel: | Orthopedic | | Regulation Number: | 888.3080 | | Product Code: | ODP | | Predicate Device: | Primary Predicate: X-Spine Calix™ Cervical Interbody | | | Spacer (K083637) | | | Other Predicate: Genesys Spine Apache™ Cervical | | | Interbody System (K103034) | | | These predicates have not been subject to a design-related<br>recall. | ## Device Description: The nv is an intervertebral body fusion device used in the cervical spine following discectomy. All devices are manufactured from PEEK Optima® LT1 per ASTM F2026 and include tantalum markers per ASTM F560 for radiographic visualization. The devices have multiple footprints to adapt to the general shape of the vertebral endplates and have a hollow centre to accommodate bone graft. The devices are implanted via an anterior (ACIF) surgical approach. Each footprint is available in multiple heights to accommodate patient variability and there are anti-migration features on the superior and inferior surfaces designed to improve fixation, stability and prevent back out. ## Intended Use: The nv is intended for spinal fusion procedures at one level, from C2-T1, in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine. DDD is defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies. One device is to be used per intervertebral space. Patients should receive six weeks of non-operative treatment prior to treatment with an intervertebral body fusion device. The ny {4}------------------------------------------------ devices must be used with supplemental fixation and are designed for use with autograft bone to facilitate fusion. The devices are to be implanted via an anterior approach. # Comparison of Technological Characteristics: The technological characteristics of the nv device, including materials, size, shape and usage were compared to a predicate device. The comparison of information provided in this Premarket Notification supports the substantial equivalence, material information, and analysis of data. ## Mechanical Testing: Performance testing was conducted in accordance with the following standardized tests: - Static and Dynamic Axial Compression per ASTM F2077 ● - Static and Dynamic Torsion per ASTM F2077 ● - Subsidence per ASTM F2267 ● - Expulsion The results of the performance testing demonstrate substantial equivalence of the nv to the predicate device. # Conclusion: Non-clinical data presented in this Premarket Submission supports the substantial equivalence of the subject device to the predicate. Based on the indications for use, technological characteristics, and comparison with the predicate device, the subject device has demonstrated substantial equivalence.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...