C-FUSE CERVICAL INTERVERTEBRAL BODY FUSION SYSTEM

K130948 · Southern Spine, LLC · ODP · Aug 14, 2013 · Orthopedic

Device Facts

Record IDK130948
Device NameC-FUSE CERVICAL INTERVERTEBRAL BODY FUSION SYSTEM
ApplicantSouthern Spine, LLC
Product CodeODP · Orthopedic
Decision DateAug 14, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Southern Spine C Fuse™ Cervical Intervertebral Body Fusion System is intended for single level spinal fusion procedures in skeletally mature patients with degenerative disc disease ("DDD") of the cervical spine. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The C-Fuse device is to be implanted via an anterior approach at the C3 to C7 disk levels using autogenous bone graft and is to be combined with supplemental fixation. Patients should have had at least six weeks of non-operative treatment prior to treatment with the device.

Device Story

C-Fuse Cervical Intervertebral Body Fusion System; trapezoidal interbody spacer; hollow center core for autogenous bone graft; machined grooves on superior/inferior surfaces. Implanted via anterior approach at C3-C7 levels; provides structural stability; adjunct to supplemental spinal fixation. Used in surgical setting by orthopedic/neurosurgeons. Benefits patient by facilitating spinal fusion in degenerative disc disease cases.

Clinical Evidence

Bench testing only. No clinical data. Testing performed per ASTM F2077 and ASTM F2267, including static/dynamic axial compression, compression shear, torsion, subsidence, and expulsion testing.

Technological Characteristics

Materials: Zeniva® ZA-500 (PEEK) per ASTM F2026 and Tantalum Wire Type R05400 per ASTM F560. Trapezoidal form factor with hollow core. Non-sterile, single-use implants; non-sterile, reusable instrumentation. Steam sterilization.

Indications for Use

Indicated for single-level spinal fusion in skeletally mature patients with cervical degenerative disc disease (DDD) confirmed by history and radiography. Requires anterior approach at C3-C7 levels, autogenous bone graft, and supplemental fixation. Patients must have failed at least six weeks of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ AUG 1 4 2013 ## Southern Spine LLC Traditional 510(k) - C Fuse™ Cervical Intervertebral Body Fusion System # 510(k) SUMMARY This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements under 21 CFR 807.92. | Submitted By: | Southern Spine LLC<br>487 Cherry Street - Third Street Tower<br>Macon, GA 31201<br>Phone: (478) 745-0000     Fax: (478) 744-9996 | |-----------------|----------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Julie Stephens, President/Consultant<br>Regulatory Resources Group, Inc. | | Date Submitted: | Revised - August 5, 2013 | ## Device Name and Classification: | Trade/Proprietary Name: | C-Fuse™ Cervical Intervertebral Body Fusion System | |-------------------------|----------------------------------------------------| | Common Name: | Cervical Intervertebral Body Fusion System | | Classification Name: | Intervertebral body fusion device | | Product Code: | ODP | | Regulation: | 21 CFR 888.3080 | | Class: | 2 | ## Legally Marketed Predicate Device: Spinal Devices, LLC [Current Name is Amendia, Inc.] - Phenix Cervical Interbody Device (Phenix® CID) - 510(k) # K083167 Lanx, Inc. - Lanx Fusion System (Includes Breckenridge Interbody/VBR Fusion System) -510(k) # K103660 ## Device Description: The C.Fuse Cervical Intervertebral Body Fusion System is a cervical interbody fusion device designed to be inserted within the intervertebral disc space in order to provide structural stability and act as an adjunct to spinal fixation. The device is trapezoidal in shape, with machined grooves on the superior and inferior surfaces, along with a hollow center core to accept autogenous bone graft. There are two materials used in the manufacture of the Southern Spine C-Fuse Cervical Intervertebral Body Fusion System devices which are the Zeniva® ZA-500 (PEEK) under ASTM F2026 and Tantalum Wire Type R05400 under ASTM F560. There is instrumentation for implantation and a sterilization trav for steam sterilization. The implants are non-sterile and single use. The instrumentation is non-sterile and reusable. ## Indications for Use: The Southern Spine C Fuse™ Cervical Intervertebral Body Fusion System is intended for single level spinal fusion procedures in skeletally mature patients with degenerative disc disease ("DDD") of the cervical spine. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The C-Fuse device is to be implanted via an anterior approach at the C3 to C7 disk levels using autogenous bone graft and is to be combined with supplemental fixation. Patients should have had at least six weeks of non-operative treatment prior to treatment with the device. {1}------------------------------------------------ ## Southern Spine LLC Traditional 510(k) - C·Fuse™ Cervical Intervertebral Body Fusion System # 510(k) SUMMARY #### Similarities and Differences to the Predicate Devices: #### Similarities The same materials, the same performance standards, and the same indications for use are used in the C.Fuse™ Cervical Intervertebral Body Fusion System and the predicate devices. The C-Fuse and predicate devices include approximately the same size ranges for use in cervical C-3 to C-7 disc levels. The devices all have a hollow center core to accept autogenous bone graft. The C-Fuse and predicate devices provide instrumentation sets and sterilization travs as accessories for use during implantation. #### Differences The PEEK resin and tantalum materials are from different manufacturers than the predicates; however, no issue with biocompatibility was shown with the C·Fuse device. The C-Fuse and the predicate devices have roughly the same device dimensions and characteristics. ## Summary of Testing: The C-Fuse Cervical Intervertebral Body Fusion System was tested to the ASTM F2077 and ASTM F2267 standards as identified in "Guidance for Industry and FDA Staff Class II Special Controls Guidance Document: Intervertebral Body Fusion Device", Dated June 12, 2007 as appropriate for characterization of intervertebral body fusion devices. Testing included static axial and compression shear testing, dynamic axial and compression shear testing, static and dynamic torsion testing, static subsidence testing, and static expulsion testing. ### Substantial Equivalence Conclusions: The C-Fuse Cervical Intervertebral Body Fusion System has the same intended use and fundamental scientific technology as the predicate devices. The minor differences do not raise any issues of safety or effectiveness. Testing results support the determination of substantial equivalence for the C-Fuse Cervical Intervertebral Body Fusion System against the predicate devices. {2}------------------------------------------------ Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 14. 2013 Southern Spine, LLC % Regulatory Resources Group, Incorporated Ms. Julie Stephens President/Consultant 111 Laurel Ridge Drive Alpharetta. Georgia 30004 Re: KI30948 Trade/Device Name: C.Fuse™ Cervical Intervertebral Body Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP Dated: June 28, 2013 Received: June 28, 2013 Dear Ms. Stephens: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device. subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Ms. Julie Stephens forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Erin I. Keith For Mark Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health 100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K130948 Device Name: C·Fuse™ Cervical Intervertebral Body Fusion System Indications for Use: The Southern Spine C Fuse™ Cervical Intervertebral Body Fusion System is intended for single level spinal fusion procedures in skeletally mature patients with degenerative disc disease ("DDD") of the cervical spine. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The C-Fuse device is to be implanted via an anterior approach at the C3 to C7 disk levels using autogenous bone graft and is to be combined with supplemental fixation. Patients should have had at least six weeks of non-operative treatment with the device. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Anton E. Dmitriev, PhD Division of Orthopedic Devices Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...