CD HORIZON™ Spinal System

K173249 · Medtronic Sofamor Danek USA, Inc. · NKB · Mar 8, 2018 · Orthopedic

Device Facts

Record IDK173249
Device NameCD HORIZON™ Spinal System
ApplicantMedtronic Sofamor Danek USA, Inc.
Product CodeNKB · Orthopedic
Decision DateMar 8, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The CD HORIZON™ Spinal System with or without SEXTANT™ instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (DDD - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion. Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON™ Spinal System may also be used for the same indications as an adjunct to fusion. With the exception of DDD, the CD HORIZON™ LEGACY™ 3.5mm rods and the CD HORIZON™ Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below. When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON™ Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e. scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the CD HORIZON™ Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/ spondylolysis and fracture caused by tumor and/or trauma. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach. The CD HORIZON™ SPIRE™ Plate is a posterior, single-level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/ attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: DDD (as previously defined), spondylolisthesis, trauma, and/or tumor. In order to achieve additional levels of fixation, the CD HORIZON™ Spinal System rods may be connected to the VERTEXT™ Reconstruction System with the VERTEX™ rod connector. Refer to the VERTEX™ Reconstruction System Package Insert for a list of the VERTEX™ indications of use.

Device Story

The CD HORIZON™ Spinal System is a modular orthopedic implant set used by surgeons in the operating room to stabilize the spine. It consists of rods, hooks, screws, plates, and connectors that are assembled into custom constructs to provide rigid fixation during spinal fusion procedures. The system is designed to treat various spinal pathologies, including degenerative disc disease, deformities, and trauma. The device is implanted surgically; it provides mechanical support to the spine to facilitate bone fusion. The system is modified to include a sterile domino connector to allow for additional levels of fixation. The components are made of titanium alloy, cobalt chrome, and silicone. The system is used as an adjunct to autograft or allograft bone material. By providing rigid, tailor-made stabilization, the device helps correct spinal alignment and supports the healing process in patients with spinal instability or deformity.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including mechanical testing (ASTM F1798 for axial grip, torsion, and flexion-extension; ASTM F1717 for static compression, torsion, and fatigue) and bacterial endotoxin testing (LAL test) per ANSI/AAMI ST72 and USP <161>.

Technological Characteristics

Constructs consist of rods, hooks, screws, plates, and connectors. Materials: Ti-6Al-4V ELI alloy (ASTM F136), Cobalt-28Chromium-6Molybdenum alloy (ASTM F1537), and silicone elastomers (ISO 14949). Mechanical fixation via rigid locking mechanisms. Permanent implant (>30 days). Sterilization method not explicitly detailed but implied to be standard for surgical implants.

Indications for Use

Indicated for skeletally mature and pediatric patients requiring posterior, non-cervical spinal fixation as an adjunct to fusion. Conditions include DDD, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, spinal curvatures (scoliosis, kyphosis, lordosis), tumor, pseudarthrosis, and failed previous fusion. Pediatric use limited to posterior pedicle screw fixation for progressive spinal deformities (idiopathic, neuromuscular, congenital scoliosis), spondylolisthesis/spondylolysis, and tumor/trauma-related fractures. Contraindicated for non-fusion procedures; pediatric pedicle screw fixation limited to posterior approach.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 8, 2018 Medtronic Sofamor Danek USA, Inc. Mr. Ankit K. Shah Senior Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132 Re: K173249 Trade/Device Name: CD HORIZON™ Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB, KWP, KWO Dated: February 6, 2018 Received: February 7, 2018 Dear Mr. Shah: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Ronald P. Jean -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ | Form Approved: OMB No. | 0910-0120 | |--------------------------|------------| | Expiration Date: | 06/30/2020 | | See PRA Statement below. | | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration **Indications for Use** | 510(k) Number ( <i>if known</i> ) | K173249 | |-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | CD HORIZON <sup>TM</sup> Spinal System | | Indications for Use (Describe) | The CD HORIZON <sup>TM</sup> Spinal System with or without SEXTANT <sup>TM</sup> instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (DDD - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.<br>Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON <sup>TM</sup> Spinal System may also be used for the same indications as an adjunct to fusion. With the exception of DDD, the CD HORIZON <sup>TM</sup> LEGACY <sup>TM</sup> 3.5mm rods and the CD HORIZON <sup>TM</sup> Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.<br>When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON <sup>TM</sup> Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e. scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the CD HORIZON <sup>TM</sup> Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/ spondylolysis and fracture caused by tumor and/or trauma. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.<br>The CD HORIZON <sup>TM</sup> SPIRE <sup>TM</sup> Plate is a posterior, single-level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/ attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: DDD (as previously defined), spondylolisthesis, trauma, and/or tumor.<br>In order to achieve additional levels of fixation, the CD HORIZON <sup>TM</sup> Spinal System rods may be connected to the VERTEXT <sup>TM</sup> Reconstruction System with the VERTEX <sup>TM</sup> rod connector. Refer to the VERTEX <sup>TM</sup> Reconstruction System Package Insert for a list of the VERTEX <sup>TM</sup> indications of use. | | Type of Use (Select one or both, as applicable) | <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | |-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff [PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov) "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {3}------------------------------------------------ ## 510(k) SUMMARY MEDTRONIC Sofamor Danek CD HORIZON™ Spinal System February 2018 | I. | Submitter | Medtronic Sofamor Danek USA, Inc. | |------|-----------------------|----------------------------------------------------------| | | | 1800 Pyramid Place | | | | Memphis, Tennessee 38132 | | | | Telephone: (901)396-3133 | | | | | | | Contact: | Ankit K. Shah | | | | Sr. Regulatory Affairs Specialist | | | | Phone: (901) 344-1272 | | | Date Prepared: | February 6, 2018 | | II. | Device | | | | Name of Device: | CD HORIZON™ Spinal System | | | Classification Names: | Thoracolumbosacral Pedicle Screw System | | | | (21 CFR 888.3070) | | | Class: | Class II | | | Product Code: | NKB, KWP, KWQ | | III. | Predicate Devices: | Primary Predicate: | | | | • CD HORIZON Spinal System, K090390 | | | | (S.E. 05/15/2009) | | | | Additional Predicates: | | | | • TSRH™ Spinal System, K030285 (S.E. 02/25/2003) | | | | • CD HORIZON Spinal System,<br>K062606 (S.E. 09/25/2006) | | | | • CD HORIZON Spinal System,<br>K042025 (S.E. 08/25/2004) | {4}------------------------------------------------ - CD HORIZON Spinal System, . K132639 (S.E 11/25/2013) - . CD HORIZON Spinal System, K132328 (S.E 12/06/2013) Reference predicates have been mentioned in certain sections. The predicates have not been subject to a design related recall. #### IV. Description: The CD HORIZON™ Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK™ Plates, staples, and connecting components, as well as implant components from other Medtronic spinal systems which can be rigidly locked into a variety of configurations with each construct being tailor-made for the individual case. A subset of CD HORIZON™ Spinal System components may be used for posterior pedicle screw fixation in pediatric cases. These constructs may be comprised of a variety of shapes and sizes of rods (ranging in diameter from 3.5mm to 6.35mm), hooks, screws, CROSSLINK™ Plates and connecting components. Similarly, to the CD HORIZON™ implants used in adult cases, these components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case. The purpose of this Traditional 510(k) is to modify Medtronic's CD HORIZON™ Spinal System to add additional components to the system, specifically, a sterile domino connector. The subject device is made of Titanium Alloy, Cobalt Chrome and Silicone. #### V. Indications for Use: The CD HORIZON™ Spinal System with or without SEXTANT™ instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (DDD - defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), {5}------------------------------------------------ spondylolisthesis, trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion. Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON™ Spinal System may also be used for the same indications as an adjunct to fusion. With the exception of DDD, the CD HORIZON™ LEGACY™ 3.5mm rods and the CD HORIZON™ Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below. When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON™ Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e. scoliosis, kyphosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the CD HORIZON™ Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis and fracture caused by tumor and/or trauma. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach. The CD HORIZON™ SPIRE™ Plate is a posterior, single-level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: DDD (as previously defined), spondylolisthesis, trauma, and/or tumor. In order to achieve additional levels of fixation, the CD HORIZON™ Spinal System rods may be connected to the VERTEX™ Reconstruction System with the VERTEX™ rod connector. Refer to the VERTEX™ Reconstruction System Package Insert for a list of the VERTEX™ indications of use. {6}------------------------------------------------ #### VI. Comparison of Technological Characteristics with the Predicate Devices: The subject CD HORIZON™ Spinal System has the same indications, intended use, fundamental scientific technology, materials, and sterilization method as the previously FDA cleared predicates. The indications and the fundamental technology is identical to the FDA cleared CD HORIZON™ Spinal System, 510(k) K090390 (S.E. 05/15/2009). Other features and materials of the device within subject CD HORIZON Spinal System submission are identical to the following predicates: - TSRH Spinal System, K030285 (S.E. 02/25/2003) . - CD HORIZON™ Spinal System, K062606 (S.E. 09/25/2006) . - CD HORIZON Spinal System, K042025 (S.E. 08/25/2004) . - CD HORIZON Spinal System K132639 (S.E 11/25/2013) ● - CD HORIZON Spinal System, K132328 (S.E 12/06/2013) ● #### VII. Performance Data: The following information is provided in support of substantial equivalence. Biocompatibility The subject CD HORIZON™ Spinal System implants are permanent implants (> 30 days) and will be classified as body contacting devices according to FDA's Draft Guidance for Industry and FDA Staff "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing". The subject implants and instruments are manufactured from identical materials as the predicate devices, in accordance with the following standards: - . ASTM F136: Standard Specification for Wrought Ti-6A1-4V ELI Alloy for Surgical Implant - ASTM F1537: Standard Specification for Wrought Cobalt-28Chromium-6Molybdenum Alloys for Surgical Implants - . ISO14949: Implants for surgery -- Two-part addition-cure silicone elastomers {7}------------------------------------------------ The materials used for manufacturing the subject device have a long history of safe and effective use in predicate spinal implants, and biocompatibility testing is not required. ## Mechanical Testing Non-clinical mechanical testing was performed in accordance with ASTM F1798 including Axial Grip, Axial Torsion and Flexion Extension Static and ASTM F1717 static compression, static torsion and compression fatigue. The subject devices met the pre-determined acceptance criteria. Medtronic believes that the subject implants do not introduce a new worst case scenario and are substantially equivalent to the predicate device. ### Non-Pyrogenicity Endotoxin Testing The bacterial endotoxin test, also known as Limulus amebocyte lysate (LAL) test, was performed utilizing worst case subject implants to verify that the subject implants meet the 20 endotoxin units (EU)/device pyrogen limit specification. Testing was successfully performed, and it was confirmed that the subject implants meet the 20 EU/device testing limit for general medical devices that are implanted as outlined in ANSI/AAMI ST72, Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing and USP <161>, Medical Devices – Bacterial Endotoxin and Pyrogen Tests. #### VIII. Conclusion: A risk analysis was completed, and non-clinical testing was performed in accordance with ASTM F1798 and ASTM F1717. Based on the test results and additional supporting information provided in this pre-market notification, the subject CD HORIZON™ Spinal System is substantially equivalent to the following predicates: Primary Predicate: - · CD HORIZON™ Spinal System, K090390 (S.E. 05/15/2009) Additional Predicates: - · TSRH™ Spinal System, K030285 (S.E. 02/25/2003) - · CD HORIZON™ Spinal System, K062606 (S.E. 09/25/2006) {8}------------------------------------------------ - CD HORIZON™ Spinal System, K042025 (S.E. 08/25/2004) - CD HORIZON Spinal System K132639 (S.E 11/25/2013) - CD HORIZON Spinal System, K132328 (S.E 12/06/2013)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...