RANGE /DENALI/MESA and SMALL STATURE and ARI are cleared for the following indications: Non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation): spinal stenosis: curvatures (i.e. scoliosis, kyphosis and/or lordosis); tumor: pseudarthrosis: and/or failed previous fusion. Except for hooks, when used as an anterolateral thoracic/lumbar system the Range Spinal System may also be used for the same indications as an adjunct to fusion. Except for the ARI staples, the Range Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical fixation in pediatric patients. The Range Spinal System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
Device Story
RANGE/DENALI/MESA Spinal Systems are top-loading, posterior thoracic-lumbar spinal fixation systems. Components include pedicle screws, rods, hooks, and rod connectors. Modular hooks consist of a hook body and adjustable blade options. Used by surgeons in clinical settings to provide stabilization as an adjunct to spinal fusion. Device components are implanted to fixate spinal segments, aiding in patient recovery from degenerative conditions, trauma, or deformities. Output is mechanical stabilization of the spine.
Clinical Evidence
Bench testing only. Mechanical performance of the K2M Modular Hook was evaluated using static and dynamic compression bending tests per ASTM F1717 and static A-P and disassembly tests per ASTM F1798. Results demonstrate performance equivalent to the predicate MESA Spinal System.
Technological Characteristics
Components manufactured from titanium alloy (Ti-6Al-4V ELI) per ASTM and ISO standards. System is a top-loading, posterior thoracic-lumbar fixation construct. Modular hook design features adjustable blade options. Mechanical testing performed per ASTM F1717 and ASTM F1798.
Indications for Use
Indicated for non-cervical spinal fixation as an adjunct to fusion in patients with degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, spinal curvatures (scoliosis, kyphosis, lordosis), tumor, pseudarthrosis, or failed previous fusion. Also indicated for adolescent idiopathic scoliosis in pediatric patients via posterior approach using autograft/allograft.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
K140765 — RANGE/MESA/DENALI SPINAL SYSTEM · K2m, Inc. · Jul 10, 2014
K131784 — RANGE SPINAL SYSTEM · K2m, Inc. · Aug 30, 2013
K112920 — RANGE SPINAL SYSTEM (MESA AND DENALI) · K2m, Inc. · Dec 21, 2011
K121630 — RANGE SPINAL SYSTEM · K2m, Inc. · Sep 11, 2012
K121016 — RANGE SPINAL SYSTEM (MESA AND DENALI) · K2m, Inc. · Jun 27, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 3 0 2014
:
# 510(k) Summary
.
.
| Date: | 5 November 2013 |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | K2M Inc.<br>751 Miller Drive SE<br>Leesburg, Virginia 20175<br>Phone: 703.777.3155 |
| Contact Person: | Nancy Giezen |
| Trade Names: | RANGE (DENALI / MESA) Spinal Systems |
| Device Classification | Class II, Class III |
| Classification Names<br>(Regulations): | Spinal Interlaminal Fixation Orthosis (888.3050)<br>Spinal intervertebral body fixation orthosis (888.3060)<br>Pedicle Screw Spinal System (888.3070) |
| Device Product<br>Codes: | KWP, KWQ, MNI, MNH, NKB, OSH |
| Submission Purpose: | This submission adds the K2M Modular Hooks to the Range<br>(Denali/Mesa) Spinal System. |
| Device Description: | The RANGE / DENALI / MESA Spinal Systems are top-loading,<br>multiple component, posterior (thoracic-lumbar) spinal fixation<br>systems which consists of pedicle screws, rods, hooks and rod<br>connectors.<br><br>The K2M Modular Hook is comprised of a hook body and a variety of<br>adjustable blade options. |
| Intended Use: | RANGE /DENALI/MESA and SMALL STATURE and ARI are cleared<br>for the following indications:<br><br>Non-cervical fixation as an adjunct to fusion for the following<br>indications: degenerative disc disease (defined as back pain of<br>discogenic origin with degeneration of the disc confirmed by history<br>and radiographic studies); spondylolisthesis; trauma (i.e., fracture or<br>dislocation): spinal stenosis: curvatures (i.e. scoliosis, kyphosis<br>and/or lordosis); tumor: pseudarthrosis: and/or failed previous fusion.<br>Except for hooks, when used as an anterolateral thoracic/lumbar<br>system the Range Spinal System may also be used for the same<br>indications as an adjunct to fusion. |
| | Except for the ARI staples, the Range Spinal System is indicated as<br>an adjunct to fusion to treat adolescent idiopathic scoliosis when<br>used for posterior noncervical fixation in pediatric patients. The<br>Range Spinal System for pediatric use is intended to be used with<br>autograft and/or allograft. Pediatric pedicle screw fixation is limited<br>to a posterior approach. |
| Materials: | The K2M Modular Hooks components are manufactured from<br>titanium alloy (Ti-6Al-4V ELI) as described by ASTM and ISO<br>standards. |
| Predicate Devices: | K2M MESA® Spinal System (K052398/K080611)<br>K2M Range (Mesa/Denali) Spinal System (K141147)<br>Medtronic CD Horizon (K962708/K031655) |
K133402 Page 1 of 2
{1}------------------------------------------------
Performance Data: The worst case K2M Modular Hook was evaluated using static and dynamic compression bending tests according to ASTM F1717, and static A-P and disassembly tests according to ASTM F1798. The mechanical test results demonstrate that the K2M Modular Hook performance is substantially equivalent to the predicate MESA Spinal System performance.
> The K2M Modular Hook components possess the same technological characteristics as the predicate devices. These include:
- basic design (rod-based hook attachment), .
- . material (titanium alloy),
- . anatomic location (non-cervical spine) and
- . sizes (widths, lengths and heights are within the range(s) offered by the predicate systems).
Technological characteristics which are different have been supported with descriptive information and/or performance data which demonstrate the safety and effectiveness has not been diminished.
#### Conclusion:
Technological
Characteristics:
In comparison to the predicate devices, the K2M Modular Hooks has
- the same intended use (as described above),
- . the same technological characteristics or different without raising safety and effectiveness issues (as described above)
Therefore the K2M Modular Hooks can be found substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. The logo is simple, using a black and white color scheme.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 30, 2014
K2M, Incorporated % Karen E. Warden, Ph.D. BackRoads Consulting, Incorporated 8202 Sherman Road Chesterland, Ohio 44026
Re: K133402
Trade/Device Name: RANGE (DENALI/MESA) Spinal Systems Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNI. MNH, KWP, KWQ Dated: June 26, 2014 Received: June, 27, 2014
Dear Dr. Warden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Karen E. Warden, Ph.D.
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Ronald P. Jean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known) K133402
Device Name
RANGE (DENALI/MESA) Spinal Systems
Indications for Use (Describe)
RANGE/DENALI/MESA and SMALL STATURE and ARI are cleared for the following indications:
Non-cervical fixation as an adjunct to following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.
Except for hooks, when used as an anterolateral the Range Spinal System may also be used for the same indications as an adjunct to fusion.
Except for the AR1 staples, the Range Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical fixation in pediatic patients. The Range Spinal System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
## Zane W=Wyatt Division of Orthopedic Devices
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to. a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.