MODIFICATION TO: S4 SPINAL SYSTEM

K090657 · Aesculap Implant Systems · NKB · Apr 20, 2009 · Orthopedic

Device Facts

Record IDK090657
Device NameMODIFICATION TO: S4 SPINAL SYSTEM
ApplicantAesculap Implant Systems
Product CodeNKB · Orthopedic
Decision DateApr 20, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The S4 Spinal System, when used as a pedicle screw fixation system, is indicated for use in patients: a) having severe spondylolisthesis (Grade 3 and 4) at the L5-S1 ioin: b) who are receiving fusion using autogenous graft only; c) who are having the device fixed or attached to the lumbar or sacral spin; and d) who are having the device removed after the development of a solid fusion mass may not go above the L5-S1 joint, the levels of pedicle screw fixation may span from L3 to the sacrum. The S4 Spinal System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spin: degenerative spondyloisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis). When used as a hook and sacral screw system (other than pedicle screw fixation system for high grade spondylolisthesis), the S4 Spinal System is intended for use in the treatment of degenerative disc disease (as defined by chronic back pain of discogenic origin with the degeneration of the disc confirmed by a history and radiographic studies). idiopathic scoliosis, spondylolisthesis, kyphotic or lordotic deformity of the spine, loss of stability due to tumors, spinal stenosis, vertebral fracture of dislocation, pseudoarthrosis, and previous failed spinal surgery/fusion. When used for this indication, screws of the S4 Spinal System are intended for sacral/illac attachment only. Hooks and transverse connectors of the system are intended for posterior throracic and/or lumbar use only. The levels of use for hook and sacral screw fixation of this system are T1 to the sacrum.

Device Story

S4 Spinal System is a posterior spinal fixation system consisting of polyaxial/monoaxial screws, hooks, rods, and rod-to-rod connectors. Components are top-loading and top-tightening. Used by surgeons in clinical settings to provide immobilization and stabilization of spinal segments as an adjunct to fusion. Device provides mechanical support to the spine to facilitate fusion in patients with various instabilities or deformities. Benefits include stabilization of the spinal column, potentially reducing pain and correcting deformities. No electronic or software components.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Materials: Titanium and Titanium alloy per ISO 5832/3 and ISO 5832/2. Components: polyaxial/monoaxial screws, hooks, rods, rod-to-rod connectors. Design: top-loading, top-tightening. Mechanical fixation system. No software or energy source.

Indications for Use

Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion. Conditions: degenerative spondylolisthesis, fracture, dislocation, scoliosis, spinal tumor, pseudarthrosis, degenerative disc disease, kyphotic/lordotic deformity, spinal stenosis. Levels: T1 to sacrum (hooks/sacral screws); L3 to sacrum (pedicle screws). Severe spondylolisthesis (Grade 3/4) at L5-S1 requires autogenous graft.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ S4 Spinal System Line Extension Page 1 of 2 510(k) SUMMARY (as required by 21 CFR 807.92) B. APR 2 0 2009 S4 Spinal System March 11, 2009 Aesculap Implant Systems, Inc. COMPANY: 3773 Corporate Parkway Center Valley, PA 18034 Establishment Reqistration Number: 3005673311 CONTACT: Lisa M. Bovle 610-984-9274 (phone) 610-791-6882 (fax) lisa.boyle@aesculap.com (email) TRADE NAME: ਟੇਖ COMMON NAME: S4 Spinal System DEVICE CLASS: Class III PRODUCT CODE: NKB, MNI, MNH, and KWP REGULATION NUMBER: 888.3070 - Orthosis, Spinal Pedicle Fixation For Degenerative Disc Disease 888.3070 - Orthosis, Spinal Pedicle Fixation 888.3070 - Orthosis, Spondyloisthesis Spinal Fixation 888.3050 - Appliance, Fixation, Spinal Fixation Orthopedics REVIEW PANEL: # SUBSTANTIAL EQUIVALENCE Aesculap Implant Systems, Inc. believes that the S4 Spinal System additions are substantially equivalent to the existing components of the S4 Spinal System (K071945/K062085/K032219). # DEVICE DESCRIPTION The S4 Spinal System consists of polyaxial screws and monoaxial screws of varying diameters and lengths, various hook styles, rods of varying lengths, and fixed and adjustable rod to rod connectors. All implant components are top loading and top tightening. The S4 Spinal System is manufactured from Titanium and Titanium alloy in accordance with ISO 5832/3 and ISO 5832/2. # INDICATIONS FOR USE The S4 Spinal System, when used as a pedicle screw fixation system, is indicated for use in patients: a) having severe spondylolisthesis (Grade 3 and 4) at the L5-S1 join; b) who are receiving fusion using autogenous graft only; c) who are having the device fixed or {1}------------------------------------------------ SPECIAL 510(k) Premarket Notification 54 Spinal System Page 2 of 2 attached to the lumbar or sacral spine; and d) who are having the device removed after the development of a solid fusion mass may not go above the L5-S1 joint, the levels of pedicle screw fixation may span from L3 to the sacrum, The S4 Spinal System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spin: degenerative spondyloisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis). When used as a hook and sacral screw system (other than pedicle screw fixation system for high grade spondylolisthesis), the S4 Spinal System is intended for use in the treatment of degenerative disc disease (as defined by chronic back pain of discogenic origin with the degeneration of the disc confirmed by a history and radiographic studies), idiopathic scoliosis, spondylolisthesis, kyphotic or lordotic deformity of the spine, loss of stability due to tumors, spinal stenosis, vertebral fracture of dislocation, pseudoarthrosis, and previous failed spinal surgery/fusion. When used for this indication, screws of the S4 Spinal System are intended for sacral/iliac attachment only. Hooks and transverse connectors of the system are intended for posterior throracic and/or lumbar use only. The levels of use for hook and sacral screw fixation of this system are T1 to the sacrum. ### TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s)) The new components of the S4 Spinal System are offered in similar shapes and sizes as the predicate devices. All the components are manufactured from Titanium and Titanium Allov, which is the same material as the predicate devices. #### PERFORMANCE DATA All required testing per "Draft Guidance for the Preparation of Premarket Notifications (510(k)s) Applications for Orthopedic Devices-The Basic Elements" were done where applicable. In addition, testing per the "Spinal System 510(k)s" was completed where applicable. {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/2/Picture/2 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract image of an eagle. Food and Drug Administration 200 Corporate Boulevard Rockville MD 20850 Aesculap Implant Systems % Lisa M. Boyle 3773 Corporate Parkway Center Valley, Pennsylvania 18034 APR 2 0 2009 Re: K090657 Trade/Device Name: S4 Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle Screw Spinal System Regulatory Class: Class III Product Code: NKB, MNI, MNH and KWP Dated: April 14, 2009 Received: April 15, 2009 Dear Ms. Boyle: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Lisa M. Boyle If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Darbere BrienD to mken Mark N. Melkerson Director Division of General, Restorative, And Neurological Devices Office of Device Evaluation Center for Devices and Radiological.Health Enclosure {4}------------------------------------------------ S4 Spinal System Line Extension K090657 Page 1 of 1 Page 1 of 1 #### A. INDICATIONS FOR USE STATEMENT K090657 510(k) Number: Device Name: S4 Spinal System Indications for Use: The S4 Spinal System, when used as a pedicle screw fixation system, is indicated for use in patients: a) having severe spondylolisthesis (Grade 3 and 4) at the L5-S1 ioin: b) who are receiving fusion using autogenous graft only; c) who are having the device fixed or attached to the lumbar or sacral spin; and d) who are having the device removed after the development of a solid fusion mass may not go above the L5-S1 joint, the levels of pedicle screw fixation may span from L3 to the sacrum. The S4 Spinal System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spin: degenerative spondyloisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis). When used as a hook and sacral screw system (other than pedicle screw fixation system for high grade spondylolisthesis), the S4 Spinal System is intended for use in the treatment of degenerative disc disease (as defined by chronic back pain of discogenic origin with the degeneration of the disc confirmed by a history and radiographic studies). idiopathic scoliosis, spondylolisthesis, kyphotic or lordotic deformity of the spine, loss of stability due to tumors, spinal stenosis, vertebral fracture of dislocation, pseudoarthrosis, and previous failed spinal surgery/fusion. When used for this indication, screws of the S4 Spinal System are intended for sacral/illac attachment only. Hooks and transverse connectors of the system are intended for posterior throracic and/or lumbar use only. The levels of use for hook and sacral screw fixation of this system are T1 to the sacrum. X Prescription Use______________________________________________________________________________________________________________________________________________________________ and/or Over-the-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANQTHER PAGE IF NEE Concurrence of CDRH, Office of Device Evalualis Division of General, Restorative, and Neurological Devices 002 510(k) Number k090057
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...