A literature review was conducted to establish the safety and effectiveness of the device for use in voids resulting from resected benign bone cysts/tumors in adult and pediatric patients (≥ 6 years old) to support an expanded indications statement.
Literature review; Benign bone cysts; Pediatric patients; Bone void filler
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Literature review for benign bone cysts/tumors; Literature review
Adult and pediatric patients (≥ 6 years old) with benign bone cysts/tumors
Not applicable for this study
Safety and effectiveness of bone graft substitute in resected benign bone cysts/tumors
Indications for Use
PRO-DENSE® resultant paste is intended for use as a bone graft substitute to be injected or digitally packed into open bone voids/gaps that are not intrinsic to the stability of bony structure of the skeletal system (i.e., the extremities and pelvis) to cure in situ. These open bone voids may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), surgically created osseous defects or osseous defects created from traumatic injury to the bone. The paste provides a bone graft substitute that resorbs and is replaced with bone during the healing process. The PRO-DENSE® paste cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K. Wires) to help support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process. PRO-DENSE® is provided sterile for single use only.
Device Story
PRO-DENSE® is a calcium sulfate/calcium phosphate composite bone graft substitute. Supplied as powder and liquid components; mixed into paste by clinician; injected or packed into non-structural bone voids/gaps. Cures in situ via hydration. Acts as osteoconductive scaffold; resorbs in three phases: calcium sulfate dissolves for early vascular infiltration; calcium phosphate resorbs via osteoclastic action; TCP granules resorb in final phase. Used in surgical settings to fill defects or augment provisional hardware (e.g., K-wires). Provides temporary support during healing; replaced by native bone. Benefits include osteoconductivity and gradual resorption.
Clinical Evidence
No clinical trials performed. Safety and effectiveness for the expanded indications (benign bone cysts/tumors in adult/pediatric patients) established via literature review of clinical use of calcium sulfate/calcium phosphate composites.
Technological Characteristics
Calcium sulfate and calcium phosphate composite. Supplied as powder and aqueous mixing solution. Hardens via hydration reaction. Triphasic resorption profile. Sterile, single-use.
Indications for Use
Indicated for adults and pediatric patients ≥ 6 years old with open bone voids/gaps in the extremities and pelvis resulting from benign bone cysts, tumors, surgically created osseous defects, or traumatic injury. Not for use in voids intrinsic to skeletal structural stability.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
<11387)
1/2
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
# MAR 2 5 2013
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CRF 807, this information serves as a Summary of Safety and Effectiveness for the use of PRO-DENSE® Bone Graft Substitute.
| Submitted By: | Wright Medical Technology, Inc. |
|-------------------------------------|------------------------------------------------------------------------------------------------------|
| Date: | March 15, 2013 |
| Contact Person: | Samir Ibrahim, PhD, RAC<br>Regulatory Affairs Specialist II<br>901-290-5909<br>samir.ibrahim@wmt.com |
| Proprietary Name: | PRO-DENSE® Bone Graft Substitute |
| Common Name: | Bone Void Filler |
| Classification Name and Reference: | 21 CFR 888.3045 - Resorbable Calcium<br>Salt Bone Void Filler Device - Class II |
| Device Product Code and Panel Code: | Orthopedic/87/MQV |
| Predicate Device: | K070437 - PRO-DENSE® Bone Graft Substitute |
| | |
#### DEVICE INFORMATION
#### A. INTENDED USE
PRO-DENSE® resultant paste is intended for use as a bone graft substitute to be injected or digitally packed into open bone voids/gaps that are not intrinsic to the stability of bony structure of the skeletal system (i.e., the extremities and pelvis) to cure in situ. These open bone voids may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), surgically created osseous defects or osseous defects created from traumatic injury to the bone. The paste provides a bone graft substitute that resorbs and is replaced with bone during the healing process.
The PRO-DENSE® paste cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K. Wires) to help support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process.
PRO-DENSE® is provided sterile for single use only.
### B. DEVICE DESCRIPTION
PRO-DENSE® is indicated as a bone graft substitute to be injected and/or digitally packed into open bone voids/gaps that are not intrinsic to the structural stability of the skeletal system to cure in-situ. It is supplied in separate powder and liquid vials along with the instruments for mixing it into a paste and delivering it to the defect site. The triphasic resorption of PRO-DENSE® results in a scaffold
{1}------------------------------------------------
that is osteoconductive allowing tissue infiltration and must eventually be degraded through osteoclastic action as bone remodels within the scaffold. The clinical use of calcium sulfate, calcium phosphate, and composites thereof as a bone void filler has been well established through many peer reviewed publications.
### C. MATERIALS
PRO-DENSE® is a calcium sulfate - calcium phosphate composite bone graft substitute consisting of a powder component and an aqueous mixing solution. When the two components are mixed according to directions, an injectable paste forms and subsequently hardens via hydration reactions. The benefits of this composite include:
- . Calcium Sulfate
- Primary osteoconductive filler :
- Resorbs first primarily through simple dissolution to allow early vascular . infiltration
- . Excellent clinical history
- Calcium Phosphate ●
- Osteoclastic resorption .
- I Secondary porous scaffold that is resorbed after primary filler
- 파 TCP granules are resorbed in the third and final phase
# D. SUBSTANTIAL EQUIVALENCE INFORMATION
#### Technological Characteristics Comparison
The technical features of PRO-DENSE® in this submission remain identical to the predicate product cleared in K070437.
#### Non-Clinical Evidence
N/A
#### Clinical Evidence
A literature review was conducted to establish the safety and effectiveness of PRO-DENSE® for use in voids resulting from resected benign bone cysts/tumors in both adult and pediatric patients. This evidence was used to change the indications statement to include this patient population.
#### Conclusion
The design characteristics of the subject device are identical to the predicate device and do not raise any new types of questions of safety or effectiveness. A literature search was conducted to support the addition of adult/pediatric benign bone cysts/tumor to the indications statement. From the evidence submitted in this 510(k), the subject devices can be expected to perform at least as well as the predicate system(s).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized emblem resembling an abstract eagle or bird-like figure with three curved lines forming its body and wings. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 25, 2013
Wright Medical Technology, Incorporated % Samir Ibrahim, Ph.D., RAC Regulatory Affairs Specialist II 5677 Airline Road Arlington, Tennessee 38002
Re: K113871
Trade/Device Name: PRO-DENSE® Bone Graft Substitute Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: March 18, 2013 Received: March 19, 2013
Dear Dr. Ibrahim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
## Page 2 - Samir Ibrahim, Ph.D., RAC
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erin Keith
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(k) Number: K113871
Device Name:
PRO-DENSE® Bone Graft Substitute
Indications For Use:
PRO-DENSE® resultant paste is intended for use as a bone graft substitute to be injected or digitally packed into open bone voids/gaps that are not intrinsic to the stability of bony structure of the scheral system (i.e., the extremities and pelvis) to cure in situ. These open bone voids may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), surgically created osseous or oefers or osseous defects created from traumatic injury to the bone. The paste provides a bone graft substitute that resorbs and is replaced with bone during the healing process.
The PRO-DENSE® paste cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K Wires) to help support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support the carder healing process.
PRO-DENSE® is provided sterile for single use only.
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number
Prescription Use V (Per21 CFR 801.109)
OR
Over-The Counter Use (Optional Format 1-2-96)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Laurence D: Coyne -A
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K113871