The VyBrate-L™ VBR System is intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e. fracture). The VyBrate-L™ VBR device is intended for use with autograft and/or allograft bone graft material and must be used with supplemental fixation systems. The VyBrate-C™ VBR System is a vertebral body replacement device indicated for use in the cervical spine (C3-C7 vertebral bodies) in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor, fracture, or osteomyelitis, or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in cervical degenerative disorders. The VyBrate-C™ VBR System is intended to be used with supplemental fixation cleared by the FDA for use in the cervical spine. These implants are intended for use with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, as an adjunct to fusion. The VyBrate-C™ VBR System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon’s discretion.
Device Story
Spinal spacer implant; inserts between vertebral bodies in anterior column of thoracolumbar or cervical spine; replaces collapsed, damaged, or unstable vertebral bodies; used in surgical setting by orthopedic or neurosurgeons; requires supplemental fixation; provides structural support to spinal column; facilitates fusion; restores spinal integrity in advanced tumor cases where fusion is not achievable; manufactured from OXPEKK with Tantalum markers.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical mechanical testing (static/dynamic axial compression, torsion, compressive shear, expulsion, and subsidence) performed on the predicate device (K142005) as a worst-case representative.
Technological Characteristics
Spinal vertebral body replacement spacer; material: OXPEKK with Tantalum markers; non-software, non-electrical; mechanical fixation via supplemental systems; intended for use with autograft/allograft bone graft.
Indications for Use
Indicated for skeletally mature patients requiring vertebral body replacement in the thoracolumbar (T1-L5) or cervical (C3-C7) spine due to tumor, trauma (fracture), or osteomyelitis; also for reconstruction following corpectomy for cervical degenerative disorders; intended for use with autograft/allograft bone graft and supplemental fixation.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
K050850 — XPAND CORPECTOMY SPACER · Globus Medical, Inc. · May 26, 2005
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 7, 2026
Vy Spine, LLC
Jordan Hendrickson
Quality Operations Manager
545 W 500 South
Suite 100
Bountiful, Utah 84010
Re: K253158
Trade/Device Name: VyBrate™ VBR System
Regulation Number: 21 CFR 888.3060
Regulation Name: Spinal Intervertebral Body Fixation Orthosis
Regulatory Class: Class II
Product Code: MQP, PLR
Dated: October 14, 2025
Received: October 14, 2025
Dear Mr. Hendrickson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253158 - Jordan Hendrickson
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253158 - Jordan Hendrickson
Page 3
Sincerely,
KATHERINE D.
KAVLOCK-S
for
Brent Showalter, Ph.D.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K253158
Device Name
VyBrate™ VBR System
Indications for Use (Describe)
The VyBrate-L™ VBR System is intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e. fracture). The VyBrate-L™ VBR device is intended for use with autograft and/or allograft bone graft material and must be used with supplemental fixation systems.
The VyBrate-C™ VBR System is a vertebral body replacement device indicated for use in the cervical spine (C3-C7 vertebral bodies) in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor, fracture, or osteomyelitis, or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in cervical degenerative disorders. The VyBrate-C™ VBR System is intended to be used with supplemental fixation cleared by the FDA for use in the cervical spine.
These implants are intended for use with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, as an adjunct to fusion. The VyBrate-C™ VBR System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon’s discretion.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
K253158
Page 1 of 2
510(k) Summary
Vy Spine, LLC
545 W 500 South
Suite 100
Bountiful, UT 84010
Telephone: 866-489-7746
Fax: 801-294-0079
7 January 2026
Contact: Jordan Hendrickson
Operations Manager
510(k) Number:
Common or Usual Name: Spinal Vertebral Body Replacement Device
Proposed Proprietary or Trade Name: VyBrate™ VBR System
Classification Name: Spinal Intervertebral Body Fixation Orthosis
Regulation Number: 21 CFR 888.3060
Product Code: MQP, PLR
Regulatory Classification: Class II
Substantial Equivalence
The VyBrate™ VBR System is substantially equivalent to the primary predicate device Oxford Performance Materials SpineFab VBR (K142005) and the secondary predicate devices NuVasive Monolith Cervical Corpectomy (K180550), as well as the reference predicates ClariVy™ Cervical IBF System (K231836) and LumiVy™ Lumbar IBF (K233807). The VyBrate™ VBR System is equivalent to these commercially available devices in terms of material, intended use, levels of attachment, size range, and use with supplemental fixation.
Device Description
The Vy Spine™ VyBrate™ VBR System is comprised of implants components. The implant component, the VyBrate™ VBR device, is a spacer, which inserts between vertebral bodies in the anterior column of the thoracolumbar or cervical spine. The spacer is made of OXPEKK with Tantalum markers.
Intended Use/Indications for Use
The VyBrate-L™ VBR System is intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e. fracture). The VyBrate-L™ VBR device is intended for use with autograft and/or allograft bone graft material and must be used with supplemental fixation systems.
The VyBrate-C™ VBR System is a vertebral body replacement device indicated for use in the cervical spine (C3-C7 vertebral bodies) in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor, fracture, or osteomyelitis, or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in cervical degenerative disorders. The VyBrate-C™ VBR System is intended to be used with supplemental fixation cleared by the FDA for use in the cervical spine.
{5}
K253158
Page 2 of 2
These implants are intended for use with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, as an adjunct to fusion. The VyBrate-C™ VBR System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon’s discretion.
## Non-Clinical Testing
The Oxford Performance Materials SpineFab VBR (K142005) has undergone Non-Clinical Testing using various ASTM Standards at a third party facility. Mechanical testing was performed on the Oxford Performance Materials SpineFab VBR following ASTM F-2077, ASTM F-2267, and ASTM Draft F-04.25.02.02 standards. These mechanical tests included static axial compression, dynamic axial compression, static axial torsion, dynamic axial torsion, static compressive shear, dynamic compressive shear, expulsion, and subsidence. Sterilization validations have also been performed. The SpineFab testing was used because it was determined to be worst-case compared to the subject VyBrate™ VBR System. The subject VyBrate™ VBR System has the same material, design, sizes, indication of use & biocompatibility as the predicate devices.
## Technological Characteristics
The subject Vy Spine™ VyBrate™ VBR System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to have equivalent technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, and function. This device does not contain software or electrical equipment.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.