MODIFICATION TO VERTE-STACK

K040422 · Medtronic Sofamor Danek · MQP · Mar 19, 2004 · Orthopedic

Device Facts

Record IDK040422
Device NameMODIFICATION TO VERTE-STACK
ApplicantMedtronic Sofamor Danek
Product CodeMQP · Orthopedic
Decision DateMar 19, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The VERTE-STACK™ Spinal System is a vertebral body replacement device intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). The VERTE-STACK™ Spinal System is to be used with supplemental fixation. Specifically, the VERTE-STACK™ device is to be used with the Medtronic Sofamor Danek ZPLATE II Anterior Fixation System, the DYNALOK™ CLASSIC Spinal System, the VANTAGE™ Anterior Fixation System, the TSRH® Spinal System, the CD HORIZON® Spinal System, the GDLH® Spinal System and/or their successors. Additionally, the VERTE-STACK™ device is intended to be used with bone graft.

Device Story

VERTE-STACK™ is a stackable spinal spacer for vertebral body replacement in the thoracolumbar spine (T1-L5). Device components are inserted between vertebral bodies following corpectomy to provide structural support. System consists of stackable components of varying sizes/heights to match patient anatomy. Fabricated from PEEK OPTIMA LT with tantalum markers or titanium alloy. Not a stand-alone device; requires supplemental anterior/posterior spinal instrumentation (e.g., ZPLATE II, DYNALOK, VANTAGE, TSRH, CD HORIZON, GDLH) and bone graft. Titanium constructs incompatible with stainless steel supplemental fixation. Used by surgeons in clinical settings to stabilize the spine after tumor or trauma-induced vertebral collapse.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Stackable vertebral body replacement spacer. Materials: PEEK OPTIMA LT with tantalum markers or titanium alloy. Dimensions: variable heights/sizes. Connectivity: None. Energy source: None. Sterilization: Not specified. Software: None.

Indications for Use

Indicated for patients with collapsed, damaged, or unstable vertebral bodies in the thoracolumbar spine (T1-L5) resulting from tumor or trauma (fracture). Must be used with supplemental fixation and bone graft.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAR 1 9 2004 ## VERTE-STACK™ Spinal System 510(k) Summary February 2004 K040422 10F2 - Medtronic Sofamor Danek I. Company: 1800 Pyramid Place Memphis, TN 38132 (901) 396-3133 #### Proprietary Trade Name: VERTE-STACK™ Spinal System II. ### Classification Name: Spinal Intervertebral Body Fixation Orthosis (21 CFR III. 888.3060) #### III. Product Description The VERTE-STACK™ device is a stackable spacer, which inserts between vertebral bodies in the anterior thoracic and lumbar spine. The device is intended for vertebral body replacement to aid in the surgical correction and stabilization of the spine. The construct is not intended to be employed as a stand-alone device. The VERTE-STACK™ device is fabricated and manufactured from POLYETHERETHERKETONE (PEEK OPTIMA LT) along with a tantalum marker. Alternatively the device may be manufactured from titanium alloy. The design of the VERTE-STACK™ device includes a variety of stackable components of different sizes and heights. The stackable components are designed to suit the individual patient pathology. The VERTE-STACK™ device may be used individually, or two or more may be stacked together in order to accommodate the individual anatomical requirements of the vertebral space created by the corpectomy. The VERTE-STACK™ Spinal System must be used with additional anterior and/or posterior spinal instrumentation to augment stability. VERTE-STACK™ constructs manufactured with PEEK may be used with stainless steel or titanium supplemental fixation devices. Titanium VERTF-STACK™ constructs may not be used with stainless steel supplemental fixation devices. The purpose of this submission is to include a titanium version of previously cleared VERTE-STACKTM Spinal System components. 000089 {1}------------------------------------------------ #### Indications IV. The VERTE-STACK™ Spinal System is a vertebral body replacement device intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). The VERTE-STACK™ Spinal System is to be used with supplemental fixation. Specifically, the VERTE-STACK™ device is to be used with the Medtronic Sofamor Danck ZPLATE II Anterior Fixation System, the DYNALOK™ CLASSIC Spinal System, the VANTAGE™ Anterior Fixation System, the TSRH® Spinal System, the CD HORIZON® Spinal System, the GDLH® Spinal System and/or their successors. Additionally, the VERTE-STACK™ device is intended to be used with bone graft. #### Substantial Equivalence V. Documentation was provided which demonstrated the VERTE-STACK™ Spinal System to be substantially equivalent to the previously cleared VERTE-STACK™ Spinal System (K031780, SE 07/30/03) and to the VERTE-SPAN™ Spinal System (K010930, SE 10/24/01). {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Public Health Service MAR 1 9 2004 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Richard Treharne, Ph.D. Senior Vice President, Regulatory Affairs Medtronic Sofamor Danek Memphis, Tennessee 38132 Re: K040422 Trade/Device Name: VERTE-STACK™ Spinal System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: MQP Dated: February 17, 2004 Received: February 18, 2004 Dear Dr. Treharne: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 .. Richard Treharne, Ph.D. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, sincerely yours, Mark A. Williams Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 0404da 510(k) Number (if known): Device Name:__________________________________________________________________________________________________________________________________________________________________ ## Indications for Use: The VERTE-STACK™ Spinal System is a vertebral body replacement device intended for use in the thoracolumbar spine (T1-L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). The VERTE-STACK™ Spinal System is to be used with supplemental fixation. Specifically, the VERTE-STACK™ device is to be used with the Medtronic Sofamor Danek ZPLATE II Anterior Fixation System, the DYNALOK™ CLASSIC Spinal System, the VANTAGE™ Anterior Fixation System, TSRH® Spinal System, CD HORIZON® Spinal System, the GDLH® Spinal System, or their successors. Additionally, the VERTE-STACK™ device is intended to be used with bone graft. # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use ***_***_ and/or (Optional 1-2-96) Mark A. Millican Review Sign-Off Division of General, Restorative, and Neurological Devices Number K040122 mxm
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...