AESCULAP SPINE SYSTEM EVOLUTION

K980484 · Aesculap, Inc. · MNI · Oct 2, 1998 · Orthopedic

Device Facts

Record IDK980484
Device NameAESCULAP SPINE SYSTEM EVOLUTION
ApplicantAesculap, Inc.
Product CodeMNI · Orthopedic
Decision DateOct 2, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

When used as a pedicle screw fixation system, Aesculap's Spine System® Evolution is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). It is also intended for the treatment of severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skelatally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine with removal of the implants after the attainment of a solid fusion. Levels of pedicle screw fixation for this indication are from L3-S1. When used as a non-pedicle screw system, Aesculap's Spine System® Evolution is indicated for use in patients with degenerative disc disease (defined as back pain of discongenic origin with degeneration of the disc confirmed by history and radiographic studies), kyphosis, spondylolisthesis, neurological scoliosis, spine tumors and fractures. It is intended for posterior fixation from levels T1 through S1.

Device Story

Spinal fixation system consisting of monoaxial/polyaxial pedicle screws, rods, linking plates, and lateral connectors; used to stabilize thoracic, lumbar, and sacral vertebrae to promote fusion. Implants are surgically placed by physicians in an OR setting. System provides mechanical immobilization of spinal segments. Benefits include stabilization of acute/chronic spinal instabilities or deformities. No software or electronic components.

Clinical Evidence

Bench testing only. Components subjected to static and fatigue mechanical testing per ASTM Standard F 1717-96 (Corpectomy Model). No clinical data provided.

Technological Characteristics

Materials: Titanium alloy (Ti6Al4V) per ISO 5832/III for implants; surgical grade stainless steel per ISO 7153/1 for instrumentation. Mechanical fixation system. No software, electronics, or energy sources.

Indications for Use

Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion. Conditions: degenerative spondylolisthesis with neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, failed fusion, and degenerative disc disease. Pedicle screw fixation levels: L3-S1 (or L2-S1 per indication statement). Non-pedicle fixation levels: T1-S1.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows a handwritten string of characters. The characters are 'K980484'. The characters are written in a cursive style. The image is in black and white. **OCT** 2 1998 Page 1 of 2 ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS in Accordance with SMDA of 1990 # AESCULAP SPINE SYSTEM® EVOLUTION September 25, 1998 ## Company Contact Aesculap®, Inc. 1000 Gateway Blvd. So. San Francisco, CA 94080 Lia S. Jones, Regulatory Associate Phone: 650-876-7000 x 350 650-589-3007 FAX: #### Trade Name Aesculap Spine System® Evolution ### Common Name Spinal Fixation System ### Product Code and Regulatory Classification | MNI: | 888.3070 | Pedicle Screw Spinal Fixation Orthosis | |------|--------------|-------------------------------------------------| | MNH: | unclassified | Spondylolisthesis Spinal Fixation Device System | | KWP: | 888.3050 | Spinal Interlaminal Fixation Orthosis | ### Product Classification Class II ### Intended Use When used as a pedicle screw fixation system, Aesculap's Spine System® Evolution is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). It is also intended for the treatment of severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skelatally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine with removal of the implants after the attainment of a solid fusion. Levels of pedicle screw fixation for this indication are from L3-S1. When used as a non-pedicle screw system, Aesculap's Spine System® Evolution is indicated for use in patients with degenerative disc disease (defined as back pain of discongenic origin with degeneration of the disc confirmed by history and radiographic studies), kyphosis, spondylolisthesis, neurological scoliosis, spine tumors and fractures. It is intended for posterior fixation from levels T1 through S1. 0002 {1}------------------------------------------------ Page 2 of 2 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS in Accordance with SMDA of 1990 # AESCULAP SPINE SYSTEM® EVOLUTION ### Device Description Spine System® Evolution is a multiple component system comprised of a variety of single use devices that allow the surgeon to build a spinal implant construct in order to stabilize the thoracic and lumbar vertebrae and promote spinal fusion. Spine System® Evolution consists of pedicle screws (monoaxial and polyaxial), rods, linking plates, lateral connectors, and other implant components used to lock the desired construct in place. The spinal implants are manufactured from titanium alloy (Ti6Al4V) in accordance to ISO 5832/III. The specialized instrumentation used to implant and explant the Spine System® Evolution implants are made from surgical grade stainless steel in accordance to ISO 7153/1. ### Performance Data No applicable performance standards have been promulgated under Section 514 of the Food, Drug and Cosmetic Act for these devices. However, the components that comprise Spine System® Evolution were subjected to mechanical testing according to the ASTM Standard F 1717-96 (Static and Fatigue Test Methods for Spinal Implant Constructs in a Corpectomy Model). ## Substantial Equivalence Aesculap believes that the Spine System® Evolution additional components presented in this submission are substantially equivalent in design, material composition, function and intended use to currently marketed spinal implant systems, such as: - Spine System® . bv Aesculap (K962757, K953599, K935113) - CD Horizon™ Spinal System . bv Sofamor Danek (K964159, K962708, K961633) - . Mirage™ Spinal System by Alphatec (K951846) - Moss® Miami Titanium Spinal System . by Depuy Motech ™(K980447, K955348) - Synergy™ Posterior Spinal System . by Cross® Medical Products, Inc. (K973836, K950099, K940631) 0003 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized caduceus symbol. The caduceus is depicted with a modern, abstract design, consisting of three intertwined figures. The text is arranged in a circular pattern around the symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 2 1998 OCT Ms. Lia S. Jones Regulatory Associate Aesculap®, Inc. 1000 Gateway Boulevard South San Francisco, California 94080-7030 Re: K980484 Spine System® Evolution Regulatory Class: II Product Codes: MNI, KWP, and MNH Dated: August 20, 1998 Received: August 21, 1998 Dear Ms. Jones: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - Lia S. Jones This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure i 11 ll {4}------------------------------------------------ Page 1 of 1 # INDICATION FOR USE STATEMENT 510(k) Number (if known): K980484 Device Name: ### Aesculap Spine System® Evolution ### Indication for Use: When used as a pedicle screw fixation system, Aesculap's Spine System® Evolution is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis). It is also intended for the treatment of severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra in skelatally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine with removal of the implants after the attainment of a solid fusion. Levels of pedicle screw fixation for this indication are from L2-S1. When used as a non-pedicle screw system, Aesculap's Spine System® Evolution is indicated for use in patients with degenerative disc disease (defined as back pain of discongenic origin with degeneration of the disc confirmed by history and radiographic studies), kyphosis, spondylolisthesis, neurological scoliosis, spine tumors and fractures. It is intended for posterior fixation from levels T1 through S1. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | | (Division Sign-Off) | |---------------|-----------------------------------------| | | Division of General Restorative Devices | | 510(k) Number | K980484 | | Prescription Use (per 21 CFR 801.109) | X | |---------------------------------------|---| |---------------------------------------|---| or | Over-the-Counter Use | | |----------------------|--| |----------------------|--| (Optional Format 1-2-96) 0001
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...