MODIFICATION TO MOSS MIAMI SPINAL SYSTEM POLYAXIAL SCREWS
Device Facts
| Record ID | K030383 |
|---|---|
| Device Name | MODIFICATION TO MOSS MIAMI SPINAL SYSTEM POLYAXIAL SCREWS |
| Applicant | Depuyacromed |
| Product Code | MNH · Orthopedic |
| Decision Date | Feb 26, 2003 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 888.3070 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
When used as a posterior, noncervical hook, and/or sacral/iliac screw fixation system, or as an anterior, thoracic/lumbar screw fixation system, the Moss Miami Spinal System is intended to treat scoliosis, kyphosis and lordosis, fracture, loss of stability due to tumor, spinal stenosis, spondylolisthesis, a previously failed fusion surgery or degenerative disc disease (i.e., discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies). When used as a pedicle screw fixation system of the noncervical spine in skeletally mature patients, the Moss Miami Spinal System is indicated for deqenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The Moss Miami Spinal System is also indicated for pedicle screw fixation in skeletally mature patients with severe spondylolisthesis (Grades 3 and 4) at the L5 -- S1 vertebral joint, having fusions with autogenous bone graft. with the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation are L3 - S1), and for whom the device system is intended to be removed after solid fusion is attained.
Device Story
Moss Miami Spinal System Polyaxial Screws are orthopedic implants designed to stabilize the noncervical spine. The system consists of polyaxial screws and polyaxial reduction screws configured to accept 5.5mm rods. Used by surgeons in clinical settings to provide rigid fixation during spinal fusion procedures. The device acts as a mechanical construct to restore spinal stability, treat deformities, and facilitate bone fusion. It is intended for use with autogenous bone graft; the system is designed for removal after solid fusion is achieved in specific severe spondylolisthesis cases.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design, intended use, and technological characteristics compared to predicate devices.
Technological Characteristics
Polyaxial screw and rod fixation system. Designed to accept 5.5mm rods. Mechanical spinal fixation orthosis. Class II device. Materials and specifications are consistent with previously cleared Moss Miami components.
Indications for Use
Indicated for skeletally mature patients requiring noncervical spinal fixation for scoliosis, kyphosis, lordosis, fracture, spinal tumor, spinal stenosis, spondylolisthesis, degenerative disc disease, or failed fusion. Pedicle screw fixation specifically indicated for degenerative spondylolisthesis with neurologic impairment, dislocation, and severe spondylolisthesis (Grades 3-4) at L5-S1.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Predicate Devices
- Moss Miami Spinal System (K953915)
- Moss Miami Spinal System (K982320)
- Moss Miami Spinal System (K982511)
- Moss Miami Spinal System (K982011)
- Moss Miami Spinal System (K983583)
- Moss Miami Spinal System (K992168)
- Moss Miami Spinal System (K011182)