NOVEL TAPERED HIP STEM

K103479 · Stryker Corp. · MEH · Mar 10, 2011 · Orthopedic

Device Facts

Record IDK103479
Device NameNOVEL TAPERED HIP STEM
ApplicantStryker Corp.
Product CodeMEH · Orthopedic
Decision DateMar 10, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The indications for use of the total hip replacement prostheses include: - Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis; 1) - Rheumatoid arthritis 2) - 3) Correction of functional deformity; - Revision procedures where other treatments or devices have failed; and, 4) - 2) Nonunions, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. Additional indication specific to use of Accolade II Femoral Stems with compatible Howmedica Osteonics Constrained Liners: - 1. When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability, Accolade II Femoral Stems are intended for cementless use only and are intended for total and hemiarthroplasty procedures.

Device Story

Accolade II Femoral Hip Stem is a tapered, non-porous coated femoral component for cementless, press-fit total or hemiarthroplasty. Device features variable medial curvature to facilitate press-fit stability and load transmission; reduced stem length to accommodate various surgical approaches (e.g., direct approach). Manufactured from Ti-6Al-4V substrate with CP Titanium and hydroxylapatite (HA) coating. Used by orthopedic surgeons in clinical settings; implanted into the femur. Compatible with Howmedica Osteonics femoral heads, acetabular components, and constrained liners. Output is a restored hip joint structure; intended to alleviate pain and restore function. Clinical benefit derived from improved femoral morphology fit and intra-operative insertion ease.

Clinical Evidence

Bench testing only. Included range of motion analysis, proximal fatigue testing, and distal fatigue testing. No clinical data provided.

Technological Characteristics

Ti-6Al-4V substrate; CP Titanium and hydroxylapatite (HA) coating. Cementless, press-fit design. 12 sizes; 127° and 132° neck angles. Sterile, single-use. Compatible with V40, C-Taper, Universal, and Unitrax femoral heads.

Indications for Use

Indicated for patients requiring primary or revision total hip arthroplasty or hemiarthroplasty due to noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, revision of failed prior treatments, or proximal femoral fractures/nonunions. Specific subset indicated for use with constrained liners in patients at high risk of dislocation due to prior history, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K103479 page 182 # 510(k) Summary Proprietary Name: Novel Tapered Hip Stem (aka Accolade II Femoral Hip Stem) MAR 1 0 2011 Common Name: Hip prosthesis ## Classification Name and Reference: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis, 21 CFR §888.3353 Regulatory Class: Class II ### System Product Codes: 87 MEH - prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calcium-phosphate 87 LZO - prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented 87 LWJ - prosthesis, hip, semi-constrained, metal/polymer, uncemented - 87 K WZ prosthesis, hip, constrained, cemented or uncemented, metal/polymer 87 KWL - prosthesis, hip, hemi-, femoral, metal 87 KWY - Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis 87 JDI - prosthesis, hip, semi-constrained, metal/polymer, cemented 87 LPH - prosthesis, hip, semi-constrained, metal/polymer, porous uncemented | Contact Person: | Estela Celi, Regulatory Affairs Associate | |-----------------|-------------------------------------------| | | Howmedica Osteonics Corp. | | | 325 Corporate Drive | | | Mahwah, NJ 07430 | | | Phone: (201) 831-6461 Fax: (201) 831-3461 | Date Prepared: March 9, 2011 Legally Marketed Device to Which Substantial Equivalence is Claimed: Accolade TMZF Hip Stem: K994366, K020572 & K023102 Restoration Modular Hip Stem: K013106 ### Description: Howmedica Osteonics is introducing the Accolade II Femoral Hip Stem, a tapered non-porous coated femoral stem intended for cementless, press-fit application. The basic design of the Accolade II Femoral Hip Stem is similar to other total hip stems commercially distributed such as the Accolade TMZF Hip Stem. There are the design of the subject hip stem. such as a variable medial curvature which increases with each size in order to faciliate press fit stability and load transmission in the proximal region of the stem geometry is designed to address variations in patient femoral morphology. The overall stem length has been reduced, compared to the Accolade TMZF hip stem, to facilitate intra-operative stem insertion. The stem is manufactured from a Ti-6Al-4V substrate materially Purc (CP) Titanium coating and Purefix hydroxylapatite (HA) coating identical to the previously cleared 2 Piece Modular Hip Stem (K013106). The 2 Piece Modular Hip Stem (K013106) is now marketed and will be referred to throughout this submission as the Restoration Modular Hip Stem. The Accolade II Femoral Hip Stem will be available in 12 sizes ranging from size 0 through 11 with two neck angles (127° and 132°) that provide dual head offsets. The stem is designed for use with the currently available compatible Howmedica Osteonics' femoral heads and their compatible acetabular components. All instrumentation is listed in the Surgical Protocol attached. The Accolade II Femoral Hip Stem is a sterile, single-use in primary and revision total hip arthroplasty to alleviate pain and restore function. The subject hip stem is compatible with V40 heads, C-1aper {1}------------------------------------------------ page 2 of 2 heads when used with the V40/C-Taper Adaptor Sleeve, Universal Heads when used with the V40/Universal Adaptor Sleeve and Unitrax Heads. # Indications: The indications for use of the total hip replacement prostheses include: - Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis; 1) - Rheumatoid arthritis 2) - 3) Correction of functional deformity; - Revision procedures where other treatments or devices have failed; and, 4) - 2) Nonunions, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. ## Additional indication specific to use of Accolade II Femoral Stems with compatible Howmedica Osteonics Constrained Liners: - 1. When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability, Accolade II Femoral Stems are intended for cementless use only and are intended for total and hemiarthroplasty procedures. ## Technological Characteristics The technological characteristics are basically the same as the predicates identified in the Legally Marketed Devices to which substantial equivalence is claimed with differences noted in design of the subject hip stem such as variable medial curvature, distal lateral reduction in overall stem length to accommodate various surgical approaches, specifically the direct approach. Stem lengths are within the range of legally marketed predicates. ### Substantial Equivalence: The Accolade II Femoral Hip Stem is substantially equivalent to other commercially available hip stems in regard to intended use, design, materials and operational principles as a hip prosthesis. The following devices are examples of predicate systems: Accolade TMZF Hip Stem and Restoration Modular Hip Stem. Based upon the mechanical testing, the Accolade II Femoral Hip Stem is substantially equivalent for its intended use to other press-fit femoral replacement hip stems currently on the market. # Summary of Non-Clinical Testing: Non-clinical testing bench testing was provided and included Range of Motion analysis, proximal fatigue, and distal fatigue testing. All of the observed results indicate that the Accolade II Hip System is substantially equivalent to devices currently marketed. Therefore, the subject device is as safective, and performs at least as safely and effectively as legally marketed predicates identified. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines representing the wings and a stylized representation of the body and legs. Food and Drug Administration 1 0903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAR 2 1 2011 Howmedica Osteonics Corp. % Ms. Estela Celi Regulatory Affairs Associate 325 Corporate Drive Mahwah, New Jersey 07430 Re: K103479 Trade/Device Name: Novel Tapered Hip Stem Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: MEH, LZO, LWJ, KWZ, KWY, KWL, JDI, LPH Dated: February 18, 2011 Received: February 22, 2011 Dear Ms. Celi: This letter corrects our substantially equivalent letter of March 10, 2011. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). Vo may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements af the Act {3}------------------------------------------------ Page 2 - Ms. Estela Celi or any Federal statutes and regulations administered by other Federal agencies. You must COMPLY with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHODRHOffices/uccn1118800.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CCFR Previous 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, A.J. V.S. D. te Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K103479 Device Name: Novel Tapered Hip Stem (aka Accolade II Femoral Hip Stem) Indications for Use: The indications for use of the total hip replacement prostheses include: - Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis; 1) - Rheumatoid arthritis 2) - Correction of functional deformity; 3) - Revision procedures where other treatments or devices have failed; and, 4) - Nonunions, femoral neck and trochanies of the proximal femur with head involvement that are unmanageable રે) using other techniques. Additional indication specific to use of Accolade II Femoral Stems with compatible Howmedica Oseonics Constrained Liners: - When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use l. in primary or revision patients at hip distocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability. Accolade II Femoral Stems are intended for cementless use only and and hemiarthroplasty procedures. Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) for M. Makerson (Division Sign Only) Division of Syrgical, Orthopedic, and Restorative Devices 510(k) Number K103479
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...