MITEK ABSORBABLE POLYDIOXANONE (PDS) H FIX

K970119 · Mitek Products · MBI · Jan 30, 1998 · Orthopedic

Device Facts

Record IDK970119
Device NameMITEK ABSORBABLE POLYDIOXANONE (PDS) H FIX
ApplicantMitek Products
Product CodeMBI · Orthopedic
Decision DateJan 30, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

Mitek meniscal fasteners "H" Fix PDS are intended for use in the repair or torn meniscal tissue in the red/red zone of the meniscus (outer 1/3) as well as the repair of a separation of the meniscus from the synovium

Device Story

Mitek Absorbable "H" Device is a cylindrical, "H"-shaped polymeric implant made of polydioxanone (PDS). It is used by surgeons to repair meniscal tears in the knee. The device is delivered via a cannulated needle applicator; once placed across the tear, the device's geometry and material spring forces provide compressive holding force to maintain tissue apposition. This facilitates natural healing. The device is absorbable via hydrolysis, with strength retention profiles documented in animal studies. It is intended for prescription use only.

Clinical Evidence

Bench testing only. Pull tests performed on cadaveric meniscus tissue resulted in an average failure load of 12.06 pounds. Animal implantation studies were cited regarding the degradation and strength retention characteristics of the PDS polymer.

Technological Characteristics

Cylindrical "H"-shaped implant; material: polydioxanone (PDS) polymer; mechanism: mechanical compression via material spring forces; delivery: cannulated needle applicator; sterilization: supplied sterile, do not re-sterilize.

Indications for Use

Indicated for repair of torn meniscal tissue in the red/red zone (outer 1/3) or meniscal-synovial separation. Contraindicated for cases requiring tissue approximation under stress beyond 6 weeks.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} JAN 30 1996 K970119 # SUMMARY OF SAFETY AND EFFECTIVENESS ## MITEK ABSORBABLE "H" DEVICE ### SUBMITTER | NAME | Mitek Products | | --- | --- | | ADDRESS | 60 Glacier Drive, Westwood, MA 02090 | | TEL # | 617-461-9700 | | CONTACT | Edward F. Kent Vice President Regulatory Affairs | | DATE | January 10, 1997 | ### NAME OF DEVICE | CLASSIFICATION NAME | | | --- | --- | | COMMON NAME | | | PROPRIETARY NAME | MITEK ABSORBABLE "H" DEVICE | ### PREDICATE DEVICE | ACUFEX "T" FIX DEVICE | | --- | | PDS II SUTURE | ### DESCRIPTION OF DEVICE #### FUNCTION To hold outer 1/3 (meniscal) soft tissue in apposition during the healing time period. #### DEVICE DESIGN The "H" Fix is cylindrical in cross section and configured to have one leg of the letter "H" in a 90 degree opposed position to the other. #### MATERIALS USED The Mitek absorbable "H" Fix is made from PDS (polydioxanone) polymer. #### INTENDED USE The device is surgically placed across (through) a tear in the meniscal tissue of the knee with a needle applicator and released. The released "H" Fix produces a holding (compressive) force across the tear causing tissue apposition, which allows the natural healing process (at the tear) to occur. The tear in the meniscus for repair with the "H" PDS device should be located in the outer 1/3 zone of the meniscus,i.e., red/red zone. #### COMPARISON TO PREDICATE DEVICE(S) The Mitek absorbable "H" Fix device is similar to the Acufex T-Fix device in function. The T-Fix is surgically placed through a tear in the knee meniscus, toggled and then tied with sutures over external tissues to hold the T-Fix in place. The "H" Fix is placed through the tear in a cannulated needle and extended into the tissue, the spring forces in the material are induced by proper placement in the tissue and device length. #### DESCRIPTION OF NON CLINICAL TESTS The Mitek absorbable "H" Fix pull tests from cadaver meniscus provided an average failure load of 12.06 pounds. {1} # MITEK MENISCAL FASTENER Catalog No. XXXXXX Catalog No. YYYYYY ## DESCRIPTION Mitek Meniscal Fasteners are molded polymeric implants used to repair tears in the meniscus. Fasteners provide the means for firmly securing together the edges of a tear in the meniscus or across the synovial meniscal junction. The fasteners are available in absorbable PDS (polydioxanone). Studies of the polydioxanone polymer have found it to be nonantigenic, nonpyrogenic, and it elicits only a slight tissue response during healing. Mitek instruments - Applicator, probe and rasp used to assist in the procedure of measuring, preparing and inserting the Mitek Meniscal Fastener. ## ACTIONS ### PDS (Polydioxanone) The absorption characteristics of polydioxanone have been extensively studied. The general mechanism by which in vivo degradation of polydioxanone occurs is hydrolysis. The in vivo hydrolysis of polydioxanone cleaves the ester linkages and initially results in the loss of tensile strength. Ultimately the hydrolyzed polymer is eliminated from the body, primarily in urine. Studies conducted on devices made from polydioxanone polymer indicate minimal absorption until about the 90th post-implantation day. The results of implantation studies of polydioxanone molded articles in animals indicate that approximately 85% of its original strength remains 5 weeks after implantation. At 7 weeks post implantation approximately 12% of its original strength is retained. ## CONTENTS A Meniscal Fastener secured in a delivery needle is supplied sterile. Contents are sterile unless the package is damaged or opened. Do not re-sterilize PDS (polydioxanone). ## INDICATIONS Mitek Meniscal Fasteners are intended for the repair of torn meniscal tissue in the red/red zone of the meniscus (outer 1/3) as well as the repair of a separation of the meniscus from the synovium. ## CONTRAINDICATIONS These fasteners (being absorbable) are not to be used where prolonged (beyond 6 weeks) approximation of tissue under stress is required. ## PRECAUTIONS 1. Care should be taken to avoid damage when handling the fasteners. Avoid the crushing or crimping of the fasteners by the application of needle holders or forceps. 2. A surgeon should not attempt clinical use of the Mitek Meniscal Fasteners before reviewing instructions for use and rehearsing the installation procedure. 3. Users should exercise caution when handling surgical needles to avoid inadvertent needle sticks. 4. Discard used needles in sharps containers. {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 30 1998 Mr. Edward F. Kent Vice President, Regulatory Affairs Mitek Products 60 Glacier Drive Westwood, Massachusetts 02090 Re: K970119 Mitek Absorbable Polydioxanone "H" Device Regulatory Class: II Product Code: MBI Dated: January 8, 1998 Received: January 9, 1998 Dear Mr. Kent: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3} Page 2 - Mr. Edward F. Kent This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4} Page ___ of ___ 510(k) Number (if known): K970119 Device Name: Mitek "H" Fix PDS device Indications For Use: Mitek meniscal fasteners "H" Fix PDS are intended for use in the repair or torn meniscal tissue in the red/red zone of the meniscus (outer 1/3) as well as the repair of a separation of the meniscus from the synovium (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General Restorative Devices 510(k) Number K970119 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...