MULTITAK SS SUTURE SYSTEM
Device Facts
| Record ID | K964532 |
|---|---|
| Device Name | MULTITAK SS SUTURE SYSTEM |
| Applicant | Bonutti Research |
| Product Code | MBI · Orthopedic |
| Decision Date | Feb 3, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3040 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The devices are intended for soft tissue to bone suture fixation for the following indications: Shoulder - Rotator cuff tear repairs - Biceps tenodesis Foot and Ankle - Lateral instability repairs/reconstructions Elbow - Ulnar or radial collateral ligament reconstructions - Tennis elbow repair - Biceps tendon reattachment Knee - Extra-capsular repairs: - Medial collateral ligament - Lateral collateral ligament
Device Story
Multitak Suture System consists of tubular bone anchor preassembled with USP size 2-0 to 2 braided polyester suture. Introduction device delivers anchor into predrilled bone hole; sutures secure soft tissue to bone. Used by surgeons in clinical settings for orthopedic soft tissue fixation. Device provides mechanical attachment of ligaments/tendons to bone to facilitate healing. Benefit: secure fixation for various orthopedic repairs.
Clinical Evidence
Bench testing only. Pullout tension testing performed in cadaver bone comparing Multitak Suture System to predicate device. Results showed performance equal to or better than predicate.
Technological Characteristics
Tubular bone anchor; braided polyester suture (USP 2-0 to 2); metallic bone fixation fastener (21 CFR 888.3040).
Indications for Use
Indicated for soft tissue to bone suture fixation in shoulder (rotator cuff repair, biceps tenodesis), foot/ankle (lateral instability repair/reconstruction), elbow (ulnar/radial collateral ligament reconstruction, tennis elbow repair, biceps tendon reattachment), and knee (medial/lateral collateral ligament extra-capsular repairs).
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Multitak SS Suture System (Bonutti Research)