The Opus® LabraLock™ P bone anchor with inserter is indicated for use in fixation of soft tissue to bone. Examples of such procedures include: Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction and midfoot reconstruction Foot: Hallux valgus reconstruction Elbow: Tennis elbow repair, biceps tendon attachment Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions
Device Story
Bone anchor with inserter handle; used in arthroscopic/orthopedic procedures for soft tissue-to-bone fixation. Device manufactured from PEEK (polyether-etherketone). Operated by surgeons in clinical/OR settings. Provides mechanical fixation of tendons/ligaments to bone. Benefits patient by enabling secure tissue reattachment during reconstructive surgeries.
Clinical Evidence
No clinical data. Substantial equivalence based on design, material characterization, and comparison to legally marketed predicate devices.
Technological Characteristics
Bone anchor with inserter handle. Material: PEEK (polyether-etherketone). Mechanical fixation device. Non-degradable. Class II, product code MBI/HWC.
Indications for Use
Indicated for fixation of soft tissue to bone in orthopedic procedures including shoulder (Bankart, SLAP, rotator cuff, biceps tenodesis, deltoid repair), ankle (instability, Achilles, midfoot), foot (hallux valgus), elbow (tennis elbow, biceps attachment), and knee (extra-capsular repairs, ligament/tendon reattachment).
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Ko61349 page 4 of 2
### 510(K) SUMMARY
JUL 1 4 2006
### ARTHROCARE CORPORATION OPUS LABRALOCK P KNOTLESS FIXATION DEVICE
### General Information
| <b>Submitter Name/Address:</b> | ArthroCare Corporation<br>680 Vaqueros Avenue<br>Sunnyvale, CA 94085-3523 |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Establishment Registration No.:</b> | 2951580 |
| <b>Contact Person:</b> | Laura N. Kasperowicz<br>Sr. Manager, Regulatory Affairs |
| <b>Date Prepared:</b> | May 12, 2006 |
| <b>Device Description</b> | |
| <b>Trade Name:</b> | Opus® LabraLock™ P |
| <b>Generic/Common Name:</b> | Bone Anchor, Fastener, Fixation, Soft Tissue |
| <b>Classification Name:</b> | Fastener, Fixation, Nondegradeable, Soft Tissue<br>(Class II per 21 CFR 888.3040, Product code: MBI) |
| <b>Predicate Devices</b> | |
| | Opus MiniMagnum K042584 (Cleared 12/14/04)<br>Mitek Mini QuickAnchor Plus K992487 (Cleared 09/21/99)<br>Arthrex PEEK Pushlock K051219 (Cleared 09/27/05)<br>Mitek Bioknotless Anchor K002639 (Cleared 05/11/01) |
### Product Description
The Opus Labral.ock™ P device is a bone anchor with inserter handle designed for specific indications in arthroscopic and orthopedic procedures.
### Indications For Use
The Opus LabraLock™ P bone anchor with inserter is indicated for use in fixation of soft tissue to bone. Examples of such procedures include:
Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction and midfoot reconstruction
Foot: Hallux valgus reconstruction
Elbow: Tennis elbow repair, biceps tendon attachment
Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions
{1}------------------------------------------------
# 510(K) SUMMARY
### Substantial Equivalence
By definition, substantial equivalence means that a device has the same intended use and technical characteristics as the predicate device, or has the same intended use and different technological characteristics, but can be demonstrated to be as safe and effective as the predicate device. The Opus® Labral.ock™ P is substantially equivalent to the existing Opus® MiniMagnum™ Knotless Firation Device cleared by the Food & Drug Administration [K042584], the Mitek Mini QuickAnchor™ Plus [K992487], the Mitek Bioknotless™ Anchor [K002639] and the Arthrex PEEK™ PushLock™ Knotless Anchor [K051219]. The differences between the Opus® LabraLock™ P and the predicate devices do not raise any questions regarding the safety and effectiveness of the implant. Furthermore, the materials are well characterized and have been used in predicate devices with similar indications. The device, as designed, is as safe and effective as predicate devices.
### Summary and Reason for 510k Notification
The purpose of this 510k is to notify the Food and Drug Administration of a new product, the Opus® LabraLock™ P Knotless Fixation Device. This new product is substantially equivalent to the Opus® MiniMagnum™ Knotless Fixation Device originally cleared under K042584, but is manufactured from PEEK (polyether-etherketone) as opposed to stainless steel.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUL 14 2006
ArthroCare Corporation % Ms. Laura N. Kasperowicz Sr. Manager, Regulatory Affairs 680 Vaqueros Avenue Sunnyvale, CA 94085-3523
Re: K061349
Trade/Device Name: Opus® LabraLock™ P Knotless Fixation Device Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI, HWC Dated: May 12, 2006 Received: May 15, 2006
Dear Ms. Kasperowicz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Laura N. Kasperowicz
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or 240-276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
uberg Buchup
to
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
page 1 of 1
# INDICATIONS FOR USE STATEMENT
510(k) Number:
KO61349
Device Name:
Opus® LabraLock™ P 'Knotless Fixation Device
Indications for Use:
The Opus® LabraLock™ P bone anchor with inserter is indicated for use in fixation of soft tissue to bone. Examples of such procedures include:
Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator Duff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction and midfoot reconstruction
Foot: Hallux valgus reconstruction
Elbow: Tennis elbow repair, biceps tendon attachment
Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions
Prescription Use (Part 21 CFR 801 Subpart D) X AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)
NO
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Saiba Kucind & Kim
---
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
**510(k) Number** K061349