AUDERE Lumbar Spacer System

K193457 · Met One Technologies · MAX · May 12, 2020 · Orthopedic

Device Facts

Record IDK193457
Device NameAUDERE Lumbar Spacer System
ApplicantMet One Technologies
Product CodeMAX · Orthopedic
Decision DateMay 12, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The AUDERE Lumbar Spacer System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients with previous non-fusion spinal surgery at the involved level may be treated with the AUDERE Lumbar Spacer System. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the AUDERE Lumbar Spacer System. The AUDERE Lumbar Spacer System is designed for use with autogenous bone graft. The system is also intended to be used with supplemental fixation systems that are cleared by the FDA for use in the lumbar spine.

Device Story

Intervertebral body fusion device; used in lumbar spine (L2-S1) to facilitate fusion; implanted via posterior (PLIF) or transforaminal (TLIF) approach; used with autogenous bone graft and supplemental fixation. Device provides structural support to disc space; maintains disc height; promotes fusion. Used by surgeons in clinical/OR settings. Benefits patient by stabilizing spinal segment; reducing discogenic pain. System includes implants and trials; compatible with previously cleared instruments.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing per ASTM standards: static and dynamic axial compression (ASTM F2077), expulsion (ASTM F-04.25.02.02), and subsidence (ASTM F2267).

Technological Characteristics

Machined structural interbody spacer; materials: PEEK (ASTM F2026) and Tantalum radiographic markers (ASTM F560). Dimensions comparable to predicate devices. No energy source or software.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous lumbar levels (L2-S1) with up to Grade 1 spondylolisthesis or retrolisthesis. Includes patients with prior non-fusion surgery at the level. Requires 6 months of failed non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Met One Technologies Mr. Evan Carbonell President & CEO 513 W. San Antonio Ave, Suite C El Paso, Texas 79901 Re: K193457 Trade/Device Name: AUDERE Lumbar Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: January 24, 2020 Received: February 13, 2020 Dear Mr. Carbonell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's May 12, 2020 {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Brent Showalter, Ph.D. Assistant Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K193457 Device Name AUDERE Lumbar Spacer System #### Indications for Use (Describe) The AUDERE Lumbar Spacer System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1) with up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients with previous non-fusion spinal surgery at the involved level may be treated with the AUDERE Lumbar Spacer System. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the AUDERE Lumbar Spacer System. The AUDERE Lumbar Spacer System is designed for use with autogenous bone graft. The system is also intended to be used with supplemental fixation systems that are cleared by the FDA for use in the lumbar spine. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------| | <div> <span> <span style="text-decoration: overline;">×</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K193457 - 510(K) Summary | Date: | 20 April 2020 | |---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Sponsor: | Met One Technologies, LLC<br>513 W. San Antonio Ave Ste. C<br>El Paso, TX 79901 | | Sponsor Contact | Kyle Atwood, Vice President, Engineering | | Proposed Trade<br>Name: | AUDERE Lumbar Spacer System | | Common Name: | Interbody Fusion System | | Device Classification: | Class II | | Regulation Name,<br>Regulation Number,<br>Product Code: | Intervertebral fusion device with bone graft, lumbar, 888.3080, MAX | | Submission Purpose: | The purpose of this submission is to address dimensional changes of the<br>AUDERE Lumbar Spacer System for the purpose of improving insertion, better<br>radio-visibility and to expand the system to include 4° and 8° lordotic<br>implantable devices. | | Device Description: | The AUDERE Lumbar Spacer System consists of implants and trials that are<br>compatible with all previously cleared AUDERE instruments. The AUDERE<br>Lumbar Spacer System may be implanted bilaterally using a posterior (PLIF)<br>approach, or as a single device employing a transforaminal (TLIF) approach. | | Indications for Use: | The AUDERE Lumbar Spacer System is indicated for spinal fusion procedures in<br>skeletally mature patients with degenerative disc disease (DDD) at one or two<br>contiguous levels in the lumbar spine (L2-S1) with up to Grade 1<br>spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as<br>discogenic back pain with degeneration of the disc confirmed by history and<br>radiographic studies. Patients with previous non-fusion spinal surgery at the<br>involved level may be treated with the AUDERE Lumbar Spacer System.<br>Patients must have undergone a regimen of at least six (6) months of non-<br>operative treatment prior to being treated with the AUDERE Lumbar Spacer<br>System.<br>The AUDERE Lumbar Spacer System is designed for use with autogenous bone<br>graft. The system is also intended to be used with supplemental fixation<br>systems that are cleared by the FDA for use in the lumbar spine. | | Materials: | The AUDERE Lumbar Spacer System is comprised of a family of implants that<br>has bodies manufactured from polyetheretherketone (PEEK), in compliance<br>with ASTM F2026, and radiographic markers manufactured from Tantalum, in<br>compliance with ASTM F560. | | Primary Predicate | AUDERE Lumbar Spacer System (Met One Technologies, LLC - K160699) | | Additional Predicates | PATRIOTTM Spacer (Globus Medical Inc. – K122097)<br>Interbody System (Tyber Medical - K130573)<br>AccuLiF® TL-PEEK Cage and AccuLiF® TL and PL (CoALign Innovation – K112095) | | Performance Data: | THE AUDERE Lumbar Spacer System has been tested in the following test<br>modes:<br>- Static axial compression per ASTM F2077<br>- Dynamic axial compression per ASTM F2077<br>- Expulsion per ASTM F-04.25.02.02<br>- Subsidence per ASTM F2267<br>The results of this non-clinical testing show that the strength of the AUDERE<br>Lumbar Spacer System is sufficient for its intended use and substantially<br>equivalent to legally marketed predicate devices. | | Technological<br>Characteristics: | The modified AUDERE Lumbar Spacer System possesses the same<br>technological characteristics as one or more of the predicate devices. These<br>include:<br>- Performance (as described above),<br>- Basic design (machined structural interbody)<br>- Material (PEEK and Tantalum) and<br>- Size (dimensions are comparable to those offered by the cleared<br>devices)<br>Therefore, the fundamental scientific technology of the modified AUDERE<br>Lumbar Spacer System is the same as previously cleared devices. | | Conclusion: | The modified AUDERE Lumbar Spacer System possesses the same intended use<br>and technological characteristics as the predicate devices. Therefore, the<br>modified AUDERE Lumbar Spacer System is substantially equivalent for its<br>intended use. | {4}------------------------------------------------
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...