MONDRIAN Lumbar Interbody Fusion Cage System

K192863 · Ctl Medical Corporation · MAX · Aug 26, 2020 · Orthopedic

Device Facts

Record IDK192863
Device NameMONDRIAN Lumbar Interbody Fusion Cage System
ApplicantCtl Medical Corporation
Product CodeMAX · Orthopedic
Decision DateAug 26, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The MONDRIAN™ Lumbar Interbody Fusion Cage system is intended for use as an interbody fusion cage device to maintain lumbar intervertebral spacing and must be used with supplemental fixation. The devices are available in a variety of different sizes and configurations to accommodate anatomical variation in different vertebral levels and/or patient anatomy. The devices are made of PEEK-OPTIMA LT1 –or- PEEK VESTAKEEP i-Grade with Tantalum marker pins; titanium alloy; or a combination of PEEK, Titanium Alloy, or both.

Device Story

Lumbar interbody fusion cage system; used to maintain intervertebral spacing during fusion procedures. Implanted in lumbar spine (L2-S1) by surgeons; requires supplemental fixation. Device acts as structural spacer; supports bone graft fusion. Available in various sizes/configurations to match patient anatomy. Fabricated from PEEK, titanium alloy, or combinations; includes tantalum marker pins for radiographic visualization. Benefits patients by stabilizing spinal segments and facilitating fusion in degenerative disc disease cases.

Clinical Evidence

Bench testing only. Mechanical performance testing included static axial compression, static compression shear, dynamic axial and compression shear, expulsion, and subsidence testing. Results demonstrate equivalent mechanical performance to predicate devices.

Technological Characteristics

Materials: PEEK (ASTM F2026), Titanium alloy Ti-6Al-4V (ASTM F136), Tantalum (ASTM F560). Instruments: Stainless steel (ASTM F899), Aluminum (ASTM B211, B221). Form factor: Interbody fusion cage with various sizes/configurations. Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD defined as discogenic back pain with degeneration confirmed by history and radiography; may include up to Grade 1 spondylolisthesis or retrolisthesis. Requires use with supplemental lumbar fixation; hyperlordotic cages (>20°) require anterior fixation. Must be used with autogenous bone graft. Patients must have at least 6 months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font. August 26, 2020 CTL Medical Corporation % Mr. Barry E. Sands President and Founder RQMIS, Inc. 110 Haverhill Road, Suite 526 Amesbury, Massachusetts 01913 Re: K192863 Trade/Device Name: MONDRIAN 114 Lumbar Interbody Fusion Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: July 17, 2020 Received: July 27, 2020 Dear Mr. Sands: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Brent L. Showalter, Ph.D. Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K192863 #### Device Name MONDRIANTM Lumbar Interbody Fusion Cage System #### Indications for Use (Describe) The MONDRIAN™ Lumbar Interbody Fusion Cage is indicated for intervertebral body fusion procedures in sketally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. The system is to be used with supplemental fixation cleared for use in the lumbar spine. Hyperlordotic cage offerings (>20°) require the use of anterior fixation. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). This device is to be used with autogenous bone graft. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------| | <div> <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size: 10px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY ## CTL Medical Corporation's MONDIRAN™ Lumbar Interbody Fusion Cage System | Submitter: | Sean Suh | |--------------------------|-----------------------------------------------| | | CTL Medical Corporation | | | 4550 Excel Parkway Suite 300 | | | Addison, TX 75001 | | | Phone: 214-545-5820 | | | Fax: 888-831-4892 | | Contact Person: | Barry Sands | | | RQMIS, Inc. | | | 110 Haverhill Road, Suite 526 | | | Amesbury, MA 01913 | | | Phone: 978-358-7307 | | Date Prepared: | Thursday, July 16, 2020 | | Name of Device: | MONDRIAN™ Lumbar Interbody Fusion Cage System | | Regulation: | 21 CFR 888.3080 | | Class: | Class II | | Name/Address of Sponsor: | Sean Suh | | | CTL Medical Corporation | | | 4550 Excel Parkway Suite 300 | | | Addison, TX 75001 | | Common or Usual Name | Intervertebral Body Fusion Device | Intervertebral Body Fusion Device, Lumbar (Product code MAX) {4}------------------------------------------------ ### Predicate Device | Primary Predicate Device | | |---------------------------------------------------|---------------------------| | • CEZANNE Lumbar Interbody Fusion Cage System, | K121567 | | Additional Predicate Devices | | | • CEZANNE-II Lumbar Interbody Fusion Cage System, | K131981 | | • PHANTOM PLUS Cage System, | K082801 | | • VALEO Spacer System | K091278, K142347, K143518 | | • VALEO-II IBF Device System | K121892, K142347, K143518 | | • VALEO-II IBF Device System, LL | K161405 | | • VALEO-II AL System | K143158 | | • VALEO-II OL/PL System | K143518 | | • Nuvasive TLX Interbody System | K171633 | ### Intended Use / Indications for Use #### Indications for Use: The MONDRIAN™ Lumbar Interbody Fusion Cage is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. The system is to be used with supplemental fixation cleared for use in the lumbar spine. Hyperlordotic cage offerings (>20°) require the use of anterior fixation. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).This device is to be used with autogenous bone graft. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage #### Intended Use: The MONDRIAN™ Lumbar Interbody Fusion Cage system is intended for use as an interbody fusion cage device to maintain lumbar intervertebral spacing and must be used with supplemental fixation. The devices are available in a variety of different sizes and configurations to accommodate anatomical variation in different vertebral levels and/or patient anatomy. The devices are made of PEEK-OPTIMA LT1 –or- PEEK VESTAKEEP i-Grade with Tantalum marker pins; titanium alloy; or a combination of PEEK, Titanium Alloy, or both. #### Technological Characteristics The MONDRIAN™ Lumbar Interbody Fusion Cage System consists of cages made out of PEEK with marker pins made of Tantalum or Titanium alloy, both of which are identical to its predicate device. All of the heights, lengths, and widths are similar and/or within ranges covered by its predicate {5}------------------------------------------------ devices. Material composition profiles are provided below with their associated standards in Table 1. | Material | Standards | |----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Polyetheretherketone (PEEK) | ASTM F2026: Standard Specification for Polyetheretherketone<br>(PEEK) Polymers for Surgical Implant Applications | | Titanium alloy (Ti-6Al-4V) | ASTM F136: Standard Specification for Wrought Titanium-<br>6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical<br>Implant Applications (UNS R56401) | | Tantalum | ASTM F560: Standard Specification for Unalloyed Tantalum for<br>Surgical Implant Applications (UNS R05200, UNS R05400) | | Stainless steel<br>(Instrument shafts) | ASTM F899: Standard Specification for Wrought Stainless Steels for<br>Surgical Instruments | | Aluminum<br>(Handles) | ASTM B211: Standard Specification for Aluminum and Aluminum-<br>Alloy Rolled or Cold Finished Bar, Rod, and Wire<br>ASTM B221: Standard Specification for Aluminum and Aluminum-<br>Alloy Extruded Bars, Rods, Wire, Profiles, and Tubes | Table 1 Material Profile with Associated Standard(s) #### Performance Data Table 2 shows the mechanical testing performed on the MONDRIAN™ Lumbar Interbody Fusion Cage System to establish substantial equivalence with predicate devices and demonstrate structural and functional efficacy. | Testing Performed | | |--------------------------------------|---| | Static Axial<br>Compression | X | | Static<br>Compression Shear | X | | Dynamic Axial &<br>Compression Shear | X | | Expulsion | X | | Subsidence | X | Table 2 Performance Testing The subject lumbar devices demonstrated equivalent mechanical performance to the cited predicate device under the same test conditions. {6}------------------------------------------------ ## Substantial Equivalence The MONDRIAN™ Lumbar Interbody Fusion Cage System is as safe and effective as the predicate CEZANNE™ and CEZANNE-II™ Lumbar Interbody Fusion Cage Systems (K121567, K131981, K082801). The MONDRIAN™ Lumbar Interbody Fusion Cage System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The technological differences between the MONDRIAN™ Lumbar Interbody Fusion Cage System and its predicate devices raise no new issues of safety or effectiveness. Thus, MONDRIAN™ Lumbar Interbody Fusion Cage System is substantially equivalent. The MONDRIAN™ Lumbar Interbody Fusion Cage System is as safe and effective as the CEZANNE™ and CEZANNE-II™ Lumbar Interbody Fusion Cage Systems (K121567, K131981, K082801). The MONDRIAN™ Lumbar Interbody Fusion Cage System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the MONDRIAN™ Lumbar Interbody Fusion Cage System and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the MONDRIAN™ Lumbar Interbody Fusion Cage System is as safe and effective as the CEZANNE™ and CEZANNE-II™ Lumbar Interbody Fusion Cage Systems (K121567, K131981, K082801). Thus, the MONDRIAN™ Lumbar Interbody Fusion Cage System is substantially equivalent.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...