K191134 · Medicrea International SA · MAX · Oct 30, 2019 · Orthopedic
Device Facts
Record ID
K191134
Device Name
IB3D ALIF
Applicant
Medicrea International SA
Product Code
MAX · Orthopedic
Decision Date
Oct 30, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
MEDICREA® INTERNATIONAL IB3D ALIF device is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I Spondylolisthesis or Retrolisthesis at the involved level(s). This device is to be used with autogenous bone graft. IB3D ALIF Device is to be used with supplemental fixation. Patients should have at least six (6) months of nonoperative treatment prior to treatment with an intervertebral cage.
Device Story
IB3D ALIF is an intervertebral body fusion device designed for anterior approach spinal fusion. It is a single implant available in various heights, lengths, and lordotic angles. The device is implanted by a surgeon between two adjacent vertebrae to provide immobilization, stabilization, and support for spinal fusion. It functions by providing mechanical resistance and stability to the operated spinal segment. The device is used in a clinical/surgical setting. It is intended to be used in conjunction with autogenous bone graft and supplemental fixation to facilitate a solid spinal fusion. The device benefits patients by stabilizing the spine and addressing discogenic back pain associated with degenerative disc disease.
Clinical Evidence
No clinical data provided. Evidence consists of non-clinical bench testing, including expulsion testing for new endplate configurations and Finite Element Analysis (FEA) to evaluate mechanical performance relative to predicate devices.
Technological Characteristics
Material: Titanium alloy Ti-6Al-4V ELI (ASTM F3001). Form factor: Intervertebral cage with varying heights, lengths, and lordosis. Sterilization: Provided sterile. Connectivity: None. Software: None.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1, including those with up to Grade I Spondylolisthesis or Retrolisthesis. Requires autogenous bone graft and supplemental fixation. Patients must have failed at least 6 months of nonoperative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K221936 — Standalone ALIF Interbody Fusion System · Eminent Spine · Oct 17, 2022
K132852 — PEEK CAGE FOR THE VERTEBRAL SPINE · Neoortho Produtos Ortopedicos S/A · May 29, 2014
K163481 — HALF DOME Posterior Lumbar Interbody System · Astura Medical · May 3, 2017
K141665 — NuVasive CoRoent System · Nuvasive, Inc. · Mar 13, 2015
K171633 — NuVasive TLX Interbody System · Nu Vasive, Incorporated · Sep 28, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.
October 30, 2019
Medicrea International S.A. David Ryan Chief Operating Officer 5389 route de Strasbourg - Vancia Rillieux-la-Pape 69140 France
Re: K191134
Trade/Device Name: IB3D ALIF Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: October 1, 2019 Received: October 3, 2019
Dear David Ryan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Ronald P. Jean, Ph.D. Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191134
Device Name IB3D ALIF
#### Indications for Use (Describe)
MEDICREA® INTERNATIONAL IB3D ALIF device is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I Spondylolisthesis or Retrolisthesis at the involved level(s). This device is to be used with autogenous bone graft. IB3D ALIF Device is to be used with supplemental fixation. Patients should have at least six (6) months of nonoperative treatment prior to treatment with an intervertebral cage.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|----------------------------------------------------------------------------------------------|
| | X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY MEDICREA INTERNATIONAL's IB3D ALIF
In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the IB3D ALIF:
Date Prepared: April, 25, 2019
- 1. Submitter:
MEDICREA INTERNATIONAL 5389 route de Strasbourg - Vancia RILLIEUX-LA-PAPE 69140 FR
Contact Person: David RYAN MEDICREA INTERNATIONAL 5389 route de Strasbourg - Vancia RILLIEUX-LA-PAPE 69140 FR +33 (0)4 72 01 87 87
- 2. Trade name: IB3D ALIF
# Regulatory Identification/ Classification
FDA classification name: Intervertebral Body Fusion Device with Bone Graft, lumbar Requlation Number : 21CFR 888.3080 Product Code: MAX Class II
#### 3. Predicate or legally marketed devices which are substantially equivalent:
### Primary predicate device:
- Cellular Titanium ALIF, (EIT, K172888) .
### Additional Predicates:
- Anterior Spine Truss System (STS) Interbody Fusion Device, (4WEB, K170851) ●
- . IMPIX 3D print (MEDICREA INTERNATIONAL, K163595)
Those three devices have not been subject to a design-related recall.
#### 4. Description of the device:
MEDICREA® INTERNATIONAL IB3D ALIF device consists of one single implant available in several heights, lengths, lordosis and footprints. It is intended for insertion between two adjacent vertebrae by an anterior approach. The MEDICREA® INTERNATIONAL implant is manufactured from titanium alloy Ti-6Al-4V ELI following standards ASTM F3001, a radio opaque material. As any orthopaedic implant, the lumbar interbody device must not be reused. The surgeon should strictly follow the recommendations provided in the surgical technique.
The purpose of this submission is to introduce IB3D ALIF.
MATERIALS: Titanium Alloy (Ti-6Al-4V) according to the ASTM F3001.
{4}------------------------------------------------
# Function:
The IB3D ALIF was developed as an implant:
- to provide immobilization and stabilization of posterior spinal segments ●
- to augment the development of a solid spinal fusion
- to provide stability to ease fusion ●
- . to be mechanically resistant to allow the fusion of the operated level
## 5. Indication for Use
MEDICREA® INTERNATIONAL IB3D ALIF device is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I Spondylolisthesis or Retrolisthesis at the involved level(s). This device is to be used with autogenous bone graft. IB3D ALIF Device is to be used with supplemental fixation. Patients should have at least six (6) months of nonoperative treatment prior to treatment with an intervertebral cage.
# 6. Substantial equivalence claimed to predicate devices
The IB3D ALIF are technologically similar to the already cleared EIT Cellular Titanium® ALIF (K172888), 4WEB Anterior Spine Truss System (STS) Interbody Fusion Device (K170851) and MEDICREA INTERNATIONAL IMPIX 3D print (K163595) in terms of intended use, material used, mechanical safety and performances.
The table below compares the features and characteristics of the submitted IB3D ALIF to their predicate devices.
| Device | MEDICREA<br>INTERNATIONAL<br>ALIF IB3D | EIT<br>Cellular<br>Titanium® ALIF | 4WEB<br>Anterior Spine<br>Truss System (STS)<br>Interbody Fusion<br>Device | MEDICREA<br>INTERNATIONAL<br>IMPIX 3D print |
|------------------|----------------------------------------|-----------------------------------|----------------------------------------------------------------------------|---------------------------------------------|
| 510(k)<br>number | K191134 | K172888 | K170851 | K163595 |
| Intended use | | | | |
| Lumbar | Yes | Yes | Yes | Yes |
| Approach | Anterior | Anterior | Anterior | Posterior |
| Material | | | | |
| Raw material | Titanium<br>Ti6Al4V ELI | Titanium<br>Ti6Al4V ELI | PET &<br>Pure titanium<br>(T40) conforming<br>to ASTM F67. | Titanium<br>Ti6Al4V ELI |
| Sterilization | Provided sterile | Provided<br>sterile | Provided sterile | Provided sterile |
# 7. Non-clinical Test Summary:
The following performance data was provided in support of the substantial equivalence determination.
Biocompatibility Testing
{5}------------------------------------------------
The IB3D ALIF devices are made from the same materials as the predicates and the manufacturing processes are the same as a predicate.
#### Mechanical testing
Finite Element Analysis (FEA) was conducted and it was determined that no new worse-case implant was created. Therefore additional testing per ASTM F2077 "Standard Test Methods for Intervertebral Body Fusion Devices" and ASTM F2267 "Standard Test Methods Measuring Load Induced Subsidence of Intervertebral Body Fusion Device under Static Axial Compression" was not conducted. However, expulsion testing was conducted to evaluate new endplate configurations.
#### 8. Conclusions
IB3D ALIF devices are substantially equivalent to legally marketed predicate devices
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.