The Lumfuse-TP cage is intended to be used in spinal fusion procedures for patients diagnosed with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to S1. DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. When used for these indications, the Lumfuse-TP cage is intended for use with supplemental fixation systems cleared for use in the lumbar spine. These patients should be skeletally mature and have had six months of nonoperative treatment. The Lumfuse-TP cage is intended to be used with autograft and lor allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion.
Device Story
LumFuse-TP is a PEEK interbody cage for lumbar spinal fusion; inserted between vertebral bodies to provide support and correction. Hollow geometry allows packing with autograft or allogenic bone graft to promote fusion. Used in spinal surgery by surgeons; requires supplemental fixation. Device provides structural support during bone healing; benefits patients by stabilizing spinal segments affected by degenerative disc disease. No software or electronic components.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including static and dynamic axial compression, compression-shear (ASTM F2077-14), and subsidence (ASTM F2267-04). Biocompatibility, cleaning, and sterilization validation also performed.
Indicated for skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to S1, including those with up to Grade 1 Spondylolisthesis or retrolisthesis, who have failed six months of nonoperative treatment. Used with supplemental lumbar fixation and bone graft.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
K172083 — Idys ALIF System · Clariance, Sas · Nov 8, 2017
K131077 — LIFE SPINE PLATEAU SPACER SYSTEM · Life Spine · Jul 17, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
PRECIFIT MEDICAL LTD % Mr. Kellen Hills Quality and Regulatory Consultant Orchid Design 4600 East Shelby Drive. Suite 1 Memphis, Tennessee 38118
July 25, 2017
Re: K171630
Trade/Device Name: LumFuse-TP Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: May 22, 2017 Received: June 2, 2017
Dear Mr. Hills:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
## Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K171630
Device Name Lumfuse-TP
#### Indications for Use (Describe)
The Lumfuse-TP cage is intended to be used in spinal fusion procedures for patients diagnosed with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to S1. DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. When used for these indications, the Lumfuse-TP cage is intended for use with supplemental fixation systems cleared for use in the lumbar spine. These patients should be skeletally mature and have had six months of nonoperative treatment. The Lumfuse-TP cage is intended to be used with autograft and lor allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion.
#### Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Sun
[As Required by 21 CFR 8
(a)(1) Submitted By: Address:
> Phone: Date: Contact Persons Primary: Secondary:
Proprietary Name: (a)(2) Common Name:
Classification Name and Reference:
Product Code:
(a)(3) Predicate Devices: Primary:
PRECIFIT MEDICAL LTD 951 Aviation Pkwy Ste 100, Morrisville, NORTH CAROLINA, 27560 901-433-1990 July 21, 2017
Kellen Hills (Orchid Design Consulting) Eric Wu (Precifit Medical Ltd)
Lumfuse-TP Lumbar interbody fusion device, interbody cage 21 CFR 888.3080: Intervertebral Fusion Device With Bone Graft, Lumbar MAX
Medtronic® CAPSTONE (K151128);
(a)(4) Device Description:
> The Lumfuse-TP cage consists of PEEK cages of various lengths and heights, which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The hollow geometry of the implants allows them to be packed with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The devices are made from PEEK radiolucent material (ASTM F2026) with embedded tantalum x-ray markers (ASTM F560 or ISO 13782).
The purpose of this submission is to gain initial marketing authorization in the United States.
- (a)(5) Indications for Use:
The Lumfuse-TP cage is intended to be used in spinal fusion procedures for patients diagnosed with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 to S1. DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. When used for these indications, the Lumfuse-TP cage is intended for use with supplemental fixation systems cleared for use in the lumbar spine. These patients should be skeletally mature and have had six months of nonoperative treatment. The Lumfuse-TP cage is intended to be used with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when the subject device is used as an adjunct to fusion.
- (a)(6) Comparison of Technological Characteristics: The Lumfuse-TP has the same intended use and is similar in basic shape, material and performance characteristics to the predicate device. The technological characteristics do not raise any new questions of safety and efficacy.
- (b)(1) Non-clinical testing:
The worst case devices were evaluated for mechanical performance in static and dynamic axial compression and compression-shear per ASTM F2077-14 and subsidence per ASTM F2267-04 (2011). End-user cleaning and sterilization as well as implantable device biocompatibility endpoints were also evaluated.
{4}------------------------------------------------
#### (b)(2) Clinical testing:
Clinical testing was not required to demonstrate substantial equivalence in this premarket notification.
- Conclusions: (b)(3)
Based on the information provided in this premarket notification, we believe that the subject Lumfuse-TP demonstrates substantial equivalence to the identified predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.