EVOS Lumbar Interbody System

K150321 · Cutting Edge Spine, LLC · MAX · Jul 13, 2015 · Orthopedic

Device Facts

Record IDK150321
Device NameEVOS Lumbar Interbody System
ApplicantCutting Edge Spine, LLC
Product CodeMAX · Orthopedic
Decision DateJul 13, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The EVOS Lumbar Interbody System (EVOS device) is an intervertebral body device intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine with up to Grade 1 spondylolisthesis at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. The EVOS Lumbar Interbody System is intended to be used with autologous bone graft to facilitate fusion. It is to be used in patients who have had six months of non-operative treatment and is to be implanted via a direct posterior or transforaminal approach. The EVOS CURVED devices are implanted singly, while the EVOS ROTATE and EVOS STRAIGHT devices may be implanted singly or in pairs in the lumbosacral spine. The EVOS Lumbar Interbody System is intended to be used with supplemental fixation.

Device Story

EVOS Lumbar Interbody System is an intervertebral fusion implant for lumbar spine stabilization. Device consists of PEEK-OPTIMA (standard or HA-enhanced) spacers with embedded tantalum beads for radiographic visualization. Implants feature internal cavities for autologous bone graft and surface ridges/teeth for vertebral endplate anchoring. Available in STRAIGHT, CURVED, and ROTATE geometries with varying heights, widths, and lordotic angles. Surgical instruments facilitate implantation via direct posterior or transforaminal approach. Used by surgeons to treat DDD; requires supplemental fixation. Benefits include structural support and facilitation of spinal fusion in patients with discogenic pain.

Clinical Evidence

Bench testing only. Mechanical testing performed on gamma-irradiated, aged devices per FDA guidance. Tests included expulsion, static axial compression (ASTM F2077), static compressive shear (ASTM F2077), static subsidence (ASTM F2267), and dynamic axial compression (ASTM F2077).

Technological Characteristics

Materials: PEEK-OPTIMA LT1 or PEEK-OPTIMA HA Enhanced; Tantalum beads for radiopacity. Design: Rectangular spacers with surface ridges/teeth. Energy: None (passive implant). Connectivity: None. Sterilization: Gamma irradiation. Standards: ASTM F2077 (axial compression, compressive shear, dynamic axial compression), ASTM F2267 (subsidence).

Indications for Use

Indicated for skeletally mature patients with lumbar Degenerative Disc Disease (DDD) and up to Grade 1 spondylolisthesis at 1-2 contiguous levels (L2-S1) who failed 6 months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is a stylized image of three human profiles facing right, layered on top of each other to create a sense of depth. July 13, 2015 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Cutting Edge Spine, LLC Mr. John Souza, Sr. Director of Engineering 101 Waxhaw Professional Park Drive, Suite A Waxhaw, North Carolina 28173 Re: K150321 Trade/Device Name: EVOS Lumbar Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: July 2, 2015 Received: July 7, 2015 Dear Mr. Souza: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ Page 2 - Mr. John Souza, Sr. forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K150321 Device Name EVOS Lumbar Interbody System ### Indications for Use (Describe) The EVOS Lumbar Interbody System (EVOS device) is an intervertebral body device intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine with up to Grade 1 spondylolisthesis at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. The EVOS Lumbar Interbody System is intended to be used with autologous bone graft to facilitate fusion. It is to be used in patients who have had six months of non-operative treatment and is to be implanted via a direct posterior or transforaminal approach. The EVOS CURVED devices are implanted singly, while the EVOS ROTATE and EVOS STRAIGHT devices may be implanted singly or in pairs in the lumbosacral spine. The EVOS Lumbar Interbody System is intended to be used with supplemental fixation. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary #### I. SUBMITTER Cutting Edge Spine, LLC 101 Waxhaw Professional Park Dr., Suite A Waxhaw, NC 28173 Tel: (704) 243-0892 ext 12 Fax: (704) 731-2559 Contact Person: John Souza Sr., Director of Engineering Date Prepared: February 06, 2015 #### II. DEVICE Name of Device: EVOS Lumbar Interbody System Common or Usual Name: Intervertebral fusion device with bone graft, lumbar Classification Name: Intervertebral body fusion device (21 CFR 888.3080) Regulatory Class: II Product Code: MAX #### III. PREDICATE DEVICE | | 510(k) Number | Device | Manufacturer | |-------------------------|---------------|---------------------------------------------|-------------------------| | Primary<br>Predicate | K102957 | EVOL Intervertebral Body<br>Fusion Device | Cutting Edge Spine, LLC | | Additional<br>Predicate | K071724 | Lucent Intervertebral Body<br>Fusion Device | Spinal Elements | | Additional<br>Predicate | K073291 | Capstone Spinal System | Medtronic | | Additional<br>Predicate | K120368 | Capstone Control Spinal<br>System | Medtronic | No reference devices were used in this submission. #### IV. DEVICE DESCRIPTION The EVOS Lumbar Interbody System is an Intervertebral Body Fusion System for the lumbar spine. The EVOS Lumbar Interbody System includes surgical instruments to deliver the EVOS devices via a Posterior or Transforaminal approach. The generally rectangular design of the EVOS device {4}------------------------------------------------ incorporates an internal cavity for insertion of biological bone growth material through the top and bottom openings, as well as four small transverse holes (two on each side) that are designed to enhance bony ingrowth. Each device has a series of ridges or "teeth" on the top and bottom surfaces that are designed to ensure strong anchoring on the vertebral body endplates and resist backing out of the implant. The EVOS devices are provided sterile (gamma irradiated) for single use. The EVOS device is made of radiolucent PEEK-OPTIMA® LT1 or PEEK-OPTIMA® HA Enhanced material. In addition, tantalum beads are embedded in the spacers to allow for radiographic visualization. The geometric shape of the EVOS devices are categorized as either STRAIGHT, CURVED, or ROTATE. All are available in a range of sizes, as well as flat and biconvex endplates, and with various degrees of lordosis to accommodate variations in patients' anatomy. Most of the EVOS STRAIGHT devices are offered with an option of a Standard nose, Wide nose, or Narrow nose. The devices are available in various heights, widths, and lengths to accommodate various surgical technique options. #### V. INDICATIONS FOR USE The EVOS Lumbar Interbody System (EVOS device) is an intervertebral body device intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) of the lumbar spine with up to Grade 1 spondylolisthesis at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. The EVOS Lumbar Interbody System is intended to be used with autologous bone graft to facilitate fusion. It is to be used in patients who have had six months of non-operative treatment and is to be implanted via a direct posterior or transforaminal approach. The EVOS CURVED devices are implanted singly, while the EVOS ROTATE and EVOS STRAIGHT devices may be implanted singly or in pairs in the lumbosacral spine. The EVOS Lumbar Interbody System is intended to be used with supplemental fixation. ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE Intervertebral Body Fusion through a Posterior or Transformanial approach is the technological principle for both the subject and predicate devices. It is based on the use of surgical instrumentation to insert PEEK implants into the intervertebral space in order to assist in fusion. At a high level, the subject and predicate devices are based on the following same technological elements: - Indications for Use - Structural support mechanism . #### VII. PERFORMANCE DATA The EVOS device was tested in compliance with FDA's guidance document titled "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device" and demonstrated substantially equivalent performance characteristics to the identified predicate devices. {5}------------------------------------------------ The EVOS device has been tested in the following test modes: - Mechanical Testing on the final device that has been gamma irradiated and aged. ● - o Expulsion Testing - o Static Axial Compression per ASTM F2077 - Static Compressive Shear per ASTM F2077 O - Static Testing in Subsidence per ASTM F2267 O - Dynamic Axial Compression per ASTM F2077 O #### VIII. CONCLUSIONS The overall technology characteristics and mechanical performance data lead to the conclusion that the EVOS Lumbar Interbody System is substantially equivalent to the EVOL Intervertebral Body Fusion Device from Cutting Edge Spine (K102957), the Lucent Intervertebral Body Fusion Device from Spinal Elements (K071724), the Medtronic Capstone Spinal System (K073291), and the Medtronic Capstone Control Spinal System (K120368).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...