SPINEOLOGY PEEK LUMBAR INTERBODY FUSION DEVICE

K123652 · Spineology, Inc. · MAX · Mar 22, 2013 · Orthopedic

Device Facts

Record IDK123652
Device NameSPINEOLOGY PEEK LUMBAR INTERBODY FUSION DEVICE
ApplicantSpineology, Inc.
Product CodeMAX · Orthopedic
Decision DateMar 22, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Rampart O and P Devices are an intervertebral body fusion device indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These devices may be implanted singly or in pairs via an open or a minimally invasive posterior or transforaminal approach. The Rampart O and P Devices are designed for use with autograft as an adjunct to fusion and is intended for use with supplemental fixation systems cleared by the FDA for use in the lumbar spine.

Device Story

Rampart O and P are intervertebral body fusion devices; implanted in lumbar spine (L2-S1) via open or minimally invasive posterior/transforaminal approach. Used with autograft to promote fusion; requires supplemental fixation. Features include variable lengths, heights, and tantalum markers for radiographic visualization; allows for post-insertion rotation. Used by orthopedic or neurosurgeons in clinical/surgical settings. Provides structural support to disc space to facilitate spinal fusion; benefits patients by stabilizing symptomatic spinal segments.

Clinical Evidence

Bench testing only. Static compression and static compression shear testing performed per ASTM F2077 to demonstrate mechanical equivalence to predicate devices.

Technological Characteristics

Intervertebral body fusion device; PEEK construction; includes tantalum markers for radiographic visualization. Designed for manual surgical implantation. Non-sterile. Complies with ASTM F2077 for mechanical performance.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) and up to Grade I spondylolisthesis at one or two contiguous lumbar levels (L2-S1) who have failed six months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summary # MAR 22 2013 | Applicant: | Spineology Inc.<br>7800 3rd Street N., Suite 600<br>Saint Paul, MN 55128<br>Phone: 651-256-8500 Fax: 651-256-8505 | |-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Bryan Becker | | Date Prepared: | November 21, 2012 | | Trade Name: | Rampart O and Rampart P | | Product Classification<br>and Code: | Class II Medical Device, Product Code MAX, 21 CFR 888.3080 | | Predicate Device(s): | Spineology PEEK Lumbar Interbody Fusion Devices (K111880,<br>K113030, and K120293) | | Device Description: | The Spineology Rampart O and P Devices are designed for use with<br>autograft as an adjunct to fusion and is intended for use with supplemental<br>fixation systems cleared for use in the lumbar spine. The device is<br>available in a range of lengths and heights. The device and associated<br>instruments are provided non-sterile. | | Intended Use: | The Rampart O and P Devices are an intervertebral body fusion device<br>indicated for intervertebral body fusion at one level or two contiguous<br>levels in the lumbar spine from L2 to S1 in patients with degenerative disc<br>disease (DDD) with up to Grade I spondylolisthesis at the involved<br>level(s). DDD is defined as back pain of discogenic origin with<br>degeneration of the disc confirmed by patient history and radiographic<br>studies. These patients should be skeletally mature and have had six<br>months of non-operative treatment.<br>These devices may be implanted singly or in pairs via an open or a<br>minimally invasive posterior or transforaminal approach.<br>The Rampart O and P Devices are designed for use with autograft as an<br>adjunct to fusion and is intended for use with supplemental fixation<br>systems cleared by the FDA for use in the lumbar spine. | | Purpose of this 510(k): | The purpose of this 510(k) is to add devices as small as 20 mm in length<br>and add products that include features that allow for rotation of the<br>devices after insertion on implants with tantalum markers. | | Summary of<br>Technological<br>Characteristics: | The device is shown to be substantially equivalent to the intended use,<br>materials, configuration, and performance characteristics of the predicate<br>products. | | Testing | The Rampart O and P Devices were tested in compliance with FDA's<br>guidance document titled "Class II Special Controls Guidance<br>Document: Intervertebral Body Fusion Device". Preclinical testing<br>according to ASTM F2077, including static compression and static<br>compression shear. This testing demonstrated similar performance<br>characteristics to the identified predicate devices. | | Conclusion: | The information submitted in this premarket notification supports a<br>determination that the Spineology Rampart O and P Devices are<br>substantially equivalent in technological characteristics and intended use<br>to the predicate devices. | {1}------------------------------------------------ · . . . . · : {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the emblem. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 22, 2013 Spineology, Incorporated % Mr. Brvan Becker Regulatory Affairs Manager 7800 3rd Street North, Suite 600 Saint Paul, Minnesota 55128 Re: K123652 Trade/Device Name: Rampart O and Rampart P Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: February 20, 2012 Received: March 8, 2012 Dear Mr. Becker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ #### Page 2 - Mr. Bryan Becker forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. ErinL.Keith Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### K123652 ### Device Name: Rampart O and Rampart P #### Indications for Use: Rampart O and Rampart P are intervertebral body fusion device indicated for intervertebral body fusion at one level or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade I spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. These patients should be skeletally mature and have had six months of nonoperative treatment. These devices may be implanted singly or in pairs via an open or a minimally invasive posterior or transforaminal approach. Rampart O and Rampart P are designed for use with autograft as an adjunct to fusion and is intended for use with supplemental fixation systems cleared by the FDA for use in the lumbar spine. Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Anton E. Dmitriev, PhD Division of Orthopedic Devices
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...