SPINESMITH CYNCH SPINAL SYSTEM

K090376 · Spine Smith Partners L.P. · MAX · Apr 1, 2009 · Orthopedic

Device Facts

Record IDK090376
Device NameSPINESMITH CYNCH SPINAL SYSTEM
ApplicantSpine Smith Partners L.P.
Product CodeMAX · Orthopedic
Decision DateApr 1, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Cynch System is indicated intervertebral body fusion of the lumbar spine, from L2 to SI, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion.

Device Story

SpineSmith Cynch Spinal System is an intervertebral body fusion device; curved design for medial-lateral placement; available in 25mm and 30mm footprints with 5° lordosis; hollow center for autograft bone placement; large anterior graft windows; superior and inferior surface teeth for stability. Used by surgeons in spinal fusion procedures to treat degenerative disc disease and Grade I spondylolisthesis. Device provides structural support to the intervertebral space to facilitate fusion; requires supplemental fixation.

Clinical Evidence

Bench testing only.

Technological Characteristics

Intervertebral body fusion device; curved geometry; 25mm/30mm footprints; 5° lordosis; hollow body with anterior graft windows; superior/inferior surface teeth for fixation. Material and sterilization details not specified.

Indications for Use

Indicated for intervertebral body fusion of the lumbar spine (L2-S1) in skeletally mature patients with degenerative disc disease (DDD) and up to Grade I spondylolisthesis, following six months of failed non-operative treatment. Used at one or two contiguous levels with supplemental fixation and autograft.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K090376 # SpineSmith Cynch Spinal System # APR - 1 2009 ## 510(k) Summary of Safety and Effectiveness SUBMITTED BY SpineSmith Partners, LP 8140 N. Mopac, Bldg II, Suite 120 Austin, TX 78759 ESTABLISHMENT 3006404071 REGISTRATION NUMBER CONTACT PERSON Robert Jones Vice President, Research and Development Phone: 512-206-0770 Fax: 512-637-6750 Email: rjones@spinesmithusa.com SUBMISSION PREPARED BY Lisa Peterson Kaedon Consulting, LLC Phone: 512-507-0746 February 16, 2009 DATE PREPARED CLASSIFICATION MAX 888.3080- Intervertebral Fusion Device with Bone Graft, Lumbar COMMON NAME Intervertebral Body Fusion Device PROPRIETARY NAME SpineSmith Cynch Spinal System SUBSTANTIAL EQUIVALENCE The SpineSmith Cynch System was determined to be substantially equivalent to several recently down classified cages: -LT-CAGE® Peek Lumbar Tapered Fusion Device (P970015, Medtronic Sofamor Danek, Approved 9/10/03) BAK® Cage (P950002, Zimmer Spine, Approved -7/8/03) - RAY® Threaded Fusion Cage (P950019, Stryker, -Approved 9/4/03) - Lumbar I/F Cage (P960025, DePuy, Approved 3/4/05) - 8140 N. Mopac, Bldg II, Suite 120 Austin, Texas 78759 SpineSmith Partners, LP Phone: 512-206-0770 Fax: 512-637-6750 {1}------------------------------------------------ K090376 ## DEVICE DESCRIPTION Cynch is curved to allow for medial-lateral placement of the device, and is available in two footprints (25mm and 30mm) with a 5° lordosis in various heights. Cynch is hollow to allow for the placement of autograft bone, and has large anterior graft windows to facilitate fusion. There are teeth on the superior and inferior surface of the device to provide increased stability and inhibit movement of the implant. ## INDICATIONS: The Cynch System is indicated intervertebral body fusion of the lumbar spine, from L2 to SI, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. ## MECHANICAL TEST DATA Mechanical test results demonstrate that the Cynch System is substantially equivalent to the predicate. 8140 N. Mopac, Bldg II, Suite 120 Austin, Texas 78759 SpineSmith Partners, LP Phone: 512-206-0770 Fax: 512-637-6750 {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, with stylized wings and tail feathers. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SpineSmith Partners, LP % Mr. Robert Jones Vice President, Research and Development 8140 N. Mopac, Building II, Suite 120 Austin. Texas 78759 APR - 1 2009 Re: K090376 Trade/Device Name: SpineSmith Cynch Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: II Product Code: MAX Dated: February 16, 2009 Received: February 17, 2009 Dear Mr. Jones: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Paris 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Sincerely yours, Calben Buchen Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## INDICATIONS FOR USE 510(k) Number (if known): K090376 Device Name: SpineSmith Cynch Spinal System Indications for Use: The Cynch System is indicated intervertebral body fusion of the lumbar spine, from L2 to SI, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |--------------------------------------------------------|------------------------------------------------------------| | | (Division Sign-Off) | | | Division of General, Restorative, and Neurological Devices | SpineSmith Partners, LP | 8140 N. Mopac, Bldg II, Suite 120<br>Austin, Texas 78759 | Phone: 512-206-0770<br>Fax: 512-637-6750 | |----------------------------------------------------------|------------------------------------------| |----------------------------------------------------------|------------------------------------------| Pg. 1 of 510(k) Number K090376
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...