Revision Femoral Stem

K151739 · Lima Corporate S.P.A. · LZO · Nov 23, 2015 · Orthopedic

Device Facts

Record IDK151739
Device NameRevision Femoral Stem
ApplicantLima Corporate S.P.A.
Product CodeLZO · Orthopedic
Decision DateNov 23, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Revision Femoral Stem is indicated for patients whose bone stock is of poor quality or inadequate for other reconstruction techniques as indicated by deficiencies of the femoral head, neck or portions of the proximal femur. It is intended for cementless revision hip arthroplasty on both uncemented and cemented femoral implants.

Device Story

Revision Femoral Stem is a modular hip prosthesis consisting of a femoral stem, modular neck, and safety screw. Used in revision hip arthroplasty to replace deficient proximal femoral bone stock. The stem features a tapered, finned profile with a sandblasted surface for cementless fixation. The modular neck attaches to the stem via a taper coupling secured by a safety screw, which includes a UHMWPE pin to prevent loosening. The device is implanted by orthopedic surgeons in a clinical/hospital setting. It functions as a structural replacement for the proximal femur, providing a stable foundation for modular femoral heads or bipolar heads. By restoring femoral geometry and stability, the device aims to improve patient mobility and function following revision surgery.

Clinical Evidence

No clinical data. Substantial equivalence was demonstrated through non-clinical bench testing, including fatigue testing (ISO 7206-4, ISO 7206-6), fretting evaluation, safety screw disassembly torque analysis, and range of motion testing (ISO 21535).

Technological Characteristics

Materials: Ti6Al4V (ASTM F1472, ISO 5832-3) for stem, neck, and screw; UHMWPE (ISO 5834-2, ASTM F648) for screw pin. Design: Modular femoral stem with tapered, finned profile and sandblasted surface. Dimensions: 14-24 mm diameters, 140-200 mm lengths. Neck angles: 131° and 135°. Fixation: Cementless. Sterilization: Not specified.

Indications for Use

Indicated for patients with poor quality or inadequate bone stock due to deficiencies of the femoral head, neck, or proximal femur requiring cementless revision hip arthroplasty.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three human profiles incorporated into its design, symbolizing the department's focus on health and human well-being. The logo is rendered in black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 23, 2015 Limacorporate S.p.A. % Mr. Stephen J. Peoples President Peoples & Associates 411 Auditorium Boulevard Winona Lake, Indiana 46590 Re: K151739 Trade/Device Name: Revision Femoral Stem Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO, LPH, KWY Dated: October 27, 2015 Received: October 30, 2015 Dear Mr. Peoples: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) K151739 Device Name Revision Femoral Stem Indications for Use (Describe) Revision Femoral Stem is indicated for patients whose bone stock is of poor quality or inadequate for other reconstruction techniques as indicated by deficiencies of the femoral head, neck or portions of the proximal femur. It is intended for cementless revision hip arthroplasty on both uncemented and cemented femoral implants. | <span style="text-decoration: overline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) | |-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------| |-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Type of Use (Select one or both, as applicable) {3}------------------------------------------------ # Summarv of Safety and Effectiveness Date:November 17, 2015 Manufacturer: Limacorporate S.p.A. Via Nazionale, 52 33038 – Villanova di San Daniele Udine - Italy U.S. Contact Person: Dr. Stephen J. Peoples Principal Consultant SPeoples VMD @ gmail.com PEOPLE&ASSOCIATED CONSULTING, LLC 5010 LODGE POLE LANE FORT WAYNE INDIANA, 46814 USA Phone: 260-645-0327 FAX: +39 0432945512 | Product | Common name | Product<br>Code | Regulation and Classification<br>Name | |--------------------------|---------------------------|-----------------|----------------------------------------------------------------------------------------------------------------------------| | Revision Femoral<br>Stem | Femoral Hip<br>Prosthesis | LZO | Hip joint metal/ceramic/polymer semi-<br>constrained cemented or nonporous<br>uncemented prosthesis per 21 CFR<br>888.3353 | | | | LPH | Hip joint metal/polymer/metal semi-<br>constrained porous-coated uncemented<br>prosthesis per 21 CFR 888.3358 | | | | KWY | Hip joint femoral (hemi-hip)<br>metal/polymer cemented or uncemented<br>prosthesis per 21 CFR 888.3390 | ## Description: The Revision Femoral Stem consists of a femoral stem and a modular neck, assembled together through a taper coupling stabilized by means of a safety screw. When used in total hip arthroplasty, the Revision Femoral Stem prosthesis is intended for use with modular femoral heads and compatible acetabular cups. When used in partial hip arthroplasty, the Revision Femoral Stem prosthesis is intended for use with a Lock Bipolar Head. The stem is made of Ti6Al4V (ASTM F1472 – ISO 5832-3). The stem is straight, with a tapered profile, round finned section and a rounded tip to facilitate insertion. The external surface of the stem features a macro-roughened finishing obtained by sandblasting. Diameters ranging from 14 to 24 mm, with increases of 2 mm, for a total of 6 diameters are available. Two lengths (140 or 200 mm) for each diameter of the stem are available. The neck is made of Ti6A14V (ASTM F1472 – ISO 5832-3). The same material is also used for the safety screw. which also has a small pin made of UHMWPE (ISO 5834-2. ASTM F648) to help prevent loosening of the safety screw. Two (2) versions of the neck are available , one with a CCD angle of 131° and another with a CCD angle of 135°. Both the neck versions are available in 7 different heights. Distally, the external surface of the {4}------------------------------------------------ neck component features a macro-roughened finishing obtained by sandblasting: proximally, the surface of the neck component is polished to reduce the chance of polyethylene wear particles if the neck accidentally rubs against the polyethylene of the acetabular component. ## Indication For Use: Revision Femoral Stem is indicated for patients whose bone stock is of poor quality or inadequate for other reconstruction techniques as indicated by deficiencies of the femoral head, neck or portions of the proximal femur. It is intended for cementless revision hip arthroplasty on both uncemented and cemented femoral implants. ### Predicate Devices: - Modular Revision Hip Stem (DJO Surgical) cleared via K092331. ● - ZMR Hip System Revision Taper (Zimmer), cleared via K992667 and ZMR Hip . System-XL, cleared via K031572. ### Comparable Features to Predicate Device(s): The Revision Femoral Stem components share the same materials, intended use and basic design features as those of the predicate devices. Non-clinical testing demonstrates that the subject components perform at least as well as the cited predicates. ### Non-Clinical Testing: Mechanical testing had demonstrated the device's ability to perform substantially equivalent to the predicate devices in: - Fatigue testing (ISO 7206-4); - - -Fatigue testing (ISO 7206-6); - Fretting evaluation: - - Analysis of disassembly torque of safety screw; - - Range of motion (ISO 21535). - ## Clinical Testing: Clinical testing was not necessary to demonstrate substantial equivalence of Revision Femoral Stem to the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...