INTER-OP POROUS REVISION SHELL WITH SEALABLE SCREWHOLES
K970705 · Sulzer Orthopedics, Inc. · LPH · May 22, 1997 · Orthopedic
Device Facts
Record ID
K970705
Device Name
INTER-OP POROUS REVISION SHELL WITH SEALABLE SCREWHOLES
Applicant
Sulzer Orthopedics, Inc.
Product Code
LPH · Orthopedic
Decision Date
May 22, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Inter-Op Porous Revision Shell with Sealable Screwholes is a porous coated hemispherical shell that is used in conjunction with a snap-in acetabular insert manufactured from ultra-high molecular weight polyethylene (UHMWPE). This device is intended to replace the acetabulum during total hip arthroplasty and may be implanted with or without bone cement. Diagnostic indications for use of this device include: - revision of a previously implanted acetabular prosthesis; - patient conditions of noninflammatory degenerative joint disease; e.g., avascular necrosis, osteoarthritis, or arthritis secondary to a variety of diseases and anomalies; and, - inflammatory joint disease; e.g., rheumatoid arthritis.
Device Story
Hemispherical metal acetabular shell; Ti-6Al-4V alloy; Cancellous Structured Titanium (CSTI) coating; used with snap-in UHMWPE insert. Features nine screwholes for fixation in ilium, ischium, or pubis; includes sealable plugs to prevent fibrous tissue ingrowth, cement extrusion, or debris migration. Implanted by orthopedic surgeons during total hip arthroplasty; cementless or cemented application. Locking mechanism secures insert to shell. Benefits include improved fixation, debris containment, and versatility in revision scenarios.
Clinical Evidence
Bench testing only. No clinical data provided. Integrity of the locking mechanism between the acetabular shell and insert was evaluated via mechanical testing to ensure attachment strength is comparable to predicate devices.
Technological Characteristics
Hemispherical shell; Ti-6Al-4V alloy; Cancellous Structured Titanium (CSTI) porous coating; UHMWPE snap-in insert. Features nine sealable screwholes. Mechanical locking mechanism for insert retention. Non-software device.
Indications for Use
Indicated for patients undergoing total hip arthroplasty requiring acetabular replacement, including revision of prior prosthesis, noninflammatory degenerative joint disease (e.g., avascular necrosis, osteoarthritis, arthritis secondary to disease/anomalies), and inflammatory joint disease (e.g., rheumatoid arthritis).
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
APR Acetabular Shell (Sulzer Orthopedics Inc.)
S-ROM ZTT II Acetabular Shell with Apical Hole Plug (Johnson & Johnson Orthopedics)
P.F.C. Multi-Holed Porous Coated Acetabular Shell (Johnson & Johnson Orthopedics)
Duraloc 1200 Series Acetabular Cup (DePuy)
Trilogy Multi-Holed Shell (Zimmer)
Related Devices
K955739 — MODULAR ACETABULAR SHELL II · Intermedics Orthopedics · Mar 27, 1996
K022985 — ASCENDENT ACETABULAR SYSTEM · Sulzer Orthopedics, Inc. · Dec 5, 2002
K141043 — CS2 MULTI-HOLE ACETABULAR SHELL · Consensus Orthopedics, Inc. · May 19, 2014
K012739 — CONVERGE ACETABULAR SYSTEM · Sulzer Orthopedics, Inc. · Nov 14, 2001
K983502 — OSTEONICS SECUR-FIT AD GENERATION II ACETABULAR COMPONENT SYSTEM · Osteonics Corp. · Dec 16, 1998
Submission Summary (Full Text)
{0}
MAY 22 1997 K970705
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21CFR 807, this is to serve as a Summary of Safety and Effectiveness for the Inter-Op™ Porous Revision Shell with Sealable Screwholes.
| Submitter: | Sulzer Orthopedics Inc.
9900 Spectrum Drive
Austin, Texas 78717
(512) 432-9687 |
| --- | --- |
| Date: | February 18, 1997 |
| Contact Person: | Jacquelyn Hughes
Manager, Regulatory Affairs |
| Classification Name: | Hip Joint Metal/Polymer/Metal Semi-constrained Porous-coated Uncemented Prosthesis, 21CFR 888.3358 |
| Common/Usual Name: | Metal-backed Acetabular Component |
| Trade/Proprietary Name: | Inter-Op™ Porous Revision Shell with Sealable Screwholes |
## PRODUCT DESCRIPTION
The Inter-Op Porous Revision Shell with Sealable Screwholes is a hemispherical metal shell that is utilized with a snap-in polyethylene insert. The shell is manufactured from Ti-6Al-4V alloy and coated with Cancellous Structured Titanium™ (CSTI™). The insert is manufactured from ultra-high molecular weight polyethylene (UHMWPE) and is available in standard, hooded and hooded protrusio designs in order to address various clinical situations.
The Revision Shell with Sealable Screwholes is designed for screw application with nine screwholes allowing for screw placement into the ilium, ischium and pubis. Furthermore, those screwholes that are not utilized may be plugged after implantation thus limiting the potential for fibrous tissue growth into the shell upon cementless implantation or cement extrusion into the shell upon cemented application. Additionally, if wear of the polyethylene acetabular insert occurs, the hole plugs restrict the debris from migrating through the acetabular shell holes into the acetabulum.
The integrity of the locking mechanism of the Revision Shell was investigated by examining the attachment strength between the acetabular shell and acetabular insert. The strength of the locking mechanism compares favorably to other currently marketed devices.
{1}
112
2
# DIAGNOSTIC INDICATIONS
This device is intended to replace the acetabulum during total hip arthroplasty and may be implanted with or without bone cement. Diagnostic indications for use of this device include:
- revision of a previously implanted acetabular prosthesis;
- patient conditions of noninflammatory degenerative joint disease; e.g., avascular necrosis, osteoarthritis, or arthritis secondary to a variety of diseases and anomalies; and,
- inflammatory joint disease; e.g., rheumatoid arthritis.
# SUBSTANTIAL EQUIVALENCE
The Inter-Op Porous Revision Shell with Sealable Screwholes is substantially equivalent to the APR Acetabular Shell (Sulzer Orthopedics Inc.), the S-ROM ZTT II Acetabular Shell with Apical Hole Plug (Johnson & Johnson Orthopedics), the P.F.C. Multi-Holed Porous Coated Acetabular Shell (Johnson & Johnson Orthopedics), the Duraloc 1200 Series Acetabular Cup (DePuy), and, the Trilogy Multi-Holed Shell (Zimmer).
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
MAY 22 1997
Mr. Shavawn Parduhn
Regulatory Affairs Associate
Sulzer Orthopedics, Inc.
9900 Spectrum Drive
Austin, Texas 78717
Re: K970705
Inter-Op™ Porous Revision Shell with Sealable Screwholes
K970706
Inter-Op™ Porous Protrusio Shell with Sealable Screwholes
Regulatory Class: II
Product Codes: LPH and LZO
Dated: February 25, 1997
Received: February 26, 1997
Dear Mr. Parduhn:
We have reviewed your Section 510(k) notification of intent to market the devices referenced above and we have determined these devices are substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval), they may be subject to such additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your devices in the Federal Register. Please note: this response to your
{3}
Page 2 - Mr. Shavawn Parduhn
premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in a classification for your devices and thus, permits your devices to proceed to the market.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Enclosures
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
MAY 22 1997
Mr. Shavawn Parduhn
Regulatory Affairs Associate
Sulzer Orthopedics, Inc.
9900 Spectrum Drive
Austin, Texas 78717
Re: K970705
Inter-Op™ Porous Revision Shell with Sealable Screwholes
K970706
Inter-Op™ Porous Protrusio Shell with Sealable Screwholes
Regulatory Class: II
Product Codes: LPH and LZO
Dated: February 25, 1997
Received: February 26, 1997
Dear Mr. Parduhn:
We have reviewed your Section 510(k) notification of intent to market the devices referenced above and we have determined these devices are substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval), they may be subject to such additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your devices in the Federal Register. Please note: this response to your
{5}
Page 2 - Mr. Shavawn Parduhn
premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in a classification for your devices and thus, permits your devices to proceed to the market.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Enclosures
{6}
Page 1 of 1
510(k) Number (if known): Unknown
Device Name: Inter-Op™ Porous Revision Shell with Sealable Screwholes
## Indications For Use:
The Inter-Op Porous Revision Shell with Sealable Screwholes is a porous coated hemispherical shell that is used in conjunction with a snap-in acetabular insert manufactured from ultra-high molecular weight polyethylene (UHMWPE). This device is intended to replace the acetabulum during total hip arthroplasty and may be implanted with or without bone cement. Diagnostic indications for use of this device include:
- revision of a previously implanted acetabular prosthesis;
- patient conditions of noninflammatory degenerative joint disease; e.g., avascular necrosis, osteoarthritis, or arthritis secondary to a variety of diseases and anomalies; and,
- inflammatory joint disease; e.g., rheumatoid arthritis.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use ☑
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
14
2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.