DEPUY GRADED POROSITY AML HIP (1371-03/06-500/1371-43/46-500)
Device Facts
| Record ID | K962923 |
|---|---|
| Device Name | DEPUY GRADED POROSITY AML HIP (1371-03/06-500/1371-43/46-500) |
| Applicant | Depuy, Inc. |
| Product Code | LPH · Orthopedic |
| Decision Date | Oct 1, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3358 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Graded Porosity AML is indicated for uncemented or cemented use as the femoral component in total hip arthroplasty (THA) for replacing the hip joints of patients whose hip joint has been damaged by inflammatory or non-inflammatory degenerative joint disease, fracture or the failure of a previous arthroplasty.
Device Story
Femoral component for total hip arthroplasty; used in cemented or uncemented procedures. Device consists of cast Co-Cr-Mo femoral stem with graded porous coating. Coating architecture features five zones of Co-Cr-Mo beads; bead size increases proximally to distally. Stem sintered and HIPed to bond beads. Available in four standard and four modified medial aspect (MMA) sizes. Implanted by orthopedic surgeons to replace damaged hip joints; restores joint function and stability.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, material, and technological similarities to predicate devices.
Technological Characteristics
Material: Cast Co-Cr-Mo alloy. Design: Femoral hip prosthesis with graded porous coating. Coating: Multi-layered Co-Cr-Mo beads of varying sizes; 5 zones proximally to distally. Manufacturing: Sintered and Hot Isostatic Pressed (HIPed). Configuration: Available in 4 standard and 4 modified medial aspect (MMA) sizes.
Indications for Use
Indicated for patients requiring total hip arthroplasty due to inflammatory or non-inflammatory degenerative joint disease, fracture, or failure of a previous arthroplasty.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
- DePuy 5/8 Porous Coated AML Hip Prosthesis
- DePuy Proximally Porous Coated AML
- DePuy Solution Hip Prosthesis
Related Devices
- K030079 — VERSYS BEADED FULLCOAT BOWED REVISION HIP PROSTHESIS · Zimmer, Inc. · Feb 5, 2003
- K974740 — DEPUY TRI-LOCK HIP STEM · Depuy, Inc. · Mar 18, 1998
- K973714 — VERSYS HIP SYSTEM BEADED HIP PROSTHESIS · Zimmer, Inc. · Dec 24, 1997
- K000207 — POROCOAT PRODIGY HIP PROSTHESIS · DePuy Orthopaedics, Inc. · Feb 4, 2000
- K061833 — DEPUY AML HIP PROSTHESIS · DePuy Orthopaedics, Inc. · Jul 17, 2006