RESTORATION ACETABULAR WEDGE AUGMENTS

K102019 · Howmedica Osteonics · LPH · Mar 3, 2011 · Orthopedic

Device Facts

Record IDK102019
Device NameRESTORATION ACETABULAR WEDGE AUGMENTS
ApplicantHowmedica Osteonics
Product CodeLPH · Orthopedic
Decision DateMar 3, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Restoration Acetabular Wedge Augments are intended for use as an alternative to structural allograft in cases of superior and superior/posterior segmental acetabular deficiencies. General Indications for Total Hip Replacement Components: - Painful, disabling joint disease of the hip resulting from: degenerative arthritis, . rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis. - Revision of previous unsuccessful femoral head replacement, cup arthroplasty or . other procedure. - Clinical management problems where arthrodesis or alternative reconstructive . techniques are less likely to achieve satisfactory results. - Where bone stock is of poor quality or is inadequate for other reconstructive . techniques as indicated by deficiencies of the acetabulum. Indications Specific to the Acetabular Wedges: - As an alternative to structural allograft in cases of superior and superior/posterior . segmental acetabular deficiencies. Acetabular Augments are intended for cementless use only to the bone interface; and 2) are affixed to the mating cup using bone cement

Device Story

Restoration Acetabular Wedge Augments are orthopedic implants designed to address acetabular bone deficiencies during hip revision surgery. The augments provide structural support for the acetabular shell in cases of superior and superior/posterior defects. Fabricated from commercially pure (CP) titanium porous sintered foam, the devices are available in various geometries and sizes to accommodate different defect ranges. The augments are intended for cementless application at the bone interface and are affixed to the mating acetabular cup using bone cement. They are compatible with specific Howmedica Osteonics uncemented shells (Restoration ADM, Secur-Fit, Trident, Tritanium) and Trident all-polyethylene cemented cups. The device is used by orthopedic surgeons in a clinical/surgical setting to restore acetabular bone stock, potentially improving implant stability and patient outcomes in complex revision scenarios.

Clinical Evidence

No clinical data. Substantial equivalence was established through non-clinical laboratory testing, including tensile, compression, and construct fatigue testing of the augment/shell construct, as well as characterization of the CP titanium porous sintered foam (volume porosity, pore size, grain size).

Technological Characteristics

Materials: Commercially pure (CP) titanium porous sintered foam. Sensing/Actuation: None (passive implant). Energy Source: None. Connectivity: None. Sterilization: Not specified. Form Factor: Various geometries/sizes for acetabular defect filling. Testing performed per FDA Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone Cement (1994).

Indications for Use

Indicated for patients with painful, disabling hip joint disease (degenerative, rheumatoid, post-traumatic arthritis, or late-stage avascular necrosis), patients requiring revision of unsuccessful prior hip procedures, or patients with inadequate bone stock/acetabular deficiencies where other reconstructive techniques are insufficient. Specific use for superior and superior/posterior segmental acetabular deficiencies as an alternative to structural allograft.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K102019 MAR - 3 2011 #### 510(k) Summary | Sponsor | Howmedica Osteonics Corp.<br>325 Corporate Drive<br>Mahwah, NJ 07430 | |----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Karen Ariemma<br>Project Manager, Regulatory Affairs/Regulatory Compliance<br>Howmedica Osteonics Corp.<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Phone: (201) 831-5718 | | Date Prepared: | February 14, 2011 | | Proprietary Name: | Restoration Acetabular Wedge Augments | | Common Name: | Acetabular Bone Augments | | Classification Name: | 21 CFR §888.3358<br>Prosthesis, hip, semi-constrained, metal/polymer, porous<br>uncemented | Legally Marketed Device to Which Substantial Equivalence is Claimed: Zimmer Trabecular Metal Augments K061067 and the Osteonics Acetabular Wedge System K971422. ### Device Description: Restoration Acetabular Wedge Augments provide options to address a wide range of bone deficiencies encountered in acetabular revision. The augments provide support for the shell in the acetabulum with superior and/or posterior defects. The augments are offered in various geometries and sizes to address wide defect ranges. The augments are made of commercially pure titanium porous sintered foam. The Restoration Acetabular Wedge Augments can be used with Restoration ADM, Secur-Fit, Trident, and Tritanium uncemented shells. The Restoration Acetabular Wedge Augments can be used with Trident all-polyethylene cemented cups ### Intended Use: The Restoration Acetabular Wedge Augments are intended for use as an alternative to structural allograft in cases of superior and superior/posterior segmental acetabular deficiencies. ### Indications: # General Indications for Total Hip Replacement Components: - Painful, disabling joint disease of the hip resulting from: degenerative arthritis, . rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis. - Revision of previous unsuccessful femoral head replacement, cup arthroplasty or . other procedure. - Clinical management problems where arthrodesis or alternative reconstructive . {1}------------------------------------------------ p. 2/2 techniques are less likely to achieve satisfactory results. - Where bone stock is of poor quality or is inadequate for other reconstructive . techniques as indicated by deficiencies of the acetabulum. K102019 Indications Specific to the Acetabular Wedges: - As an alternative to structural allograft in cases of superior and superior/posterior segmental 0 acetabular deficiencies. Acetabular Augments are intended for cementless use only to the bone interface; and 2) are affixed to the mating cup using bone cement Summary of Technologies: Device comparison showed that the proposed device is substantially equivalent in intended, use, materials and performance characteristics to the predicate device. Non-Clinical Testing: Non-clinical laboratory testing was performed for the CP Titanium porous sintered foam to determine substantial equivalence. Non-clinical testing was provided as outlined in the FDA Guidance Document entitled "Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone Cement (28 April 1994)". The following mechanical testing was performed: tensile, compression, and construct fatigue testing of the augment/shell construct. The following parameters of the CP Titanium porous sintered foam were determined: volume porosity, average pore size, mean intercept length and grain size. In addition, static compression strength, abrasion and morphological properties were determined post fatigue testing. There was no affect on mechanical morphological properties based on fatigue loading. The testing demonstrated that the Restoration Acetabular Wedge Augments are substantially equivalent to devices currently cleared for market. Clinical Testing: Clinical testing was not required for this submission. Conclusion: The Restoration Acetabular Wedge Augments are substantially equivalent to the predicate devices identified in this premarket notification. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAR - 3 2011 Howmedica Osteonics Corp. % Ms. Karen Ariemma Project Manager, Regulatory Affairs and Regulatory Compliance 325 Corporate Drive Mahwah, New Jersey 07430 Re: K102019 Trade/Device Name: Restoration Acetabular Wedge Augments Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis . Regulatory Class: Class II Product Code: LPH Dated: February 14, 2011 Received: February 16, 2011 Dear Ms. Ariemma: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Or to and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the 1.0 The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contractly and warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ Page 2 - Ms. Karen Ariemma or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting on modical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the election is product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. 305%m lase note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR, Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, A. B. n h for Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 102019 510(k) Number (if known): K102019 Device Name: Restoration Acetabular Wedge Augments Indications for Use: The indications for use of the Restoration Acetabular Wedge Augments: General Indications for Total Hip Replacement Components: - Painful, disabling joint disease of the hip resulting from: degenerative arthritis, . rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis. - Revision of previous unsuccessful femoral head replacement, cup arthroplasty or . other procedure. - Clinical management problems where arthrodesis or alternative reconstructive . techniques are less likely to achieve satisfactory results. - Where bone stock is of poor quality or is inadequate for other reconstructive . techniques as indicated by deficiencies of the acetabulum. Indications Specific to the Acetabular Wedges: - As an alternative to structural allograft in cases of superior and superior/posterior . segmental acetabular deficiencies. Acetabular Augments are intended for cementless use only to the bone interface; and 2) are affixed to the mating cup using bone cement Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 for M.melkerson Division Sign Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K102019
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...