RIOPRO COBALT TRAPEZIOMETARCARPAL RESURFACING COMPONENT

K964472 · Biologically Oriented Prostheses · KYI · Feb 3, 1997 · Orthopedic

Device Facts

Record IDK964472
Device NameRIOPRO COBALT TRAPEZIOMETARCARPAL RESURFACING COMPONENT
ApplicantBiologically Oriented Prostheses
Product CodeKYI · Orthopedic
Decision DateFeb 3, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3770
Device ClassClass 2
AttributesTherapeutic

Intended Use

The BIOPRO Cobalt Trapeziometacarpal Replacement is intended to resurface the joint in cases of rheumatoid arthritis, traumatic arthritis, osteoarthritis, or post fracture deformation or bone loss which present themselves as either a painful, instable thumb or a thumb with limited range of motion.

Device Story

One-piece resurfacing implant for basal thumb joint; tri-flanged stem for medullary canal fixation; spherical head (20° varus angle) articulates with resurfaced trapezium and second metacarpal. Designed for surgical implantation by orthopedic surgeons to treat arthritis or post-traumatic thumb dysfunction. Provides pain relief and restores joint motion. Tri-flanged stem design prevents rotation; spherical head geometry optimizes range of motion; resurfacing of second metacarpal prevents impingement and dislocation. Cobalt chrome material provides harder articulating surface compared to titanium alternatives.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Material: Cast Cobalt Chrome (ASTM F75). Design: One-piece resurfacing implant; tri-flanged stem; spherical head with 20° varus angle. Function: Resurfaces basal thumb joint and second metacarpal. Non-software device.

Indications for Use

Indicated for patients with rheumatoid arthritis, traumatic arthritis, osteoarthritis, or post-fracture deformation/bone loss presenting with painful, unstable thumbs or limited range of motion.

Regulatory Classification

Identification

A wrist joint carpal trapezium polymer prosthesis is a one-piece device made of silicone elastomer or silicone elastomer/polyester material intended to be implanted to replace the carpal trapezium bone of the wrist.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 510(k) Summary K964472 FEB - 3 1997 Submitter: BIOPRO 17 Seventeenth Street Port Huron, MI 48060 (810) 982-7777 Contact: Valerie Gardner Date: October 25, 1996 Device Name: BIOPRO Cobalt Trapeziometacarpal Replacement Classification: The FDA has not yet created a classification for this type of device. Predicate Device: Swanson Titanium Condylar Implant (K864488) The BIOPRO Cobalt Trapeziometacarpal Replacement is a one-piece resurfacing component for the basal thumb joint. It has an anatomically configured tri-flanged stem that fits into the medullary canal at the proximal end of the first metacarpal. This tri-flanged configuration allows for compaction of bone around the implant to provide stable fixation and to prevent rotation of the device. The head is spherical and extends just beyond a hemisphere, slightly medialized on the stem and is at a 20° varus angle to the stem to properly space the joint and allow for sufficient range of motion. The trapezium is resurfaced to contain the head of the implant and this resurfacing is continued into the second metacarpal. Resurfacing the second metacarpal prevents impingement of the implant against that bone and decreases the possibility of dislocation of the joint. The BIOPRO Cobalt Trapeziometacarpal Replacement is intended to resurface the joint in cases of rheumatoid arthritis, traumatic arthritis, osteoarthritis, or post fracture deformation or bone loss which present themselves as either a painful, instable thumb or a thumb with limited range of motion. These indications are the same as those for the predicate device, the Swanson Titanium Condylar Implant (K864488). The other similarities, and also the differences, between the subject device and the predicate device include their functions, materials and shapes. The components function the same; they both resurface the joint to provide pain relief and increase motion in the joint. Both components have a hemispherical head, however the ratio of head height to head diameter is slightly larger for the Swanson device than for the BIOPRO component. Also, while the head of the Swanson device is cylindrical below the hemisphere, the BIOPRO head continues the spherical radius to the edge of the head to increase the range of motion of the joint. Both devices' stems are intended to prevent rotation; however, their cross-sections are different. The Swanson device has a square cross-section and the BIOPRO device has a tri-flanged cross-section. The only other difference between the devices is the material. The BIOPRO component is manufactured from cast Cobalt Chrome (to ASTM F75) while the Swanson device Page 5 {1} is manufactured from unalloyed titanium (to ASTM F 67). The use of cobalt as opposed to titanium allows for casting of the device. Also, cobalt chrome is not as soft a material as titanium which better suits the metal for use as an articulating surface. Page 6
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...