The CONSERVE® Femoral Resurfacing Component is indicated for use in hemi resurfacing for reduction or relief of pain and/or improved hip function in skeletally mature patients with non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusion acetabuli, and painful hip dysplasia. The CONSERVE® Femoral Resurfacing Component is indicated for cemented use only.
Device Story
CONSERVE® Femoral Resurfacing Component is a metallic hemi-hip resurfacing prosthesis; used for hip joint reconstruction. Device replaces femoral head surface to reduce pain and improve function in patients with degenerative joint disease. Implanted by orthopedic surgeons in clinical/surgical settings. Device manufactured from Cobalt Chrome Alloy; available in 11 sizes with 2mm increments. Requires cement for fixation. Provides mechanical structural support to the femoral head.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, material, and technological characteristics.
Indicated for skeletally mature patients requiring hemi-resurfacing for pain relief or improved hip function due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, ankylosis, protrusion acetabuli, painful hip dysplasia). Cemented use only.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metallic resurfacing prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral resurfacing component made of alloys, such as cobalt-chromium-molybdenum.
K021799 — CEMENTED FEMORAL HEAD RESURFING DEVICE · Biomet Orthopedics, Inc. · Jun 26, 2002
K043542 — BIOPRO TARA FEMORAL RESURFACING COMPONENT · Biopro, Inc. · May 18, 2005
K032659 — DEPUY ASR RESURFACING FEMORAL HEADS · DePuy Orthopaedics, Inc. · Dec 4, 2003
K023188 — PRESS-FIT HEAD RESURFACING DEVICE · Biomet Orthopedics, Inc. · Dec 11, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
062960
p. 1/2
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS DEC - 1 2006
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CRF 807, this information serves as a Summary of Safety and Effectiveness for the use of the CONSERVE® Femoral Resurfacing Component.
| Submitted By: | Wright Medical Technology, Inc. |
|-------------------------------------|-----------------------------------------------------------------------------------------|
| Date: | September 28, 2006 |
| Contact Person: | Theresa Leister |
| | Regulatory Affairs Specialist II |
| Proprietary Name: | CONSERVE® Femoral Resurfacing Component |
| Common Name: | Hemi-resurfacing femoral component |
| Classification Name and Reference: | 21 CFR 888.3400 Hip joint femoral (hemi-hip) metallic resurfacing prosthesis – Class II |
| Device Product Code and Panel Code: | Orthopedics/87/ KXA |
### DEVICE INFORMATION
### A. INTENDED USE
The CONSERVE® Femoral Resurfacing Component is indicated for use in hemi resurfacing for reduction or relief of pain and/or improved hip function in skeletally mature patients with non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusion acetabuli, and painful hip dysplasia.
The CONSERVE® Femoral Resurfacing Component is indicated for cemented use only.
## B. DEVICE DESCRIPTION
The design features of the CONSERVE® Femoral Resurfacing Component are summarized below:
- Manufactured from Cobalt Chrome Alloy .
- 11 Sizes in 2mm increments .
{1}------------------------------------------------
062960
\$p^{2}/2\$
## C. SUBSTANTIAL EQUIVALENCE INFORMATION
The indications for use and design of the CONSERVE® Femoral Resurfacing Component are identical to Orthomet Resurfacing Femoral Component. The materials of the CONSERVE® Femoral Resurfacing Component are substantially equivalent to those of the predicate device. The fundamental scientific technology of the modified device has not changed relative to the predicate device. The safety and effectiveness of the CONSERVE® Femoral Resurfacing Component are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES - USA
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Wright Medical Technology, Inc. % Ms. Theresa Leister Regulatory Affairs Specialist II 5677 Airline Road Arlington, Tennessee 38002
DEC - 1 2006
Re: K062960
Trade Name: CONSERVE® Femoral Resurfacing Component Regulation Number: 21 CFR 888.3400 Regulation Name: Hip joint femoral (hemi-hip) metallic resurfacing prosthesis Regulatory Class: Class II Product Code: KXA Dated: November 2, 2006 Received: November 3, 2006
Dear Ms. Leister:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Theresa Leister
This letter will allow you to begin marketing your device as described in your Scction 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Mel Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: _CONSERVE® Femoral Resurfacing Component
Indications For Use:
The CONSERVE® Femoral Resurfacing Component is indicated for use in hemi resurfacing for reduction or relief of pain and/or improved hip function in skeletally mature patients with noninflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusion acetabuli, and painful hip dysplasia.
The CONSERVE® Femoral Resurfacing Component is indicated for cemented use only.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbere Brielm for Man
Page 1 of 1
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
K062960 ೆ 10(k) Number
1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.