ACROMED PEDICLE SCREW
Device Facts
| Record ID | K963639 |
|---|---|
| Device Name | ACROMED PEDICLE SCREW |
| Applicant | Acromed Corp. |
| Product Code | KWP · Orthopedic |
| Decision Date | Nov 14, 1996 |
| Decision | SN |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3050 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The ISOLA Spinal System, when used with pedicle screws, is intended for use in grade 3 or 4 spondylolisthesis at L5-S1 utilizing autologous bone graft and intended to be removed after solid fusion is attained. The ISOLA Spinal System, when not used with pedicle screws, is intended for use in grade 3 or 4 spondylolisthesis at L5-S1 utilizing autologous bone graft and intended to be removed after solid fusion is attained.
Device Story
AcroMed Pedicle Screw (4.75 mm) is a spinal fixation component for the ISOLA Spine System. Device features two-section design: long cancellous section with integral fixed lower nut; machine-threaded section with hexagonal drive head. Integral nut provides bolt connection strength and reduces stress concentration. Supplied with adjustable tapered nut and optional lock nut for rod-to-screw connection. Used by surgeons in spinal fusion procedures to stabilize L5-S1 segments in spondylolisthesis patients. Removed after solid fusion achieved.
Clinical Evidence
Bench testing only. Static and fatigue testing performed to demonstrate consistency with previously cleared components.
Technological Characteristics
Material: Implant grade stainless steel (ASTM F1314). Dimensions: 4.75 mm diameter; cancellous thread lengths 25-50 mm (5 mm increments). Design: Two-section screw with integral fixed lower nut, machine-threaded section, and hexagonal drive head. Mechanical fixation.
Indications for Use
Indicated for patients with grade 3 or 4 spondylolisthesis at L5-S1 requiring spinal fixation with autologous bone graft.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
Predicate Devices
- AcroMed pedicle screws (K951116)
Reference Devices
- ISOLA Spine System (K944737)
Related Devices
- K965102 — ACROMED PEDICLE SCREW · Acromed Corp. · May 8, 1997
- K964007 — ACROMED CABLE/WIRE PEDICLE SCREW · Acromed Corp. · Dec 16, 1996
- K152098 — ODALYS Pedicle Screw System · Kisco International · Sep 17, 2015
- K023438 — MODIFICATION TO MONARCH SPINE SYSTEM · Depuyacromed · Nov 13, 2002
- K964533 — EBI SPINELINK SYSTEM · Electro-Biology, Inc. · Apr 9, 1997