SP-FIX SPINOUS PROCESS FIXATION PLATE

K102195 · Globus Medical, Inc. · KWP · Jan 19, 2011 · Orthopedic

Device Facts

Record IDK102195
Device NameSP-FIX SPINOUS PROCESS FIXATION PLATE
ApplicantGlobus Medical, Inc.
Product CodeKWP · Orthopedic
Decision DateJan 19, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3050
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SP-Fix™ Spinous Process Fixation Plate is a posterior non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), and/or tumor. The SP-Fix™ Spinous Process Fixation Plate is intended for use with bone graft material and is not intended for stand-alone use.

Device Story

SP-Fix™ Spinous Process Fixation Plate is a posterior non-pedicle supplemental fixation system; provides stabilization of spinal segments to support fusion. Components include plates, rods, and barrels; available in various sizes to accommodate patient anatomy. Used in surgical settings by surgeons; attached to spinous processes to supplement fusion constructs. Not for stand-alone use; requires concurrent bone graft material. Benefits patients by providing supplemental stability in degenerative, traumatic, or oncological spinal conditions.

Clinical Evidence

Bench testing only. Mechanical performance was validated through static and dynamic compression, static torsion, static and dynamic plate dissociation, and static tension testing to demonstrate substantial equivalence.

Technological Characteristics

Constructed from titanium alloy (ASTM F136) and PEEK radiolucent polymer (ASTM F2026). Posterior non-pedicle fixation system consisting of plates, rods, and barrels. Mechanical fixation principle.

Indications for Use

Indicated for patients requiring supplemental spinal fusion at a single level in the non-cervical spine (T1-S1) due to degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), or tumor. Not for stand-alone use; must be used with bone graft material.

Regulatory Classification

Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summary: SP-Fix™ Spinous Process Fixation Plate | Company: | Globus Medical Inc. | | | |----------|-----------------------------------------|-----------|--| | | 2560 General Armistead Ave. | | | | | Audubon, PA 19403 | | | | | (610) 930-1800 | JAN 19 20 | | | Contact: | Kelly J. Baker, Ph.D | | | | | Director, Clinical Affairs & Regulatory | | | Date Prepared: November 10, 2010 SP-Fix™ Spinous Process Fixation Plate Device Name: - Classification: Regulation Number: 21 CFR §888.3050 Regulation Name: Spinal Interlaminal Fixation Orthosis Product Code: KWP Regulatory Class II, Panel code 87. - Medtronic Spire™ (K032037), LANX Aspen™ (K071877), Predicate(s): NuVasive AFFIX™ (K073278) ## Purpose: The purpose of this submission is the addition of the SP-Fix™ Spinous Process Fixation Plate to the REVERE® Stabilization System. ## Device Description: The SP-Fix™ Spinous Process Fixation Plate consists of plates, rods and barrels that are used to build a construct to provide supplemental stabilization of spinal segments to support fusion. The components are available in a range of sizes to fit the anatomical needs of a variety of patients. SP-Fix™ implants are composed of titanium alloy (per ASTM F136) and PEEK radiolucent polymer (per ASTM F2026). ## Indications for Use: The SP-Fix™ Spinous Process Fixation Plate is a posterior non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), and/or tumor. The SP-Fix™ Spinous Process Fixation Plate is intended for use with bone graft material and is not intended for stand-alone use. {1}------------------------------------------------ ## Performance Data: Mechanical testing (static and dynamic compression, static torsion, static and dynamic plate dissociation, and static tension) was conducted to demonstrate substantial equivalence to the predicate system(s). ## Basis of Substantial Equivalence: The SP-Fix™ has been demonstrated to be substantially equivalent to predicate system(s) with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence of the subject device to the predicate device(s). {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the eagle. The text is in all caps and is in a simple, sans-serif font. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JAN 19 2011 Globus Medical, Inc. % Kelly J. Baker, Ph.D. Director, Clinical Affairs and Regulatory Valley Forge Business Center 2560 General Armistead Avenue Audubon, Pennsylvania 19403 Re: K102195 Trade Name: SP-Fix™ Spinous Process Fixation Plate Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: KWP Dated: December 30, 2010 Received: January 03, 2011 Dear Dr. Baker: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 – Kelly J. Baker, Ph.D. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use Statement 510(k) Number: K102195 Device Name: SP-Fix™ Spinous Process Fixation Plate #### INDICATIONS: The SP-Fix™ Spinous Process Fixation Plate is a posterior non-pedicle supplemental fixation device, intended for use at a single level in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous processes for the purpose of achieving supplemental fusion in the following conditions: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), and/or tumor. The SP-Fix™ Spinous Process Fixation Plate is intended for use with bone graft material and is not intended for stand-alone use. OR Prescription Use × (Per 21 CFR §801.109) Over-The-Counter Use 1 (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Digit Sign Off (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K102195 Globus Medical K102195 SP-Fix™ CONFIDENTIAL
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...