K023086 · Nexa Orthopedics, Inc. · KWH · Oct 4, 2002 · Orthopedic
Device Facts
Record ID
K023086
Device Name
CLASSIC GREAT TOE IMPLANT
Applicant
Nexa Orthopedics, Inc.
Product Code
KWH · Orthopedic
Decision Date
Oct 4, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3720
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Futura Biomedical Classic Great Toe Implant is indicated for: - Hallux limitus or hallux rigidus . - . Painful rheumatoid arthritis - Hallux abducto valgus associated with arthritis ● - Unstable or painful joint from previous surgery .
Device Story
Classic Great Toe Implant is a double-stemmed silicone prosthesis; functions as a dynamic joint spacer for first metatarsophalangeal joint arthroplasty. Implanted surgically by physician; stems placed into intramedullary canals of metatarsal and proximal phalanx. Device provides constrained joint support; intended for long-term implantation (>30 days). Benefits include pain relief and joint stabilization for patients with degenerative or post-surgical toe conditions.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, material, and intended use comparison to predicate device.
Technological Characteristics
Double-stemmed silicone prosthesis; constrained design; four sizes (20, 30, 40, 50); intended for intramedullary canal fixation.
Indications for Use
Indicated for patients with hallux limitus, hallux rigidus, painful rheumatoid arthritis, hallux abducto valgus associated with arthritis, or unstable/painful joints resulting from previous surgery.
Regulatory Classification
Identification
A toe joint polymer constrained prosthesis is a device made of silicone elastomer or polyester reinforced silicone elastomer intended to be implanted to replace the first metatarsophalangeal (big toe) joint. This generic type of device consists of a single flexible across-the-joint component that prevents dislocation in more than one anatomic plane.
Predicate Devices
Futura Biomedical Flexible Great Toe Implant (k981194)
Related Devices
K023562 — MODIFICATION TO CLASSIC GREAT TOE IMPLANT · Nexa Orthopedics, Inc. · Nov 13, 2002
K023531 — LESSER METATARSAL PHALANGEAL JOINT IMPLANT · Nexa Orthopedics, Inc. · Jan 17, 2003
K981194 — FUTURA BIOMEDICAL FLEXIBLE GREAT TOE IMPLANT · Nexa Orthopedics, Inc. · Jun 30, 1998
K191966 — NewPrim System · Brm Extremities · Feb 13, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
# OCT 0 4 2002
## 510 (k) Summary
Summary of 510 (k) safety and effectiveness information upon which the subs determination is based:
| Prepared: | September 13, 2002 |
|-----------|--------------------|
|-----------|--------------------|
- Futura Biomedical Applicant: 990 Park Center Drive, Suite H Vista, CA 92083
| Telephone: | 760-599-1670 |
|------------|-----------------|
| Fax: | 760-599-1675 |
| Contact: | Louise M. Focht |
| Device Name: | Prosthesis, Toe, Constrained, Polymer |
|----------------------------|---------------------------------------------------------|
| Device Trade Name: | Classic Great Toe Implant |
| Device Classification: | Class II |
| Reviewing Panel: | Orthopedic |
| Regulation Number | 888.3720 |
| Product Code: | 87 KWH |
| Original Predicate Device: | Futura Biomedical Flexible Great Toe<br>Implant-k981194 |
| Registration Number: | 2030833 |
| Owner Operator Number: | 9028319 |
#### Device Description:
The Classic Great Toe Implant is a double-stemmed silicone prosthesis, intended to supplement first metatarsophalangeal joint srthroplasty. The implant is designed to act as a dynamic joint spacer between the resected head of the first metatarsal and base of the proximal phalanx.
## Indications for Use:
Futura Biomedical Classic Great Toe Implant is indicated for:
- Hallux limitus or hallux rigidus .
- . Painful rheumatoid arthritis
- Hallux abducto valgus associated with arthritis ●
- Unstable or painful joint from previous surgery .
{1}------------------------------------------------
Comparison to the Original Predicate Device:
The legally marketed predicate device to which this device is substantially equivalent is the Futura Biomedical Flexible Great Toe Implant
Regulatory Class: II Product Code: 87 KWH
Comparison of original Futura Biomedical Flexible Great Toe Implant to the new configuration the Classic Great Toe Implant.
| Item | Original Futura Product | Proposed product configuration |
|---------------------|----------------------------------------------------|----------------------------------------------------|
| Product Name | Flexible Great Toe Implant | Classic Great Toe Implant |
| Use | Single use | Single use |
| Fixation | Stem in intramedullary canal | stem in intramedullary canal |
| Constraint | constrained | constrained |
| Material | Silicone | Silicone |
| Sizes | 4 sizes, 20, 30, 40, 50 | 4 sizes 20, 30, 40, 50 |
| Indications for use | Hallux limitus or hallux rigidus | Hallux limitus or hallux rigidus |
| | Painful rheumatoid arthritis | Painful rheumatoid arthritis |
| | Hallux abducto valgus associated<br>with arthritis | Hallux abducto valgus associated<br>with arthritis |
| | Unstable or painful joint from<br>previous surgery | Unstable or painful joint from<br>previous surgery |
Similarities of the Futura Flexible Great Toe Implant and Classic Great Toe Implant include; Both devices are intended for single use only; Both devices are intended for surgical implantation longer than 30 days; Both devices are placed into the intramedullary canal of the metatarsal and phalangeal bones; Both devices are made of the same industry standard materials. No new materials are introduced in either product; Both devices are comparably sized; Both devices have the identical for use.
## Summary:
The device and the predicate device have similar design characteristics and intended use. The new device is substantially equivalent to the predicate device.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling three stylized human figures or birds in flight, stacked vertically.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 0 4 2002
Mr. Louise M. Focht Futura Biomedical 990 Park Center Drive, Suite H Vista, California 92083
Re: K023086 Trade Name: Classic Great Toe Implant Regulation Number: 21 CFR 888.3720 Regulation Name: Toe joint polymer constrained prosthesis Regulatory Class: II Product Code: KWH Dated: September 13, 2002 Received: September 17, 2002
Dear Mr. Focht:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Louise M. Focht
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Q. Mark N. Millbern
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page of
510 (k) Number (If Known):_ Device Name: Classic Great Toe Implant
# Indications for Use:
Futura Biomedical Classic Great Toe Implant is indicated for:
- Hallux limitus or hallux rigidus .
- Painful rheumatoid arthritis ●
- Hallux abducto valgus associated with arthritis ●
- Unstable or painful joint from previous surgery ●
Prescription Use
Yes/No
Or
Over the counter use
Yes/No
Mark A. Milken
(Division Sign-Of Division of General in storative and Neurological L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5. Number K023086
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.