Smith & Nephew, Inc. ANTHEM Total Knee System, Genesis Uni Knee System, JOURNEY BCS and II Knee Systems, JOURNEY II CR Knee System, JOURNEY II Uni Knee System, Journey II UK Knee System, JOURNEY II XR Bi-Cruciate Retaining Knee System, JOURNEY PFJ System, JOURNEY Uni Knee System, Legion Hinge Total Knee System, Legion Total Knee System, ZUK Unicompartmental Knee System, ENGAGE Partial Knee System and Porous Patella and Porous Tibial Baseplates
K230653 · Smith & Nephew, Inc. · JWH · May 4, 2023 · Orthopedic
Device Facts
Record ID
K230653
Device Name
Smith & Nephew, Inc. ANTHEM Total Knee System, Genesis Uni Knee System, JOURNEY BCS and II Knee Systems, JOURNEY II CR Knee System, JOURNEY II Uni Knee System, Journey II UK Knee System, JOURNEY II XR Bi-Cruciate Retaining Knee System, JOURNEY PFJ System, JOURNEY Uni Knee System, Legion Hinge Total Knee System, Legion Total Knee System, ZUK Unicompartmental Knee System, ENGAGE Partial Knee System and Porous Patella and Porous Tibial Baseplates
Applicant
Smith & Nephew, Inc.
Product Code
JWH · Orthopedic
Decision Date
May 4, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ANTHEM PS Total Knee System is indicated for use only with cement and is a single use device. Unicompartmental knee implants are indicated for restoring either compartment of a knee that has been affected by the following: 1. Noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; 2. Correction of functional deformity; 3. Revision procedures where other treatments or devices have failed: and 4. Treatment of fractures that are unmanageable using other techniques. Unicompartmental knee implants are single use only and are intended for implantation only with bone cement. HA coated unicompartmental knee implants are available outside the US for use without bone cement. Smith & Nephew, Inc. Journey II BCS Knee components are indicated for use only with cement and are single use devices. The Journey II Total Knee system components are indicated for use only with cement and are single use devices. The Smith & Nephew Patello-Femoral Implants are intended for implantation with bone cement. The Legion Hinge Knee System is for Cemented Use Only. The LEGION Knee System - Finned Tibial Wedges are for single use only and are intended for implantation with bone cement. The devices are indicated for cemented use only. The Engage™ Partial Knee System is intended for medial unicompartmental knee arthroplasty to treat one or more of the following conditions: Moderately disabling joint disease of the knee resulting from painful osteo- or post traumatic arthritis. Revision of previous unsuccessful surgical procedures, including prior unicompartmental knee arthroplasty. As an alternative to tibial osteotomy in patients with Unicompartmental osteoarthritis. The femoral component and tibial tray are intented fixation. The porous surfaces of both the femoral and tibial tray components provide biological fixation when used in a cementless application. When the tibial tray is implanted without the use of bone cement, the tibial anchor should be used. When the tibial tray is implanted with bone cement, the tibial anchor should not be used. The Porous Patella and the Porous Tibia Baseplate are indicated for use with or without bone cement, and are single use devices.
Device Story
Smith & Nephew knee systems (total, unicompartmental, patello-femoral, hinge) are orthopedic implants used in primary and revision knee arthroplasty. These devices replace damaged joint surfaces to restore function and reduce pain. Implants are surgically placed by orthopedic surgeons in hospital or surgical center settings. This submission updates labeling to include Magnetic Resonance (MR) safety information. No changes were made to device design, materials, manufacturing, or sterilization. The update ensures clinicians have appropriate MR safety parameters for patients with these implants. Benefits include continued clinical utility of existing knee systems with clarified MR compatibility, facilitating safe diagnostic imaging for patients.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing for MR safety, including magnetically induced displacement force (ASTM F2052), torque (ASTM F2213), RF-induced heating (ASTM F2182-19e2, IEC 60601-2-33, ISO/TS 10974:2018E), and MR image artifact (ASTM F2119).
Technological Characteristics
Orthopedic knee implants (total, unicompartmental, hinge, patello-femoral). Materials and design are identical to previously cleared predicates. MR safety testing conducted per ASTM F2052, ASTM F2213, ASTM F2182-19e2, and ASTM F2119. Sterilization and manufacturing processes remain unchanged.
Indications for Use
Indicated for patients with rheumatoid arthritis, post-traumatic arthritis, osteoarthritis, degenerative arthritis, failed osteotomies, failed unicompartmental or total knee replacement, functional deformity, or fractures unmanageable by other techniques. Specific systems indicated for primary/revision surgery where cruciate/collateral ligaments are incompetent or absent. Patello-femoral system indicated for degenerative arthritis in distal femur/patella, patellar dislocation/fracture, or failed previous surgery.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Porous Patella and Porous Tibial Baseplates (K211221)
Submission Summary (Full Text)
{0}------------------------------------------------
May 4, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Smith & Nephew, Inc. Tyler Kulcsar Regulatory Affairs Specialist II 1450 East Brooks Rd. Memphis, Tennessee 38116
### Re: K230653
Trade/Device Name: Smith & Nephew, Inc. ANTHEM Total Knee System, Genesis Uni Knee System, JOURNEY BCS and II Knee Systems, JOURNEY II CR Knee System, JOURNEY II Uni Knee System, Journey II UK Knee System, JOURNEY II XR Bi-Cruciate Retaining Knee System, JOURNEY PFJ System, JOURNEY Uni Knee System, Legion Hinge Total Knee System, Legion Total Knee System, ZUK Unicompartmental Knee System, ENGAGE Partial Knee System and Porous Patella and Porous Tibial Baseplates Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, MBH, HSX, KRR, KRO, KWH, OIY Dated: March 8, 2023 Received: March 9, 2023
Dear Tyler Kulcsar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jesse Muir -S
Jesse Muir, Ph.D. Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K230653
Device Name Anthem Total Knee System
Indications for Use (Describe)
1. Rheumatoid arthritis.
2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
The ANTHEM PS Total Knee System is indicated for use only with cement and is a single use device.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{3}------------------------------------------------
Form Approved: OMB No. 0910-0120
510(k) Number (if known)
#### K230653
Device Name Genesis Uni Knee System
### Indications for Use (Describe)
Unicompartmental knee implants are indicated for restoring either compartment of a knee that has been affected by the following:
- 1. Noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis;
- 2. Correction of functional deformity;
- 3. Revision procedures where other treatments or devices have failed: and
- 4. Treatment of fractures that are unmanageable using other techniques
Unicompartmental knee implants are single use only and are intended for implantation only with bone cement. HA coated unicompartmental knee implants are available outside the US for use without bone cement.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known)
#### K230653
Device Name JOURNEY BCS and II Knee Systems
Indications for Use (Describe)
1. Rheumatoid arthritis.
2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
Smith & Nephew, Inc. Journey II BCS Knee components are indicated for use only with cement and are single use devices.
A recent study published in the journal Science has revealed a surprising connection between the consumption of dark chocolate and improved cognitive function. Researchers found that participants who consumed a small amount of dark chocolate daily for three months experienced significant improvements in memory, attention span, and overall cognitive performance, suggesting that the flavanols present in dark chocolate may have neuroprotective effects.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K230653
Device Name JOURNEY II CR Knee System
Indications for Use (Describe)
1. Rheumatoid arthritis.
2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
The Journey II Total Knee system components are indicated for use only with cement and are single use devices.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{6}------------------------------------------------
510(k) Number (if known)
#### K230653
Device Name JOURNEY II Uni Knee System Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
# Indications for Use (Describe)
Unicompartmental knee implants are indicated for restoring either compartment of a knee that has been affected by the following:
- 1. Noninflammatory degenerative joint disease including osteoarthritis, or avascular necrosis;
- 2. Correction of functional deformity;
- 3. Revision procedures where other treatments or devices have failed; and
- 4. Treatment of fractures that are unmanageable using other techniques
Unicompartmental knee implants are single use only and are intended for implantation only with bone cement. HA coated unicompartmental knee implants are available outside the US for use without bone cement.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{7}------------------------------------------------
**Use**
510(k) Number (if known)
### K230653
Device Name JOURNEY II UK Knee System
### Indications for Use (Describe)
Unicompartmental knee implants are indicated for restoring either compartment of a knee that has been affected by the following:
- · Non-inflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis:
- · Correction of functional deformity;
- · Revision procedures where other treatments or devices have failed; and
- · Treatment of fractures that are unmanageable using other techniques.
Unicompartmental knee implants are single use only and are intended for implantation only with bone cement.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{8}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K230653
Device Name JOURNEY II XR Bi-Cruciate Retaining Knee System
Indications for Use (Describe)
Total knee components are indicated for rheumatic arthritis, osteoarthritis, osteoarthritis, or degenerative arthritis; failed osteotomies, unicompartmental replacement, or total knee replacement.
The Journey II Total Knee system components are indicated for use only with cement and are single use devices.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{9}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K230653
Device Name JOURNEY PFJ System
Indications for Use (Describe)
1. Degenerative arthritis in the distal femur and patella;
2. A history of patellar dislocation or patellar fracture; and
3. Failed previous surgery (arthroplasty, tibial tubercule elevation, lateral release) where pain, deformity or dysfunction persists.
The Smith & Nephew Patello-Femoral Implants are intended for implantation with bone cement.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{10}------------------------------------------------
#### K230653
Device Name JOURNEY Uni Knee System
Indications for Use (Describe)
Unicompartmental knee implants are indicated for restoring either compartment of a knee that has been affected by the following:
- 1. Noninflammatory degenerative joint disease including osteoarthritis, or avascular necrosis;
- 2. Correction of functional deformity;
- 3. Revision procedures where other treatments or devices have failed; and
- 4. Treatment of fractures that are unmanageable using other techniques
Unicompartmental knee implants are single use only and are intended for implantation only with bone cement. HA coated unicompartmental knee implants are available outside the US for use without bone cement.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{11}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
### K230653
Device Name Legion Hinge Total Knee System
Indications for Use (Describe)
1. Rheumatoid arthritis.
2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
5. Constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral collateral ligament) are incompetent.
6. Hinge knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral ligament) are absent or incompetent.
The Legion Hinge Knee System is for Cemented Use Only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{12}------------------------------------------------
510(k) Number (if known)
### K230653
Device Name Legion Total Knee System
Indications for Use (Describe)
1. Rheumatoid arthritis.
2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
5. Constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral collateral ligament) are incompetent.
6. Hinge knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral ligament) are absent or incompetent.
The LEGION Knee System - Finned Tibial Wedges are for single use only and are intended for implantation with bone cement.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{13}------------------------------------------------
510(k) Number (if known)
#### K230653
Device Name ZUK Unicompartmental Knee System
Indications for Use (Describe) These devices are indicated for patients with:
- 1. Painful and/or disabling knee joints due to osteoarthritis or traumatic arthritis.
- 2. Previous tibial condyle or plateu fractures with loss of anatomy or function.
- 3. Varus or Valgus deformities.
- 4. Revision of previous arthroplasty procedures.
The devices are indicated for cemented use only.
| The Zimmer Unicompartmental Knee System is designed for use when load bearing ROM is |
|--------------------------------------------------------------------------------------|
| expected to be less than or equal to 155 degrees. |
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{14}------------------------------------------------
510(k) Number (if known)
#### K230653
Device Name The Engage™ Partial Knee System
#### Indications for Use (Describe)
The Engage™ Partial Knee System is intended for medial unicompartmental knee arthroplasty to treat one or more of the following conditions:
- Moderately disabling joint disease of the knee resulting from painful osteo- or post traumatic arthritis.
- Revision of previous unsuccessful surgical procedures, including prior unicompartmental knee arthroplasty
- As an alternative to tibial osteotomy in patients with Unicompartmental osteoarthritis.
The femoral component and tibial tray are intented fixation. The porous surfaces of both the femoral and tibial tray components provide biological fixation when used in a cementless application. When the tibial tray is implanted without the use of bone cement, the tibial anchor should be used. When the tibial tray is implanted with bone cement, the tibial anchor should not be used.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{15}------------------------------------------------
510(k) Number (if known)
#### K230653
Device Name Porous Patella and Porous Tibia Baseplate
Indications for Use (Describe) Total knee components are indicated for:
1. Rheumatoid arthritis.
2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
4. The posterior-stabilized knee system is designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
LEGION Porous CR Narrow Femoral Components are indicated for use without bone cement and are single use devices.
The Porous Patella and the Porous Tibia Baseplate are indicated for use with or without bone cement, and are single use devices.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;">☒</span> Prescription Use (Part 21 CER 801 Subpart D) |
|-----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CER 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{16}------------------------------------------------
# 510(k) Summary
| Submitted by: | Smith & Nephew, Inc. Orthopaedic Division<br>7135 Goodlett Farms Parkway Cordova,<br>Tennessee 38016 |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | |
| Date of Submission: | March 8, 2023 |
| | |
| Contact Person: | Tyler Kulcsar<br>Regulatory Affairs Specialist II<br>Tyler.Kulcsar@smith-nephew.com<br>Mobile: (732)- 678- 3967 |
| | Rose Beifuss<br>Senior Manager Regulatory Affairs<br>Rose.Beifuss@smith-nephew.com<br>Mobile: (385)-253-2551 |
| | |
| Name of Device: | Smith & Nephew, Inc. ANTHEM Total Knee System,<br>Genesis Uni Knee System, JOURNEY BCS and II Knee<br>Systems, JOURNEY II CR Knee System, JOURNEY II<br>Uni Knee System, Journey II UK Knee System,<br>JOURNEY II XR Bi-Cruciate Retaining Knee System,<br>JOURNEY PFJ System, JOURNEY Uni Knee System,<br>Legion Hinge Total Knee System, Legion Total Knee<br>System, ZUK Unicompartmental Knee System, ENGAGE<br>Partial Knee System and Porous Patella and Porous<br>Tibial Baseplates |
| Common Name: | Prosthesis, Knee, Patellofemorotibial, Semi-<br>Constrained, Cemented, Polymer/Metal/Polymer |
{17}------------------------------------------------
| Device Classification Name and<br>Reference: | 21 CFR 888.3560 – Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained cemented<br>prosthesis |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 21 CFR 888.3565 Knee joint patellofemorotibial<br>metal/polymer porous-coated uncemented prosthesis |
| | 21 CFR 888.3520 Knee joint femorotibial<br>metal/polymer non-constrained cemented prosthesis |
| | 21 CFR 888.3540 Knee joint patellofemoral<br>polymer/metal semi-constrained cemented prosthesis |
| | 21 CFR 888.3510 Knee joint femorotibial<br>metal/polymer constrained cemented prosthesis |
| | 21 CFR 888.3720 Toe joint polymer constrained<br>prosthesis |
| Device Class: | Class II |
| Panel Code: | Orthopaedics/87 |
| Product Code: | JWH, MBH, HSX, KRR, KRO, KWH, OIY |
| Predicate Device: | ANTHEM Total Knee System, Genesis Uni Knee<br>System, JOURNEY BCS and II Knee Systems,<br>JOURNEY II CR Knee System, JOURNEY II Uni Knee<br>System, JOURNEY II UK Knee System JOURNEY II<br>XR Bi-Cruciate Retaining Knee System, JOURNEY<br>PFJ System, JOURNEY Uni Knee System, Legion<br>Hinge Total Knee System, Legion Total Knee System,<br>ZUK Unicompartmental Knee System, ENGAGE Partial<br>Knee System and Porous Patella and Porous Tibial<br>Baseplates |
{18}------------------------------------------------
# Device Description:
The purpose of this Traditional 510(k) is to add MR safety information to the labeling (instructions for use, product labeling and patient implant card) of the subject Smith & Nephew Knee Systems. The technological characteristics, function of the devices, packaging and sterilization remain unchanged. As these products have been in distribution for several years, some parts have undergone changes in the technological characteristics that were assessed in compliance with the FDA's guidanceon "Deciding When to Submit a 510(k) for a Change to an Existing Device" or the FDA predecessor guidance on how to review changes to an existing device. These are not cumulative changes for all Smith & Nephew Knee Systems, but are rather specific changes that only affect certain parts. In the context of this submission, no modifications have been made to the device design, materials, sterilization, or the manufacturing process of the previously cleared devices due to the addition of MR Labeling for this 510(k).
# Indications for Use
# Legion Hinge
- 1. Rheumatoid arthritis.
- 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis,
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
- 4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
- 5. Constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral ligament) are incompetent.
- 6. Hinge knee systems are designed for use in patients in primary and revision surgery. where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are absent or incompetent.
The Legion Hinge Knee System is for Cemented Use Only.
{19}------------------------------------------------
# Legion
- 1. Rheumatoid arthritis.
- 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
- 4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
- 5. Constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral ligament) are incompetent.
- 6. Hinge knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e. medial collateral and/or lateral collateral ligament) are absent or incompetent.
The LEGION Knee System - Finned Tibial Wedges are for single use only and are intended for implantation with bone cement.
# Journey BCS and Journey II
- 1. Rheumatoid arthritis.
- 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
- 4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
Smith & Nephew, Inc. Journey II BCS Knee components are indicated for use only with cement and are single use devices.
# Anthem Total Knee System
- 1. Rheumatoid arthritis.
- 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
- 4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
The ANTHEM PS Total Knee System is indicated for use only with cement and is a single use device.
{20}------------------------------------------------
# Journey II CR Knee System
- 1. Rheumatoid arthritis.
- 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis.
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement.
- 4. Posterior stabilized knee systems are designed for use in patients in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
The Journey II Total Knee system components are indicated for use only with cement and are single use devices.
### Unicompartmental Knee Systems
Genesis Uni, Journey II Uni, Journey II UK, Journey Uni
Unicompartmental knee implants are indicated for restoring either compartment of a knee that has been affected by the following:
- 1. Noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis;
- 2. Correction of functional deformity;
- 3. Revision procedures where other treatments or devices have failed; and
- 4. Treatment of fractures that are unmanageable using other techniques
Unicompartmental knee implants are single use only and are intended for implantation only with bone cement. HA coated unicompartmental knee implants are available outside the US for use without bone cement.
# Journey Patello-Femoral Replacement System
- 1. Degenerative arthritis in the distal femur and patella;
- 2. A history of patellar dislocation or patellar fracture; and
- 3. Failed previous surgery (arthroplasty, tibial tubercule elevation, lateral release) where pain, deformity or dysfunction persists.
The Smith & Nephew Patello-Femoral Implants are intended for implantation with bone cement.
{21}------------------------------------------------
# Journey II XR Bi-Cruciate Retaining Knee System
Total knee components are indicated for rheumatoid arthritis; post-traumatic arthritis, osteoarthritis, or degenerative arthritis; failed osteotomies, unicompartmental replacement,or total knee replacement.
The Journey II Total Knee system components are indicated for use only with cement and are single use devices.
### ZUK Unicompartmental Knee System
These devices are indicated for patients with:
- 1. Painful and/or disabling knee joints due to osteoarthritis or traumatic arthritis.
- 2. Previous tibial condyle or plateu fractures with loss of anatomy or function.
- 3. Varus or Valgus deformities.
- 4. Revision of previous arthroplasty procedures.
The devices are indicated for cemented use only.
The Zimmer Unicompartmental Knee System is designed for use when load bearing ROM is expected to be less than or equal to 155 degrees.
### Engage™ Partial Knee System
The Engage™ Partial Knee System is intended for medial unicompartmental knee arthroplasty to treat one or more of the following conditions:
- 1. Moderately disabling joint disease of the knee resulting from painful osteo- or post traumatic arthritis.
- 2. Revision of previous unsuccessful surgical procedures, including prior unicompartmental knee arthroplasty
- 3. As an alternative to tibial osteotomy in patients with Unicompartmental osteoarthritis.
The femoral component and tibial tray are intended for cementless or cemented fixation. The porous surfaces of both the femoral and tibial tray components provide biological fixation when used in a cementless application. When the tibial tray is implanted without the use of bone cement, the tibial anchor should be used. When the tibial trav is implanted with bone cement, the tibial anchor should not be used.
{22}------------------------------------------------
# Porous Patella and Porous Tibial Baseplates
Total Knee components are indicated for:
- 1. Rheumatoid arthritis
- 2. Post- traumatic arthritis, osteoarthritis, or degenerative arthritis
- 3. Failed osteotomies, unicompartmental replacement, or total knee replacement
- 4. The posterior stabilized knee system is designed for use in primary and revision surgery, where the anterior and posterior cruciate ligaments are incompetent and the collateral ligaments remain intact.
LEGION POROUS CR Narrow Femoral Components are indicated for use without bone cement and are single use devices.
The Porous Patella and the Porous Tibia Baseplate are indicated for use with or without bone cement, and are single use devices.
# Technological Characteristics
The device design and material of the subject devices are the same as the predicate Smith & Nephew Knee Systems, including substantially equivalent letter to file cleared under the premarket notifications listed in Table 6.1.
# Performance Data
Below listed Magnetic Resonance Imaging (MRI) compatibility testing was conducted asper the FDA's guidance and the standards listed below .
- 1. Magnetically induced displacement force (ASTM F2052)
- 2. Magnetically induced torque (ASTM F2213)
- 3. Radiofrequency (RF) induced heating (ASTM F2182-19e2, IEC 60601-2-33, ISO/TS 10974:2018E)
- 4. MR image artifact (ASTM F2119)
MR Safety testing/ assessments support the appropriate MR parameters and symbols found in the subject device labeling.
As these products have been in distribution for several years, bacterial endotoxin testing was completed in previous submissions. The addition of the MR safety information to thelabeling does not affect the pyrogenicity, therefore, the subject Smith & Knee Systems are still expected to met the acceptable endotoxin limits as stated in the FDA Guidance, "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile," "Pyrogen and Endotoxins Testing: Questions and Answers," and ANSI/AAMI ST72.
{23}------------------------------------------------
# Substantial Equivalence Information
The subject Smith & Nephew Knee Systems are identical in design, technological characteristics, function of the devices, packaging and sterilization to the commercially available predicate devices, including substantially equivalent letter to file parts, listed in Table 6.1 below.
| Manufacturer | Description | Submission<br>Number | Clearance<br>Date |
|----------------------|----------------------------------------------------|------------------------------------------|------------------------------------------------------|
| Smith & Nephew,Inc. | ANTHEM Total Knee System | K142807<br>K170648 | 12/22/2014<br>05/11/2017 |
| Smith & Nephew,Inc. | Genesis Uni Knee System | K030301<br>K912735 | 02/25/2003<br>12/27/1991 |
| Smith & Nephew,Inc. | JOURNEY BCS and II Knee<br>Systems | K042515<br>K111711<br>K140555 | 03/14/2005<br>09/16/2011<br>05/29/2014 |
| Smith & Nephew,Inc. | JOURNEY II CR Knee System | K113482<br>K121443 | 02/27/2012<br>08/13/2012 |
| Smith & Nephew,Inc. | JOURNEY II Uni Knee System | K152315 | 10/28/2015 |
| Smith & Nephew, Inc. | JOURNEY II UK Knee System | K190085 | 2/11/2019 |
| Smith & Nephew,Inc. | JOURNEY II XR Bi-Cruciate<br>Retaining Knee System | K152726 | 10/21/2015 |
| Smith & Nephew,Inc. | JOURNEY PFJ System | K051086 | 05/31/2005 |
| Smith & Nephew,Inc. | JOURNEY Uni Knee System | K061011<br>K061779<br>K073175<br>K102069 | 07/11/2006<br>08/04/2006<br>12/20/2007<br>10/05/2010 |
| Smith & Nephew, Inc. | Legion Hinge Total Knee System | K151118<br>K081111 | 07/28/2015<br>07/23/2008 |
{24}------------------------------------------------
| Manufacturer | Description | Submission<br>Number | Clearance<br>Date |
|----------------------|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Smith & Nephew,Inc. | Legion Total Knee System | K041106<br>K043440<br>K060742<br>K071071<br>K072531<br>K073325<br>K091543<br>K100897<br>K112941<br>K162775<br>K180334 | 07/01/2004<br>02/18/2005<br>05/03/2006<br>09/19/2007<br>12/06/2007<br>12/20/2007<br>12/21/2009<br>05/13/2010<br>12/20/2011<br>01/13/2017<br>03/05/2018 |
| Smith & Nephew,Inc. | ZUK Unicompartmental Knee<br>System | K033363<br>K122529 | 01/16/2004<br>11/16/2012 |
| Smith & Nephew, Inc. | ENGAGE Partial Knee System | K190439 | 11/21/2019 |
| Smith & Nephew, Inc. | Porous Patella and Porous Tibial<br>Baseplates | K211221 | 10/01/2021 |
# Conclusion
In summary, the only differences between the subject devices, including substantially equivalent letter to file catalog items and the commercially available predicate devices were supporting MR safety testing/assessment and the addition of MR safety information to the labeling. These differences are not critical to the intended therapeutic, diagnostic, prosthetic, or surgical use of the subject devices. The subject Smith & Nephew Knee Systems are identical in function, design features, materials, sterilization, manufacturing methods and operational principles to what was previously 510(k) cleared and do not affect the safety and effectiveness of the subject devices when used as labeled. Due to the supporting documentation within this filing, it is conducted that the subject device(s) are substantially equivalent to the predicate device(s).